Glucose Monitoring in Individuals With Diabetes Using a Long-Term Implanted Sensor/Telemetry System and Model

被引:61
|
作者
Lucisano, Joseph Y. [1 ]
Routh, Timothy L. [2 ,3 ]
Lin, Joe T. [1 ]
Gough, David A. [1 ,4 ]
机构
[1] GlySens Inc, San Diego, CA 92121 USA
[2] GlySens Inc, Prod Dev, San Diego, CA USA
[3] GlySens Inc, Operat & Consumer Prod, San Diego, CA USA
[4] Univ Calif San Diego, Dept Bioengn, La Jolla, CA 92093 USA
基金
美国国家卫生研究院;
关键词
Blood glucose sensor signal model; fully implanted sensor/telemetry system; long-term glucose monitoring; BLOOD; ACCURACY; SENSOR;
D O I
10.1109/TBME.2016.2619333
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The use of a fully implanted firstgeneration prototype sensor/telemetry system is described for long-term monitoring of subcutaneous tissue glucose in a small cohort of people with diabetes. Methods: Sensors are based on a membrane containing immobilized glucose oxidase and catalase coupled to oxygen electrodes and a telemetry system, integrated as an implant. The devices remained implanted for up to 180 days, with signals transmitted every 2 min to external receivers. Results: The data include signal recordings from glucose clamps and spontaneous glucose excursions, matched, respectively, to reference blood glucose and finger-stick values. The sensor signals indicate dynamic tissue glucose, for which there is no independent standard, and a model describing the relationship between blood glucose and the signal is, therefore, included. The values of all model parameters have been estimated, including the permeability of adjacent tissues to glucose, and equated to conventionalmass transfer parameters. As a group, the sensor calibration varied randomly at an average rate of -2.6%/week. Statistical correlation indicated strong association between the sensor signals and reference glucose values. Conclusion: Continuous long-term glucose monitoring in individuals with diabetes is feasible with this system. Significance: All therapies for diabetes are based on glucose control, and therefore, require glucose monitoring. This fully implanted long-term sensor/telemetry system may facilitate a new era of management of the disease.
引用
收藏
页码:1982 / 1993
页数:12
相关论文
共 50 条
  • [1] Glucose Variability of Individuals without Diabetes Using a Long-Term Continuous Glucose Monitoring System
    Deiss, Dorothee
    Abtahi, Tahmid
    Rastogi, Ravi
    Kelley, E. L.
    [J]. DIABETES, 2018, 67
  • [2] Development of the telemetry system for the long-term borehole monitoring system
    Namba, Y.
    Ito, H.
    Kyo, K.
    Kato, K.
    Sugiyama, H.
    [J]. OCEANS 2008 - MTS/IEEE KOBE TECHNO-OCEAN, VOLS 1-3, 2008, : 318 - +
  • [3] Long-term blood glucose monitoring with implanted telemetry device in conscious and stress-free cynomolgus monkeys
    Wang, B.
    Sun, G.
    Qiao, W.
    Liu, Y.
    Qiao, J.
    Ye, W.
    Wang, H.
    Wang, X.
    Lindquist, R.
    Wang, Y.
    Xiao, Y. -F.
    [J]. JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION, 2017, 40 (09): : 967 - 977
  • [4] Long-Term Continuous Glucose Monitoring in Conscious Stress-Free Nonhuman Primates with Implanted Telemetry Device
    Wang, Bingdi
    Sun, Guofeng
    Wang, Xiaoli
    Ding, Xunhua
    Tang, Weimin
    Wang, Yi-Xin
    Xiao, Yong-Fu
    [J]. DIABETES, 2015, 64 : A246 - A246
  • [5] Long-term blood glucose monitoring with implanted telemetry device in conscious and stress-free cynomolgus monkeys
    B. Wang
    G. Sun
    W. Qiao
    Y. Liu
    J. Qiao
    W. Ye
    H. Wang
    X. Wang
    R. Lindquist
    Y. Wang
    Y.-F. Xiao
    [J]. Journal of Endocrinological Investigation, 2017, 40 : 967 - 977
  • [6] Mouse telemetry for long-term monitoring
    Kramer, K
    Hachtman, S
    [J]. PHYSIOLOGICAL GENOMICS OF THE CRITICALLY ILL MOUSE, 2004, 16 : 127 - 141
  • [7] IMPLANTABLE TELEMETRY SYSTEM FOR LONG-TERM PH MONITORING IN STOMACH
    BAURSCHMIDT, P
    GREIFENSTEIN, J
    STELZNER, F
    SCHALDACH, M
    [J]. BIOTELEMETRY, 1974, 1 (02) : 86 - 87
  • [8] Long-term monitoring of intracranial pressure in freely-moving rats using telemetry system
    Eftekhari, Sajedeh
    Johansen, Katrine P.
    Bruun, Signe Rath
    Jensen, Rigmor H.
    [J]. CEPHALALGIA, 2019, 39 : 143 - 143
  • [9] Long-Term Continuous Glucose Monitoring Using a Fluorescence-Based Biocompatible Hydrogel Glucose Sensor
    Sawayama, Jun
    Takeuchi, Shoji
    [J]. ADVANCED HEALTHCARE MATERIALS, 2021, 10 (03)
  • [10] A telemetry-instrumentation system for long-term implantable glucose and oxygen sensors
    Salehi, C
    Atanasov, P
    Yang, SP
    Wilkins, E
    [J]. ANALYTICAL LETTERS, 1996, 29 (13) : 2289 - 2308