Safety and efficacy of high-intensity focused ultrasound in severe or refractory glaucoma

被引:7
|
作者
Hugo, Juliette [1 ]
Matonti, Frederic [2 ,3 ]
Beyleriani, Marie [1 ]
Zanin, Emilie [1 ]
Aptel, Florent [4 ,5 ]
Denise, Daniele [1 ]
机构
[1] Nord Hosp, AP HM, Dept Ophthalmol, F-13015 Marseille, France
[2] Aix Marseille Univ, CNRS, INT, Marseille, France
[3] Ctr Paradis Monticelli, Marseille, France
[4] Univ Hosp Grenoble, Dept Ophthalmol, Grenoble, France
[5] Univ Grenoble Alpes, Grenoble, France
关键词
Refractory glaucoma; ultrasonic cyclocoagulation; intraocular inflammation; intraocular pressure; uveoscleral pathway; ENDOTHELIAL-CELL LOSS; DIODE-LASER CYCLOPHOTOCOAGULATION; QUALITY-OF-LIFE; CILIARY BODIES; PHACOEMULSIFICATION; CYCLOCOAGULATION; RISK;
D O I
10.1177/1120672119874594
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The aim of this study was to evaluate the efficacy and safety of ultrasonic cyclocoagulation in severe or refractory glaucoma, and to analyze the procedure-related inflammation. Patients and Methods: In this retrospective study, 15 eyes of 13 patients suffering from severe or refractory glaucoma with uncontrolled intraocular pressure of > 21 mmHg underwent 8-s ultrasonic cyclocoagulation. A complete ophthalmic evaluation, quality of life assessment using Glau-Qol-17, endothelial cell count, and flare measurement were performed. Primary outcome was qualified surgical success (defined as intraocular pressure reduction from baseline > 20% and intraocular pressure >5 mmHg without hypotensive medication adjunction). Secondary outcomes were flare, endothelial cell loss, and quality of life. Results: Qualified success was achieved in 67% of eyes at 6 months (mean intraocular pressure reduction = 42% in these eyes). During the first month after the procedure, the mean flare reached its maximum value when the intraocular pressure was minimal; the flare slowly decreased until normalization at month 3 when the maximal intraocular pressure was noted. At month 3, there was moderate but significant endothelial cell loss (11%), and no significant alteration in quality of life was demonstrated. Conclusion: The efficacy of ultrasonic cyclocoagulation in the present study is comparable to that reported in the literature. The kinetics of intraocular pressure and flare suggest that postoperative inflammation could be partly responsible for the early intraocular pressure decrease. The moderate endothelial cell loss, which could be caused by localized heating, and the preservation of quality of life confirm the safety of ultrasonic cyclocoagulation.
引用
收藏
页码:130 / 137
页数:8
相关论文
共 50 条
  • [1] Safety and Efficacy of High-intensity Focused Ultrasound Cyclocoagulation in Refractory Glaucoma
    Mansouri, Kaweh
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [2] TREATMENT OF REFRACTORY GLAUCOMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    VALTOT, F
    KOPEL, J
    PETIT, E
    MOULIN, F
    HAUT, J
    [J]. JOURNAL FRANCAIS D OPHTALMOLOGIE, 1995, 18 (01): : 3 - 12
  • [3] Efficacy and safety of high-intensity focused ultrasound cyclo-plasty in glaucoma
    Chen, Di
    Guo, Xiu-Juan
    Luo, Shu-Ke
    Lu, Yan
    Tang, Xiu-Rong
    [J]. BMC OPHTHALMOLOGY, 2022, 22 (01)
  • [4] Efficacy and safety of high-intensity focused ultrasound cyclo-plasty in glaucoma
    Di Chen
    Xiu-Juan Guo
    Shu-Ke Luo
    Yan Lu
    Xiu-Rong Tang
    [J]. BMC Ophthalmology, 22
  • [5] High-intensity focused ultrasound treatment in patients with refractory glaucoma
    Giuseppe Giannaccare
    A. Vagge
    C. Gizzi
    A. Bagnis
    S. Sebastiani
    C. Del Noce
    M. Fresina
    C. E. Traverso
    E. C. Campos
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2017, 255 : 599 - 605
  • [6] High-intensity focused ultrasound treatment in patients with refractory glaucoma
    Giannaccare, Giuseppe
    Vagge, A.
    Gizzi, C.
    Bagnis, A.
    Sebastiani, S.
    Del Noce, C.
    Fresina, M.
    Traverso, C. E.
    Campos, E. C.
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2017, 255 (03) : 599 - 605
  • [7] Treatment of refractory glaucoma using high-intensity focused ultrasound cycloablation
    Del Noce, Chiara
    Bagnis, Alessandro
    Lester, Michele M.
    Traverso, Carlo Enrico
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [8] Safety and efficacy of multiple cyclocoagulation of ciliary bodies by high-intensity focused ultrasound in patients with glaucoma
    De Gregorio, Alessandra
    Pedrotti, Emilio
    Stevan, Giulia
    Montali, Margherita
    Morselli, Simonetta
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2017, 255 (12) : 2429 - 2435
  • [9] Safety and efficacy of multiple cyclocoagulation of ciliary bodies by high-intensity focused ultrasound in patients with glaucoma
    Alessandra De Gregorio
    Emilio Pedrotti
    Giulia Stevan
    Margherita Montali
    Simonetta Morselli
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2017, 255 : 2429 - 2435
  • [10] TREATMENT OF GLAUCOMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    VALTOT, F
    KOPEL, J
    HAUT, J
    [J]. INTERNATIONAL OPHTHALMOLOGY, 1989, 13 (1-2) : 167 - 170