A pulsed Schlieren system for visualizing beams from phased-array HIFU applicators

被引:0
|
作者
Christensen, Douglas A. [1 ]
Chao, Aborn [1 ]
机构
[1] Univ Utah, Dept Bioengn, Salt Lake City, UT 84112 USA
关键词
Schlieren; ultrasound beam patterns; pulsed laser;
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
We have assembled a Schlieren system for imaging the beam patterns from HIFU transducers in water to verify the desired beam focusing and direction. It consists of a high power (65 mW) short-pulse (as short as 2 ns) laser emitting at 658 nm, a beam expander/spatial filter, water tank, 400 mu m round stop, and a 12-bit CCD camera (1035 x 1317 pixels). The laser is synchronized to the driving electronics of the transducer but its pulse is delayed, effectively freezing of the acoustic wavefronts in the imaged space.
引用
收藏
页码:15 / +
页数:2