Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function

被引:98
|
作者
Lee, Sangjin [2 ]
Kamdar, Forum [2 ]
Madlon-Kay, Richard [2 ]
Boyle, Andrew [2 ]
Colvin-Adams, Monica [2 ]
Pritzker, Marc [2 ]
John, Ranjit [1 ]
机构
[1] Univ Minnesota, Div Cardiothorac Surg, Dept Surg, Minneapolis, MN 55455 USA
[2] Univ Minnesota, Dept Med, Div Cardiol, Minneapolis, MN 55455 USA
来源
关键词
ventricular assist device; right ventricular function; tricuspid annular motion; heart failure; TRICUSPID ANNULAR MOTION; FAILURE; IMPLANTATION; PULSATILE; SUPPORT; DYSFUNCTION; TRANSPLANT; RECIPIENTS;
D O I
10.1016/j.healun.2009.11.599
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Continuous-flow devices have become the standard of care for mechanical circulatory support for end-stage heart failure patients because of improved survival and durability. The effects of these devices, such as the HeartMate II (HMII) left ventricular assist device (LVAD), on right ventricular (RV) function have not been evaluated in detail. This study evaluated the incidence of RV failure, alterations in RV function, severity of tricuspid regurgitation (TR), and cardiac hemodynamics after HMII implantation. METHODS: Echocardiograms (n = 22) and right heart catheterizations (n = 40) were performed before and after 4 to 6 months of HMII Support in 40 bridge-to-transplant patients. Right heart failure was defined as the requirement for inotropes and/or nitric oxide requirement after LVAD implantation for >14 days or the need for right-sided mechanical circulatory support. RESULTS: Overall, RV failure after HMII implantation occurred in 2 of 40 patients (5%). Significant improvements occurred in cardiac index, with reductions in right atrial pressure, RV stroke work index, tricuspid annular motion, mean pulmonary artery pressure, and pulmonary vascular resistance after HMII support. There was a trend towards reduction in TR after LVAD support (p = 0.075). CONCLUSIONS: The incidence of RV failure after support with continuous-flow devices such as the HMII is low. The favorable effects of the HMII on cardiac hemodynamics result in improved RV function, improved right- and left-sided hemodynamic profiles, and a reduction in TR severity. These findings may have important implications for LVAD patients needing longer-term support. J Heart Lung Transplant 2010;29:209-15 (C) 2010 International Society for Heart and Lung Transplantation. All rights reserved.
引用
收藏
页码:209 / 215
页数:7
相关论文
共 50 条
  • [1] Effects of the HeartMate II Continuous-Flow Left Ventricular Assist Device on RV Function
    Lee, Sangjin
    Kamdar, Forum
    Madlon-Kay, Richard
    John, Ranjit
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 53 (10) : A180 - A180
  • [2] Effects of the HeartMate II Continuous-Flow Left Ventricular Assist Device on RV Function
    Lee, S.
    Kamdar, F.
    Madlon-Kay, R.
    John, R.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : S277 - S277
  • [3] HeartMate® II continuous-flow left ventricular assist system
    Sheikh, Farooq H.
    Russell, Stuart D.
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2011, 8 (01) : 11 - 21
  • [4] Impact of continuous-flow left ventricular assist device support on right ventricular function
    Morgan, Jeffrey A.
    Paone, Gaetano
    Nemeh, Hassan W.
    Murthy, Raghav
    Williams, Celeste T.
    Lanfear, David E.
    Tita, Cristina
    Brewer, Robert J.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : 398 - 403
  • [5] Right Ventricular Function in Patients With Left Ventricular Assist Device Support by Pulsatile Polvad MEV and Continuous-Flow Pumps Heartware and Heartmate II
    Nadziakiewicz, P.
    Borkowski, J.
    Szygula-Jurkiewicz, B.
    Niklewski, T.
    Pacholewicz, J.
    Zakliczynski, M.
    Hrapkowic, T.
    Zennbala, M.
    [J]. TRANSPLANTATION PROCEEDINGS, 2016, 48 (05) : 1786 - 1790
  • [6] Device thrombosis in HeartMate II continuous-flow Left ventricular assist devices: A multifactorial phenomenon
    Uriel, Nir
    Han, Jason
    Morrison, Kerry A.
    Nahumi, Nadav
    Yuzefpolskaya, Melana
    Garan, Arthur R.
    Duong, Jimmy
    Colombo, Paolo C.
    Takayama, Hiroo
    Thomas, Sunu
    Naka, Yoshifumi
    Jorde, Ulrich P.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (01): : 51 - 59
  • [7] Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: Incidence, risk factors, and effect on outcomes
    Kormos, Robert L.
    Teuteberg, Jeffrey J.
    Pagani, Francis D.
    Russell, Stuart D.
    John, Ranjit
    Miller, Leslie W.
    Massey, Todd
    Milano, Carmelo A.
    Moazami, Nader
    Sundareswaran, Kartik S.
    Farrar, David J.
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (05): : 1316 - 1324
  • [8] Continuous-Flow Left Ventricular Assist Device and the Right Ventricle
    Puhlman, Mark
    [J]. AACN ADVANCED CRITICAL CARE, 2012, 23 (01) : 86 - 90
  • [9] Impact of a continuous flow left ventricular assist device on right ventricular function
    Murthy, Raghav
    Morgan, Jeffrey
    Brewer, Robert
    Nemeh, Hassan
    Henry, Scott
    Williams, Celeste
    Lanfear, David
    Tita, Cristina
    Sabbah, Hani
    Paone, Gaetano
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2010, 211 (03) : S31 - S32
  • [10] Outcomes and Readmissions After Continuous Flow Left Ventricular Assist Device: Heartmate II Versus Heartware Ventricular Assist Device
    Tuncer, O. N.
    Kemaloglu, C.
    Erbasan, O.
    Golbasi, I.
    Turkay, C.
    Bayezid, O.
    [J]. TRANSPLANTATION PROCEEDINGS, 2016, 48 (06) : 2157 - 2161