Surveillance of Endoscopes: Comparison of Different Sampling Techniques

被引:25
|
作者
Cattoir, Lien [1 ]
Vanzieleghem, Thomas [2 ]
Florin, Lisa [1 ]
Helleputte, Tania [3 ]
De Vos, Martine [3 ]
Verhasselt, Bruno [1 ,4 ]
Boelens, Jerina [1 ,4 ,5 ]
Leroux-Roels, Isabel [1 ,4 ,5 ]
机构
[1] Ghent Univ Hosp, Dept Lab Med, Ghent, Belgium
[2] OneLIFE, Louvain La Neuve, Belgium
[3] Ghent Univ Hosp, Dept Gastroenterol, Endoscopy Unit, Ghent, Belgium
[4] Univ Ghent, Dept Clin Chem Microbiol & Immunol, Ghent, Belgium
[5] Ghent Univ Hosp, Infect Control Team, Ghent, Belgium
来源
关键词
FLEXIBLE GASTROINTESTINAL ENDOSCOPES; ADENOSINE-TRIPHOSPHATE; INFECTION; GUIDELINE; BIOFILMS; QUALITY; DISINFECTION; CULTURES;
D O I
10.1017/ice.2017.115
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
OBJECTIVE To compare different techniques of endoscope sampling to assess residual bacterial contamination. DESIGN Diagnostic study. SETTING The endoscopy unit of an 1,100-bed university hospital performing similar to 13,000 endoscopic procedures annually. METHODS In total, 4 sampling techniques, combining flushing fluid with or without a commercial endoscope brush, were compared in an endoscope model. Based on these results, sterile physiological saline flushing with or without PULL THRU brush was selected for evaluation on 40 flexible endoscopes by adenosine triphosphate (ATP) measurement and bacterial culture. Acceptance criteria from the French National guideline (<25 colony-forming units [CFU] per endoscope and absence of indicator microorganisms) were used as part of the evaluation. RESULTS On biofilm-coated PTFE tubes, physiological saline in combination with a PULL THRU brush generated higher mean ATP values (2,579 relative light units [RLU]) compared with saline alone (1,436 RLU; P=.047). In the endoscope samples, culture yield using saline plus the PULL THRU (mean, 43 CFU; range, 1-400 CFU) was significantly higher than that of saline alone (mean, 17 CFU; range, 0-500 CFU; P<.001). In samples obtained using the saline+PULL THRU brush method, ATP values of samples classified as unacceptable were significantly higher than those of samples classified as acceptable (P=.001). CONCLUSION Physiological saline flushing combined with PULL THRU brush to sample endoscopes generated higher ATP values and increased the yield of microbial surveillance culture. Consequently, the acceptance rate of endoscopes based on a defined CFU limit was significantly lower when the saline+PULL THRU method was used instead of saline alone. Infect Control Hosp Epidemiol 2017;38:1062-1069
引用
收藏
页码:1062 / 1069
页数:8
相关论文
共 50 条
  • [1] A Comparison of Communication Techniques for Capsule Endoscopes
    Kim, Kiyun
    Won, Kyunghoon
    Shin, Jooyoup
    Choi, Hyung-Jin
    17TH ASIA-PACIFIC CONFERENCE ON COMMUNICATIONS (APCC 2011), 2011, : 761 - 764
  • [2] Comparison of the Jitter Performance of different Photonic Sampling Techniques
    Krune, Edgar
    Krueger, Benjamin
    Zimmermann, Lars
    Voigt, Karsten
    Petermann, Klaus
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2016, 34 (04) : 1360 - 1367
  • [3] Transabdominal chorionic villous sampling (CVS): comparison of two different techniques
    Patane, Luisa
    Cavalli, Giorgia
    Cattaneo, Daniela
    Strobelt, Nicola
    Barresi, Santa
    Ciriello, Elena
    Comerio, Daniela
    Comotti, Stefano
    Frigerio, Luigi
    Fruscella, Paolo
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2014, 210 (01) : S125 - S125
  • [4] Comparison Of The Different Sampling Techniques For Imbalanced Classification Problems In Machine Learning
    Peng Zhihao
    Yan Fenglong
    Li Xucheng
    2019 11TH INTERNATIONAL CONFERENCE ON MEASURING TECHNOLOGY AND MECHATRONICS AUTOMATION (ICMTMA 2019), 2019, : 431 - 434
  • [5] COMPARISON OF WASTEWATER SAMPLING TECHNIQUES
    TARAZI, DS
    HISER, LL
    CHILDERS, RE
    BOLDT, CA
    JOURNAL WATER POLLUTION CONTROL FEDERATION, 1970, 42 (05): : 708 - &
  • [6] In vitro comparison of 3 different brushes for manual cleaning of endoscopes
    Colman, Sofie
    Vanzieleghem, Thomas
    Leroux-Roels, Isabel
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2022, 43 (08): : 1076 - 1078
  • [7] COMPARISON BETWEEN DIFFERENT SAMPLING TECHNIQUES IN DETERMINING THE EMPIDID FAUNA (INSECTA, DIPTERA)
    GROOTAERT, P
    POLLET, M
    ANNALES DE LA SOCIETE ROYALE ZOOLOGIQUE DE BELGIQUE, 1988, 118 (01): : 79 - 79
  • [8] Comparison of several techniques for sampling macroinvertebrates in different habitats of a North Iberian pond
    F. García-Criado
    C. Trigal
    Hydrobiologia, 2005, 545 : 103 - 115
  • [9] Comparison of high-speed terahertz optical sampling techniques at different wavelengths
    Zouaghi, W.
    Ospald, F.
    Molter, D.
    Beigang, R.
    2012 37TH INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER, AND TERAHERTZ WAVES (IRMMW-THZ), 2012,
  • [10] Comparison of different sampling techniques for the evaluation of pesticide spray drift in apple orchards
    Briand, O
    Bertrand, F
    Seux, R
    Millet, M
    SCIENCE OF THE TOTAL ENVIRONMENT, 2002, 288 (03) : 199 - 213