Measurement of the three-dimensional distribution of radiation dose in grid therapy

被引:19
|
作者
Trapp, JV
Warrington, AP
Partridge, M
Philps, A
Glees, J
Tait, D
Ahmed, R
Leach, MO
Webb, S
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Fdn Trust, Sutton SM2 5PT, Surrey, England
[3] RMIT Univ, Dept Appl Phys, Melbourne, Vic 3001, Australia
[4] Inst Canc Res, Dept Radiotherapy, Sutton SM2 5PT, Surrey, England
[5] Inst Canc Res, Canc Res UK Clin MR Res Grp, Sutton SM2 5PT, Surrey, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2004年 / 49卷 / 19期
关键词
D O I
10.1088/0031-9155/49/19/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A single large dose of megavoltage x-rays delivered through a grid is currently being utilized by some centres for palliative radiotherapy treatments of large tumours. In this note, we investigate the dosimetry of grid therapy using two-dimensional film dosimetry and three-dimensional gel dosimetry. It is shown that the radiation dose is attenuated more rapidly with depth in a grid field than an open field, and that even shielded regions receive approximately 25% of the dose to the unshielded areas.
引用
收藏
页码:N317 / N323
页数:7
相关论文
共 50 条
  • [1] Three-dimensional dose verification for radiation therapy
    Guo, P.
    Adamovics, J.
    Oldham, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S710 - S711
  • [2] Neural Representation for Three-Dimensional Dose Distribution and its Applications in Precision Radiation Therapy
    Vasudevan, V.
    Shen, L.
    Huang, C.
    Chuang, C.
    Islam, M. T.
    Ren, H.
    Yang, Y.
    Dong, P.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E552 - E552
  • [3] Three-dimensional cluster formation and structure in heterogeneous dose distribution of intensity modulated radiation therapy
    Chao, Ming
    Wei, Jie
    Narayanasamy, Ganesh
    Yuan, Yading
    Lo, Yeh-Chi
    Penagaricano, Jose A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 (02) : 197 - 205
  • [4] Measurement of Three-Dimensional Dose Distribution Using Scintillator Imaging System
    Tsuneda, M.
    Nishio, T.
    Ezura, T.
    Karasawa, K.
    MEDICAL PHYSICS, 2019, 46 (06) : E513 - E513
  • [5] Three-dimensional measurement of dose distribution produced by an intensity modulated radiation therapy system using magnetic resonance of BANG polymer gel dosimeters
    Wong, PS
    Lin, CY
    Cardenas, RL
    Cheng, KK
    RADIOLOGY, 1998, 209P : 397 - 397
  • [6] Three-dimensional conformal radiation therapy
    Leibel, SA
    RADIATION PROTECTION IN MEDICINE: CONTEMPORARY ISSUES, 1999, (21): : 229 - 260
  • [7] Three-dimensional optical tomography of dose distribution using radiation sensitive transparent gels
    Jenneson, PM
    Bero, MA
    Gilboy, WB
    Doran, SJ
    Morton, EJ
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 538 - 540
  • [8] Accuracy of Absolute Three-Dimensional Dose Distribution Measurement Using the Delta4
    Uehara, R.
    Tachibana, H.
    Ohyoshi, H.
    Matsumoto, S.
    Baba, H.
    Tanaka, F.
    Ariji, T.
    MEDICAL PHYSICS, 2015, 42 (06) : 3351 - 3351
  • [9] Three-dimensional conformal radiation therapy for ependymoma
    Merchant, Thomas E.
    CHILDS NERVOUS SYSTEM, 2009, 25 (10) : 1261 - 1268
  • [10] Three-dimensional approach of breast radiation therapy
    Huibers, M
    Berfelo, S
    de Groot, L
    Valkering, I
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S464 - S464