Reliability and Validity of In Vitro Volume Calculations by 3-Dimensional Ultrasonography Using the Multiplanar, Virtual Organ Computer-Aided Analysis (VOCAL), and Extended Imaging VOCAL Methods

被引:2
|
作者
de Sa Barreto, Enoch Quindere [1 ]
Figuinha Milani, Herbene Jose [1 ]
Araujo Junior, Edward [1 ]
Haratz, Karina Krajden [1 ]
Rolo, Liliam Cristine [1 ]
Machado Nardozza, Luciano Marcondes [1 ]
Moron, Antonio Fernandes [1 ]
机构
[1] Univ Fed Sao Paulo, Dept Obstet, Sao Paulo, Brazil
关键词
extended imaging virtual organ computer-aided analysis; in vitro assessment; multiplanar; reliability; validity; virtual organ computer-aided analysis; SONOGRAPHIC MEASUREMENT; LIVER VOLUME; ULTRASOUND; ACCURACY; REPRODUCIBILITY; PREGNANCIES; AGREEMENT; DISTANCE; AGE;
D O I
10.7863/jum.2010.29.5.767
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective. The purpose of this study was to assess the reliability and validity of in vitro volume calculations by 3-dimensional ultrasonography. Methods. This observational study was performed by 2 examiners to obtain volumes of 3 objects of different shapes and sizes filled with ultrasound gel and immersed in water. The examiners used the multiplanar (5-mm interval), virtual organ computer-aided analysis (VOCAL, 300) and extended imaging (XI) VOCAL (5, 10, 15, and 20 planes) methods to estimate the volumes of each object. A paired Student t test (P) and intraclass correlation coefficients (ICCs) were used to assess reproducibility of the methods. Validity was assessed comparing the percent differences between the estimated and the real volumes using the P value, mean differences, and ICC for each method. Results. All methods were highly reliable and valid. There were no significant differences in interobserver variability; there was a strong interobserver correlation. There were no significant differences in the percent differences between the estimated and real volumes of the objects using the 3 methods. The XI VOCAL method was superior to the multiplanar and VOCAL methods in the measurement of irregularly shaped objects. The XI VOCAL method with 10 planes estimated volumes closest to the real volumes. Conclusions. All 3 methods were reliable and valid; however, XI VOCAL was superior to the other methods in the measurement of irregularly shaped objects.
引用
收藏
页码:767 / 774
页数:8
相关论文
共 42 条
  • [1] Nomogram of fetal upper arm volume by three-dimensional ultrasound using extended imaging virtual organ computer-aided analysis (XI VOCAL)
    Cavalcante, Rafael Oliveira
    Araujo Junior, Edward
    Machado Nardozza, Luciano Marcondes
    Rolo, Liliam Cristine
    Moron, Antonio Fernandes
    [J]. JOURNAL OF PERINATAL MEDICINE, 2011, 39 (06) : 717 - 724
  • [2] Use of virtual organ computer-aided analysis (VOCAL) for volume calculation compared to traditional volume calculations of fetal thoracic tumors
    Ehrig, Jessica
    Paulsen, Karen
    Tong, Suhong
    Meier, Maxene R.
    Lagueux, Megan
    Galan, Henry L.
    Zaretsky, Michael V.
    Behrendt, Nicholas
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2018, 218 (01) : S257 - S257
  • [3] The reliability of virtual organ computer-aided analysis (VOCAL) for the semiquantification of ovarian, endometrial and subendometrial perfusion
    Raine-Fenning, NJ
    Campbell, BK
    Clewes, JS
    Kendall, NR
    Johnson, IR
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (06) : 633 - 639
  • [4] Comparison of gestational sac volume by 3D-sonography using planimetric, virtual organ computer-aided analysis and extended imaging virtual organ computer-aided analysis methods between 7 and 11 weeks of pregnancy
    Machado Nardozza, Luciano Marcondes
    Rolo, Liliam Cristine
    Araujo Junior, Edward
    Nowak, Paulo Martin
    Bortoletti Filho, Joao
    Moron, Antonio Fernandes
    [J]. ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 2010, 89 (03) : 328 - 334
  • [5] A nomogram of fetal lung volumes estimated by 3-dimensional ultrasonography using the rotational technique (virtual organ computer-aided analysis)
    Ruano, R
    Joubin, L
    Aubry, MC
    Thalabard, JC
    Dommergues, M
    Dumez, Y
    Benachi, A
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2006, 25 (06) : 701 - 709
  • [6] Reproducibility of transvaginal three-dimensional endometrial volume measurements with virtual organ computer-aided analysis (VOCAL) during ovarian stimulation
    Bordes, A
    Bory, AM
    Benchaïb, M
    Rudigoz, RC
    Salle, B
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2002, 19 (01) : 76 - 80
  • [7] Normal fetal urine production rate estimated with 3-dimensional ultrasonography using the rotational technique (virtual organ computer-aided analysis)
    Touboul, Cyril
    Boulvain, Michel
    Picone, Olivier
    Levaillant, Jean-Marc
    Frydman, Rene
    Senat, Marie-Victoire
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2008, 199 (01) : 57.e1 - 57.e5
  • [8] Assessment of Intracranial Structure Volumes in Fetuses With Growth Restriction by 3-Dimensional Sonography Using the Extended Imaging Virtual Organ Computer-Aided Analysis Method
    Rabachini Caetano, Ana Carolina
    Perez Zamarian, Ana Cristina
    Araujo Junior, Edward
    Cavalcante, Rafael Oliveira
    Simioni, Christiane
    Silva, Carolina Pacheco
    Rolo, Liliam Cristine
    Moron, Antonio Fernandes
    Machado Nardozza, Luciano Marcondes
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2015, 34 (08) : 1397 - 1405
  • [9] Reference Intervals for Fetal Heart Volume From 3-Dimensional Sonography Using the Extended Imaging Virtual Organ Computer-Aided Analysis Method at Gestational Ages of 20 to 34 Weeks
    de Sa Barreto, Enoch Quindere
    Figuinha Milani, Herbene Jose
    Haratz, Karina Krajden
    Araujo Junior, Edward
    Machado Nardozza, Luciano Marcondes
    Moron, Antonio Fernandes
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2012, 31 (05) : 673 - 678