Study on high resolution time interval measurement module in pulsed laser ranging system

被引:0
|
作者
Wu Gang [1 ]
Li Chun-Lai [1 ]
Liu Yin-Nian [1 ]
Dai Ning [1 ]
Wang Jian-Yu [1 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Tech Phys, Shanghai 200083, Peoples R China
关键词
pulsed laser ranging; time interval measurement; time-to-digital conversion; delay line interpolation method;
D O I
暂无
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The precision of the pulsed laser ranging system was decided by the precision of the time interval measurement. Therefore, a high resolution time interval measurement module was developed. The module is based on the special time-to-digital conversion chip which adopts the delay line interpolation method. The maximum measuring time of the module is 200ms, and the maximum time resolution is 125ps, of which the corresponding distance resolution is 18.75mm. The module is especially suit for the large distance measurement. The hardware and the software of the module as well as the testing results are also presented.
引用
收藏
页码:213 / +
页数:5
相关论文
共 7 条
  • [1] Chen YW, 2004, J INFRARED MILLIM W, V23, P169
  • [2] HOU YJ, 2001, LASER INFRARED, V31, P136
  • [3] Hu YH, 2001, J INFRARED MILLIM W, V20, P335
  • [4] HU YH, 1997, LASER TECHNOLOGY, V21, P189
  • [5] JIN GF, 1998, LASER MEASUREMENT, P719
  • [6] Xu DS, 2006, J INFRARED MILLIM W, V25, P77
  • [7] Zhang B, 2004, DYNAM CONT DIS SER A, V11, P31