Monte Carlo as quality control tool of stereotactic body radiation therapy treatment plans

被引:4
|
作者
Alhamada, Husein [1 ]
Simon, Stephane [2 ]
Gulyban, Akos [2 ]
Gastelblum, Pauline [3 ]
Pauly, Nicolas [1 ]
VanGestel, Dirk [2 ]
Reynaert, Nick [2 ]
机构
[1] ULB, Nucl Metrol Dept, Ecole Polytech, Brussels, Belgium
[2] Inst Jules Bordet, Radiotherapy Dept, Brussels, Belgium
[3] CHIREC Canc Inst, Radiotherapy Dept, Brussels, Belgium
关键词
Monte Carlo (MC); Treatment Planning System (TPS); DOSE CALCULATION; MEGAVOLTAGE PHOTON; CLINICAL IMPLEMENTATION; ELECTRON-BEAM; RADIOTHERAPY; DISTRIBUTIONS; SIMULATIONS; SENSITIVITY; IMRT; VMAT;
D O I
10.1016/j.ejmp.2021.02.025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose/objective: The objective of this study was to verify the accuracy of treatment plans of stereotactic body radiation therapy (SBRT) and to verify the feasibility of the use of Monte Carlo (MC) as quality control (QC) on a daily basis. Material/methods: Using EGSnrc, a MC model of Agility T linear accelerator was created. Various measurements (Percentage depth dose (PDD), Profiles and Output factors) were done for different fields sizes from 1x1 up to 40x40 (cm(2)). An iterative model optimization was performed to achieve adequate parameters of MC simulation. 40 SBRT patient's dosimetry plans were calculated by Monaco T 3.1.1. CT images, RT-STRUCT and RT-PLAN files from Monaco T being used as input for Moderato MC code. Finally, dose volume histogram (DVH) and paired t-tests for each contour were used for dosimetry comparison of the Monaco T and MC. Results: Validation of MC model was successful, as <2% difference comparing to measurements for all field's sizes. The main energy of electron source incident on the target was 5.8 MeV, and the full width at half maximum (FWHM) of Gaussian electron source were 0.09 and 0.2 (cm) in X and Y directions, respectively. For 40 treatment plan comparisons, the minimum absolute difference of mean dose of planning treatment planning (PTV) was 0.1% while the maximum was 6.3%. The minimum absolute difference of Max dose of PTV was 0.2% while the maximum was 8.1%. Conclusion: SBRT treatment plans of Monaco agreed with MC results. It possible to use MC for treatment plans verifications as independent QC tool.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 50 条
  • [1] Monte Carlo Dose Evaluation for Pancreatic Stereotactic Body Radiation Therapy
    Gan, H.
    Zheng, D.
    Lei, Y.
    Li, S.
    Zhang, Q.
    Zhou, S.
    Li, J.
    Lin, C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S363 - S363
  • [2] Validation of a commercial Monte Carlo code used for stereotactic radiosurgery and stereotactic body radiation therapy
    Milroy, D.
    Dos Santos, G.
    Seuntjens, J.
    Patrocinio, H.
    [J]. MEDICAL PHYSICS, 2012, 39 (07) : 4620 - 4620
  • [3] Monte Carlo characterization of target doses in stereotactic body radiation therapy (SBRT)
    Rassiah-Szegedi, Premavathy
    Salter, Bill J.
    Fuller, Clifton D.
    Blough, Melissa
    Papanikolaou, Niko
    Fuss, Martin
    [J]. ACTA ONCOLOGICA, 2006, 45 (07) : 989 - 994
  • [4] Characterization of doses in stereotactic body radiation therapy (SBRT) via Monte Carlo calculation
    Salter, B
    Rasiah, P
    Fuss, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S539 - S540
  • [5] Dosimetric Evaluation of Lung Stereotactic Body Radiation Therapy (SBRT) Treatment Plans
    Giaddui, T.
    Wang, B.
    Hollander, A.
    Micaily, B.
    Miyamoto, C.
    Li, S.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [6] Monte Carlo Calculations of Dose Distribution in NSCLC Patients Planned for Stereotactic Body Radiation Therapy
    Moiseenko, V.
    Popescu, I. A.
    Liu, M.
    Kristensen, S.
    Gill, B.
    Teke, T.
    Bergman, A.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [7] Formal quality control for a proton Monte Carlo system in radiation therapy
    Perl, J.
    [J]. XVII INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY (ICCR 2013), 2014, 489
  • [8] ExacTrac Snap Verification: A New Tool for Ensuring Quality Control for Lung Stereotactic Body Radiation Therapy
    Udrescu, Corina
    Mornex, Francoise
    Tanguy, Ronan
    Chapet, Olivier
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (01): : E89 - E94
  • [9] Stereotactic Body Radiation Therapy (SBRT) Using Robotic Radiosurgery for Lung Cancer: Comparison of Treatment Plans Using Ray Tracing (RT) and Monte Carlo (MC) Algorithm
    Pennington, A.
    Selvaraj, R.
    Leventouri, T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S913 - S914
  • [10] MCDOSE - A Monte Carlo dose calculation tool for radiation therapy treatment planning
    Ma, CM
    Li, JS
    Pawlicki, T
    Jiang, SB
    Deng, J
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 123 - 125