Gaze-Contingent Flicker Pupil Perimetry Detects Scotomas in Patients With Cerebral Visual Impairments or Glaucoma

被引:25
|
作者
Naber, Marnix [1 ]
Roelofzen, Carlien [1 ,2 ]
Fracasso, Alessio [2 ,3 ]
Bergsma, Douwe P. [4 ]
van Genderen, Mies [5 ]
Porro, Giorgio L. [6 ]
Dumoulin, Serge O. [1 ,2 ,7 ]
机构
[1] Univ Utrecht, Helmholtz Inst, Expt Psychol, Utrecht, Netherlands
[2] Royal Netherlands Acad Arts & Sci, Spinoza Ctr Neuroimaging, Amsterdam, Netherlands
[3] Univ Glasgow, Inst Neurosci & Psychol, Glasgow, Lanark, Scotland
[4] Radboud Univ Nijmegen Med Ctr, Dept Cognit Neurosci, Nijmegen, Netherlands
[5] Bartimeus Diagnost Ctr Complex Visual Disorders, Zeist, Netherlands
[6] Univ Med Ctr Utrecht, Ophthalmol, Utrecht, Netherlands
[7] Vrije Univ Amsterdam, Expt & Appl Psychol, Amsterdam, Netherlands
来源
FRONTIERS IN NEUROLOGY | 2018年 / 9卷
关键词
pupillary response; perimetry; open-angle glaucoma; cerebral visual impairment; pupillometry; neuro-ophthalmic disease; visual field defect; PUPILLOGRAPHIC OBJECTIVE PERIMETRY; FIELD LOSS; LIGHT RESPONSE; CAMPIMETRY; VARIABILITY; CHILDREN; DEFECTS; STIMULI; VISION; REFLEX;
D O I
10.3389/fneur.2018.00558
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: The pupillary light reflex is weaker for stimuli presented inside as compared to outside absolute scotomas. Pupillograph perimetry could thus be an objective measure of impaired visual processing. However, the diagnostic accuracy in detecting scotomas has remained unclear. We quantitatively investigated the accuracy of a novel form of pupil perimetry. Methods: The new perimetry method, termed gaze-contingent flicker pupil perimetry, consists of the repetitive on, and off flickering of a bright disk (2 hz; 320 cd/m(2); 4 degrees diameter) on a gray background (160 cd/m(2)) for 4 seconds per stimulus location. The disk evokes continuous pupil oscillations at the same rate as its flicker frequency, and the oscillatory power of the pupil reflects visual sensitivity. We monocularly presented the disk at a total of 80 locations in the central visual field (max. 15 degrees). The location of the flickering disk moved along with gaze to reduce confounds of eye movements (gaze-contingent paradigm). The test lasted similar to 5 min per eye and was performed on 7 patients with cerebral visual impairment (CVI), 8 patients with primary open angle glaucoma (age > 45), and 14 healthy, age/gender-matched controls. Results: For all patients, pupil oscillation power (FFT based response amplitude to flicker) was significantly weaker when the flickering disk was presented in the impaired as compared to the intact visual field (CVI: 12%, AUC = 0.73; glaucoma: 9%, AUC = 0.63). Differences in power values between impaired and intact visual fields of patients were larger than differences in power values at corresponding locations in the visual fields of the healthy control group (CVI: AUC = 0.95; glaucoma: AUC = 0.87). Pupil sensitivity maps highlighted large field scotomas and indicated the type of visual field defect (VFD) as initially diagnosed with standard automated perimetry (SAP) fairly accurately in CVI patients but less accurately in glaucoma patients. Conclusions: We provide the first quantitative and objective evidence of flicker pupil perimetry's potential in detecting CVI-and glaucoma-induced VFDs. Gaze-contingent flicker pupil perimetry is a useful form of objective perimetry and results suggest it can be used to assess large VFDs with young CVI patients whom are unable to perform SAP.
引用
收藏
页数:12
相关论文
共 2 条
  • [1] Visual performance at the PRL in monocular versus binocular viewing using simulated gaze-contingent scotomas
    Martire, Joella
    Arango, Tiffany
    Bex, Peter J.
    Ross, Nicole C.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [2] Making the invisible visible: Exploration of different gaze-contingent simulated scotomas on facial recognition, visual search and reading
    Munoz, Joel
    Bex, Peter J.
    Cui, Bliss
    Idman-Rait, Cecilia
    Ross, Nicole
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)