Low uncertainty determination of manganese and vanadium in environmental and biological reference materials by instrumental neutron activation analysis

被引:8
|
作者
Kucera, J [1 ]
Soukal, L [1 ]
机构
[1] Acad Sci Czech Republ, Inst Nucl Phys, CZ-25068 Rez, Czech Republic
来源
关键词
Manganese; Vanadium; Reference Material; Activation Analysis; Neutron Activation;
D O I
10.1007/s002160050725
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The elements Mn and V were determined by INAA in about 5 mg and 100 mg aliquots of NIST SRM 1648 to elucidate discrepancies between our previous results for the 0.5 mg to 15 mg aliquots and the NIST certified and/or information values. Simultaneously, other NIST SRMs 1633a, 2704, and BCR CRMs 038, 101 and 143 were also analyzed, Special attention was given to evaluating and minimizing uncertainties of all steps of analysis. Our results compared very well with the respective certified and/or information values (if available) of all SRMs and CRMs studied, except for NIST SRM 1648. For this SRM we have found significantly lower results than the NIST values which suggests that the NIST values are positively biased by about 10%. A new value for V in BCR CRM 143 was also obtained.
引用
收藏
页码:415 / 418
页数:4
相关论文