Use of diverging apertures to minimize the edge scatter in passive scattering proton therapy

被引:6
|
作者
Zhao, Tianyu [1 ]
Cai, Bin [1 ]
Sun, Baozhou [1 ]
Grantham, Kevin [1 ]
Mutic, Sasa [1 ]
Klein, Eric [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63110 USA
来源
关键词
proton; passive scattering; aperture; edge scatter; COLLIMATOR SCATTER; DOSE CALCULATIONS; BEAMS;
D O I
10.1120/jacmp.v16i5.5675
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to evaluate the use of diverging-cut aperture to minimize collimator contamination in proton therapy. Two sets of apertures with nondivergent and divergent edge were fabricated to produce a 10 cm x 10 cm field at the radiation isocenter of a single-room proton therapy unit. Transverse profiles were acquired in a scanning water tank with both aperture sets. Up to 9.5% extra dose was observed from aperture scattering near the field edges with the nondivergent aperture set at 2 cm above the water surface and remained 3.0% at depth of 10 cm. For the divergent set, the contamination was reduced to less than 3.5% and 1.3%, respectively. Our study demonstrated that scattering from apertures contaminated the dose distribution near the field edge at shallow depth. A diverging-cut aperture was capable of reducing the contamination and is recommended for use in passive scattering proton therapy, especially when critical organs are lateral and proximal to the target at shallow depth.
引用
收藏
页码:367 / 372
页数:6
相关论文
共 50 条
  • [1] Conformal proton therapy with passive scattering
    Akulinichev, S. V.
    Ilic, R. D.
    Yakovlev, I. A.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S2 - S2
  • [2] Monte Carlo Investigation of Collimator Scatter of Proton-Therapy Beams Produced Using the Passive Scattering Method
    Titt, U.
    Zheng, Y.
    Vassiliev, O.
    Newhauser, W.
    MEDICAL PHYSICS, 2008, 35 (06) : 2797 - +
  • [3] Monte Carlo investigation of collimator scatter of proton-therapy beams produced using the passive scattering method
    Titt, Uwe
    Zheng, Yuanshui
    Vassiliev, Oleg N.
    Newhauser, Wayne D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (02): : 487 - 504
  • [4] Dosimetric impact of range uncertainty in passive scattering proton therapy
    Liu, Ruirui
    Sun, Baozhou
    Zhang, Tiezhi
    Williamson, Jeffery F.
    O'Sullivan, Joseph A.
    Zhao, Tianyu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (05): : 6 - 14
  • [5] Smearing Parameter for Smooth Compensator in Passive Scattering Proton Therapy
    Poenisch, F.
    Roemhild, P.
    Sahoo, N.
    Gillin, M.
    Zhu, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E267 - E267
  • [6] Considerations when treating lung cancer with passive scatter or active scanning proton therapy
    St James, Sara
    Grassberger, Clemens
    Lu, Hsiao-Ming
    TRANSLATIONAL LUNG CANCER RESEARCH, 2018, 7 (02) : 210 - 215
  • [7] Feasibility studies of a passive scatter proton therapy nozzle without a range modulator wheel
    Harvey, Mark C.
    Polf, Jerimy C.
    Smith, Alfred R.
    Mohan, Radhe
    MEDICAL PHYSICS, 2008, 35 (06) : 2243 - 2252
  • [8] The development of a treatment control system for a passive scattering proton therapy installation
    Hofverberg, P.
    Bergerot, J. -M.
    Trimaud, R.
    Herault, J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 1002
  • [9] Dose Evaluation From Scattered Protons in Passive Scattering Proton Therapy
    Kim, C.
    Lee, N.
    Lee, S.
    Jeong, J.
    Kim, H.
    Shin, D.
    Lim, Y.
    Shin, W.
    Min, C.
    MEDICAL PHYSICS, 2018, 45 (06) : E259 - E260
  • [10] An Evaluation of Patient Scatter Factor and Compensator Scatter Factor in Monitor Unit Calculation for Double Scattering Proton Therapy System
    Prusator, M.
    Ahmad, S.
    Chen, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E158 - E159