Simultaneous measurement of arterial and left ventricular pressure in conscious freely moving rats by telemetry

被引:8
|
作者
Segreti, Jason A. [1 ]
Polakowski, James S. [1 ]
Blomme, Eric A. [1 ]
King, Andrew J. [1 ]
机构
[1] Abbvie Inc, Res & Dev, N Chicago, IL 60064 USA
关键词
Left-ventricular contractility; Telemetry; Rat; Cardiovascular safety; BLOOD-PRESSURE; RADIOTELEMETRY;
D O I
10.1016/j.vascn.2016.01.003
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Comprehensive cardiovascular assessment in conscious rodents by utilizing telemetry has been limited by the restriction of current devices to one pressure channel. The purpose of this study was to test and validate a dual pressure transmitter that allows the simultaneous measurement of arterial pressure (AP) and left ventricular pressure (LVP) in conscious freely moving rats. Six rats were surgically implanted with dual pressure transmitters. Baseline hemodynamics and circadian rhythm were observed to return within 7 days. AP, heart rate (HR), LVP and indices of left ventricular contractility were stable and demonstrated a prominent circadian rhythm over a two-week period of uninterrupted recordings. Administration of the vasodilator nifedipine produced the anticipated dose-dependent decrease in AP which was accompanied by a baroreflex mediated increase in HR and cardiac contractility. The negative inotrope verapamil produced the expected dose-dependent decreases in AP and cardiac contractility. Finally, a terminal validation of the dual pressure transmitter was performed under anesthesia by measuring AP and LVP simultaneously via telemetry and from a fluid filled arterial catheter and an intraventricular Millar catheter, respectively. A range of pressures and cardiac contractility were studied by administering sequential intravenous infusions of the positive inotrope dobutamine followed by verapamil. Linear regression analysis revealed a high level of agreement between pressures measured by the dual pressure transmitter and the exteriorized catheters. Histopathologic analysis of the heart revealed mild peri-catheter fibrosis. In conclusion, the simultaneous measurement of AP and LVP offers the potential for more detailed cardiovascular assessment in conscious rats. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:23 / 33
页数:11
相关论文
共 50 条
  • [1] Left ventricular pressure measurement in conscious rats via telemetry
    Guild, Sarah-Jane
    Lim, Mathew
    Pauly, Bob
    McCormick, Daniel
    Kirton, Robert
    Budgett, David
    Kondo, Masahiro
    Stehlin, Ellyce
    Barrett, Carolyn J.
    Malpas, Simon C.
    [J]. FASEB JOURNAL, 2011, 25
  • [2] Chronic measurement of left ventricular pressure in freely moving rats
    Stehlin, Ellyce
    Malpas, Simon C.
    Budgett, David M.
    Barrett, Carolyn J.
    McCormick, Daniel
    Whalley, Gillian
    Fu, Fumin
    Beil, Michael
    Rigel, Dean F.
    Guild, Sarah-Jane
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2013, 115 (11) : 1672 - 1682
  • [3] Dual recordings of left ventricular and arterial pressure in conscious rats via telemetry
    McCormick, Daniel
    Guild, Sarah-Jane
    Lim, Mathew
    Pauly, Bob
    Kirton, Robert
    Budgett, David
    Barrett, Carolyn J.
    Kondo, Masahiro
    Stehlin, Ellyce
    Malpas, Simon C.
    [J]. FASEB JOURNAL, 2011, 25
  • [4] Dual recordings of left ventricular and arterial pressure in conscious rats via telemetry
    Malpas, Simon
    Stehlin, Ellyce
    Guild, Sarah-Jane
    Budgett, David
    McCormick, Daniel
    Rigel, Dean
    [J]. JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 2012, 66 (02) : 199 - 200
  • [5] A self-calibrating telemetry system for measurement of ventricular pressure-volume relations in conscious, freely moving rats
    Uemura, K
    Kawada, T
    Sugimachi, M
    Zheng, C
    Kashihara, K
    Sato, T
    Sunagawa, K
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2004, 287 (06): : H2906 - H2913
  • [6] Telemetry monitoring of pulmonary arterial pressure in freely moving rats
    Hess, P
    Clozel, M
    Clozel, JP
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1996, 81 (02) : 1027 - 1032
  • [7] Measurement of right ventricular pressure by telemetry in conscious moving rabbits
    Ronisz, Alicja
    Delcroix, Marion
    Quarck, Rozenn
    [J]. LABORATORY ANIMALS, 2013, 47 (03) : 184 - 193
  • [8] Simultaneous Pulmonary and Systemic Blood Pressure and ECG Interval Measurement in Conscious, Freely Moving Rats
    Rey, Markus
    Weber, Edgar W.
    Hess, Patrick D.
    [J]. JOURNAL OF THE AMERICAN ASSOCIATION FOR LABORATORY ANIMAL SCIENCE, 2012, 51 (02): : 231 - 238
  • [9] Intracranial pressure and mean arterial pressure monitoring in freely moving rats via telemetry; pilot study
    Kozler, Petr
    Maresova, Dana
    Pokorny, Jaroslav
    [J]. NEUROENDOCRINOLOGY LETTERS, 2019, 40 (7-8) : 319 - 324
  • [10] TELEMETRY OF CARDIAC LEFT-VENTRICULAR PRESSURE IN CONSCIOUS DOGS
    GEROLD, M
    EIGENMANN, R
    WOLFGANG, H
    [J]. BIOTELEMETRY AND PATIENT MONITORING, 1981, 8 (1-2) : 28 - 32