Quality assurance of HDR 192Ir sources using a Fricke dosimeter

被引:10
|
作者
Austerlitz, C.
Mota, H.
Almeida, C. E.
Allison, R.
Sibata, C.
机构
[1] E Carolina Univ, Brody Sch Med, Dept Radiat Oncol, Greenville, NC 27858 USA
[2] Univ Estado Rio De Janeiro, Rio De Janeiro, RJ, Brazil
关键词
HDR; audit; quality control; Fricke; dosimetry; Ir-192;
D O I
10.1118/1.2714472
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A prototype of a Fricke dosimetry system consisting of a 15 X 15 X 15 cm 3 water phantom made of Plexiglas (R) and a 11.3-ml Pyrex balloon fitted with a 0.2 cm thick Pyrex sleeve in its center was created to assess source strength and treatment planning algorithms for use in high dose rate (HDR) Ir-192 afterloading units. In routine operation, the radioactive source is positioned at the end of a sleeve, which coincides with the center of the spherical balloon that is filled with Fricke solution, so that the solution is nearly isotropically irradiated. The Fricke system was calibrated in terms of source strength against a reference well-type ionization chamber, and in terms of radial dose by means of an existing algorithm from the HDR's treatment planning system. Because the system is based on the Fricke dosimeter itself, for a given type and model of Ir-192 source, the system needs initial calibration but no recalibration. The results from measurements made over a 10 month period, including source decay and source substitutions, have shown the feasibility of using such a system for quality control (QC) of HDR afterloading equipment, including both the source activity and treatment planning parameters. The benefit of a large scale production and the use of this device for clinical HDR QC audits via mail are also discussed. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1348 / 1353
页数:6
相关论文
共 50 条
  • [1] Determination of Dose at the Reference Point for 192Ir HDR Brachytherapy Sources Using Fricke Dosimetry
    Austerlitz, Carlos
    Benhabib, Sidi
    Zhu, Dengsong
    Mota, Helvecio
    Campos, Diana
    Allison, Ron
    Sibina, Claudio
    [J]. AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2008, 31 (06): : 617 - 617
  • [2] A Feasibility Study of Fricke Dosimetry as an Absorbed Dose to Water Standard for 192Ir HDR Sources
    deAlmeida, Carlos Eduardo
    Ochoa, Ricardo
    de Lima, Marilene Coelho
    David, Mariano Gazineu
    Pires, Evandro Jesus
    Peixoto, Jose Guilherme
    Salata, Camila
    Bernal, Mario Antonio
    [J]. PLOS ONE, 2014, 9 (12):
  • [3] Response of An Implantable MOSFET Dosimeter to 192Ir HDR Radiation
    Fagerstrom, J.
    Micka, J.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Response of an implantable MOSFET dosimeter to 192Ir HDR radiation
    Fagerstrom, Jessica M.
    Micka, John A.
    DeWerd, Larry A.
    [J]. MEDICAL PHYSICS, 2008, 35 (12) : 5729 - 5737
  • [5] On the Determination of the Absolute Dose for a 192Ir HDR Source Using a Fricke Dosimeter: Correction Factors for Several Radii Using MC Calculation
    Benhabib, S.
    Mota, H.
    Austerlitz, C.
    Ferreira, M.
    Allison, R.
    Sibata, C.
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 2522 - 2523
  • [6] Determination of the correction factors used in Fricke dosimetry for HDR 192Ir sources employing the Monte Carlo method
    David, Mariano G.
    Salata, Camila
    de Almeida, Carlos E.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 50 - 55
  • [7] Energy dependence of a radiophotoluminescent glass dosimeter for HDR 192Ir brachytherapy source
    Hashimoto, Shimpei
    Nakajima, Yujiro
    Kadoya, Noriyuki
    Abe, Kota
    Karasawa, Katsuyuki
    [J]. MEDICAL PHYSICS, 2019, 46 (02) : 964 - 972
  • [8] Intercomparison of calibration procedures for 192Ir HDR sources in Brazil
    de Almeida, CE
    Pereira, AJ
    Marechal, MH
    Pereira, G
    Cruz, JC
    Ferraz, JC
    Giordani, AJ
    Khalil, CM
    Martins, RH
    Menegussi, G
    Moreira, D
    Rocha, JR
    Pinto, MA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (03): : N31 - N38
  • [9] Positional and angular tracking of HDR 192Ir source for brachytherapy quality assurance using radiochromic film dosimetry
    Aldelaijan, Saad
    Devic, Slobodan
    Bekerat, Hamed
    Papaconstadopoulos, Pavlos
    Schneider, James
    Seuntjens, Jan
    Cormack, Robert A.
    Buzurovic, Ivan M.
    [J]. MEDICAL PHYSICS, 2020, 47 (12) : 6122 - 6139
  • [10] Suitability Assessment of CsPbI3 Dosimeter with 192Ir Source for Brachytherapy Quality Assurance
    Yang, S. W.
    Park, S. K.
    Moon, C. W.
    Kim, K. H.
    [J]. JOURNAL OF INSTRUMENTATION, 2023, 18 (07)