Derivatization-free multi-step extraction for trace haloacetic acids analysis with ion chromatography: Performance and mechanisms

被引:12
|
作者
Ma, Wei [1 ]
Li, Wenyu [1 ]
Yang, Yang [1 ]
Yang, Jie [1 ]
Chen, Baiyang [1 ]
Xie, Yuefeng [2 ]
机构
[1] Harbin Inst Technol, State Key Lab Urban Water Resource & Environm, Shenzhen Key Lab Organ Pollut Prevent & Control, Shenzhen 518055, Peoples R China
[2] Penn State Univ, Civil & Environm Engn Programs, Middletown, PA 17057 USA
基金
中国国家自然科学基金;
关键词
Haloacetic acids; De-extraction; Ion chromatography; Preconcentration; Anions separation; DISINFECTION BY-PRODUCTS; SOLID-PHASE MICROEXTRACTION; SINGLE-DROP MICROEXTRACTION; DRINKING-WATER; SENSITIVE METHOD; CHLOROACETIC ACIDS; MASS-SPECTROMETRY; TAP WATER; TOXICITY; SAMPLES;
D O I
10.1016/j.jhazmat.2022.129166
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Haloacetic acids (HAAs) are a type of disinfection byproducts commonly found in drinking water with carcinogenic, mutagenic, or teratogenic risks to humans. Currently, the analytical methods of trace HAAs are either labor-intensive or very expensive. We herein propose a facile multiple-step extraction strategy for HAAs analysis with common ion chromatography (IC). This study is based on a fundamental water chemistry principle that HAAs become protonated featuring positive logK(ow) values (> 0.34) under pH < pKa but deprotonated featuring negative logK(ow) values (<-2.37) under pH > pKa. By taking advantage of the species and property switches, HAAs can be extracted and enriched into methyl tert-butyl ether first at pH < 0.5 and then back-extracted into neutral water and enriched again. Equally important, interfering anions in IC chromatogram are eliminated because they have negative logK(ow) values. Verification results show that HAAs were enriched by 11.4 times in average while interfering anions were almost eliminated (> 99%). Although similar to USEPA Method 552.3 in method detection limits (0.033-0.246 mu g/L), recoveries (70%~110%), and relative standard deviations (< 9.91%), this method took <= 70 min to run a batch of samples without derivatization, which takes over 2 h. The methodology may be applicable to other pollutants that also have contrasting K-ow values at different pH.
引用
收藏
页数:10
相关论文
共 25 条
  • [1] Determination of Trace Haloacetic Acids in Drinking Water Using Ion Chromatography Coupled with Solid Phase Extraction
    Sun Yingxue
    Huang Jianjun
    Gu Ping
    [J]. CHINESE JOURNAL OF CHROMATOGRAPHY, 2006, 24 (03) : 298 - 301
  • [2] Trace analysis of very long chain free fatty acids in plasma by fluorogenic derivatization and liquid chromatography
    Chang, CF
    Wu, HL
    Wu, SM
    Chen, SH
    Kou, HS
    [J]. JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1998, 21 (05) : 669 - 679
  • [3] Rapid simultaneous analysis of 17 haloacetic acids and related halogenated water contaminants by high-performance ion chromatography-tandem mass spectrometry
    Xue, Runmiao
    Donovan, Ariel
    Shi, Honglan
    Yang, John
    Hua, Bin
    Inniss, Enos
    Eichholz, Todd
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (24) : 6613 - 6622
  • [4] Rapid simultaneous analysis of 17 haloacetic acids and related halogenated water contaminants by high-performance ion chromatography-tandem mass spectrometry
    Runmiao Xue
    Ariel Donovan
    Honglan Shi
    John Yang
    Bin Hua
    Enos Inniss
    Todd Eichholz
    [J]. Analytical and Bioanalytical Chemistry, 2016, 408 : 6613 - 6622
  • [5] SIMPLIFIED DETERMINATION OF FREE FATTY-ACIDS USING DERIVATIZATION TECHNIQUE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY THROUGH THE MODIFIED EXTRACTION COLUMN
    IKEDA, M
    SHIMADA, K
    SAKAGUCHI, T
    [J]. BUNSEKI KAGAKU, 1982, 31 (06) : E199 - E206
  • [6] Simple and sensitive analysis of long-chain free fatty acids in milk by fluorogenic derivatization and high-performance liquid chromatography
    Lu, CY
    Wu, HL
    Chen, SH
    Kou, HS
    Wu, SM
    [J]. JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2002, 50 (01) : 71 - 73
  • [7] Amino-functionalized graphene oxide/neutral alumina nanocomposite based solid-phase extraction coupled with ion chromatography-mass spectrometry for the determination of trace haloacetic acids in drinking water
    Zhong, Zhixiong
    Li, Gongke
    Shao, Yijuan
    Zhu, Binghui
    Liu, Zhe
    Deng, Jianchao
    Mo, Jingtian
    [J]. ANALYTICAL METHODS, 2017, 9 (16) : 2425 - 2432
  • [8] ANALYSIS OF FREE AMINO-ACIDS IN MICROBIALLY COLONIZED SANDSTONE BY PRECOLUMN PHENYL ISOTHIOCYANATE DERIVATIZATION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    SIEBERT, J
    PALMER, RJ
    HIRSCH, P
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1991, 57 (03) : 879 - 881
  • [9] Microwave-assisted one-step extraction-derivatization for rapid analysis of fatty acids profile in herbal medicine by gas chromatography-mass spectrometry
    Liu, Rui-Lin
    Zhang, Jing
    Mou, Zhao-Li
    Hao, Shuang-Li
    Zhang, Zhi-Qi
    [J]. ANALYST, 2012, 137 (21) : 5135 - 5143
  • [10] Improving methodological aspects of the analysis of five regulated haloacetic acids in water samples by solid-phase extraction, ion-pair liquid chromatography and electrospray tandem mass spectrometry
    Prieto-Blanco, M. C.
    Alpendurada, M. F.
    Lopez-Mahia, P.
    Muniategui-Lorenzo, S.
    Prada-Rodriguez, D.
    Machado, S.
    Goncalves, C.
    [J]. TALANTA, 2012, 94 : 90 - 98