Organ motion in linac-based SBRT for glottic cancer

被引:6
|
作者
Perillo, Annarita [1 ]
Landoni, Valeria [2 ]
Farneti, Alessia [1 ]
Sanguineti, Giuseppe [1 ]
机构
[1] IRCSS Regina Elena Natl Canc Inst, Dept Radiat Oncol, Rome, Italy
[2] IRCSS Regina Elena Natl Canc Inst, Dept Med Phys, Via Elio Chianesi 53, I-00144 Rome, Italy
关键词
Organ motion; Daily CBCT imaging; SBRT; Glottic cancer; VOCAL CORD IRRADIATION; MANAGEMENT; SURGERY;
D O I
10.1186/s13014-021-01833-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate inter- and intra-fraction organ motion as well as to quantify clinical target volume (CTV) to planning target volume (PTV) margins to be adopted in the stereotactic treatment of early stage glottic cancer. Methods and materials: Stereotactic body radiotherapy (SBRT) to 36 Gy in 3 fractions was administered to 23 patients with early glottic cancer T1N0M0. Patients were irradiated with a volumetric intensity modulated arc technique delivered with 6 MV FFF energy. Each patient underwent a pre-treatment cone beam computed tomography (CBCT) to correct the setup based on the thyroid cartilage position. Imaging was repeated if displacement exceeded 2 mm in any direction. CBCT imaging was also performed after each treatment arc as well as at the end of the delivery. Swallowing was allowed only during the beam-off time between arcs. CBCT images were reviewed to evaluate inter- and intra-fraction organ motion. The relationships between selected treatment characteristics, both beam-on and delivery times as well as organ motion were investigated. Results: For the population systematic (Sigma) and random (sigma) inter-fraction errors were 0.9, 1.3 and 0.6 mm and 1.1, 1.3 and 0.7 mm in the left-right (X), cranio-caudal (Y) and antero-posterior (Z) directions, respectively. From the analysis of CBCT images acquired after treatment, systematic (Sigma) and random (sigma) intra-fraction errors resulted 0.7, 1.6 and 0.7 mm and 1.0, 1.5 and 0.6 mm in the X, Y and Z directions, respectively. Margins calculated from the intra-fraction errors were 2.4, 5.1 and 2.2 mm in the X, Y and Z directions respectively. A statistically significant difference was found for the displacement in the Z direction between patients irradiated with > 2 arcs versus <= 2 arcs, (MW test, p = 0.038). When analyzing mean data from CBCT images for the whole treatment, a significant correlation was found between the time of delivery and the three dimensional displacement vector (r = 0.489, p = 0.055), the displacement in the Y direction (r = 0.553, p = 0.026) and the subsequent margins to be adopted (r = 0.626, p = 0.009). Finally, displacements and the subsequent margins to be adopted in Y direction were significantly greater for treatments with more than 2 arcs (MW test p = 0.037 and p = 0.019, respectively). Conclusions: In the setting of controlled swallowing during treatment delivery, intra-fraction motion still needs to be taken into account when planning with estimated CTV to PTV margins of 3, 5 and 3 mm in the X, Y and Z directions, respectively. Selected treatments may require additional margins.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Organ motion in linac-based SBRT for glottic cancer
    Annarita Perillo
    Valeria Landoni
    Alessia Farneti
    Giuseppe Sanguineti
    Radiation Oncology, 16
  • [2] Dosimetric Impact of Intrafraction Prostate Motion in DoseEscalated Linac-Based SBRT
    Panizza, D.
    Faccenda, V.
    Lucchini, R.
    Daniotti, M. C.
    Trivellato, S.
    Caricato, P.
    Arcangeli, S.
    De Ponti, E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E575 - E575
  • [3] Intrafraction prostate motion management in dose-escalated linac-based SBRT
    Panizza, D.
    Lucchini, R.
    Faccenda, V.
    Daniotti, M. C.
    Caricato, P.
    Trivellato, S.
    Arcangeli, S.
    De Ponti, E.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1516 - S1517
  • [4] Intrafraction Prostate Motion Management during Dose-Escalated Linac-Based SBRT
    Panizza, D.
    Faccenda, V.
    Lucchini, R.
    Daniotti, M. C.
    Trivellato, S.
    Caricato, P.
    Arcangeli, S.
    De Ponti, E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E570 - E571
  • [5] Treatment Planning Comparison of CyberKnife and Linac-Based SBRT
    Yoon, J.
    Schwer, M.
    Sharma, S.
    Ju, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E299 - E299
  • [6] Assessing the dosimetric impact of intrafraction prostate motion in dose-escalated linac-based SBRT
    Faccenda, V.
    Panizza, D.
    Daniotti, M. C.
    Trivellato, S.
    Caricato, P.
    Lucchini, R.
    Arcangeli, S.
    De Ponti, E.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1515 - S1516
  • [7] Linac-Based SBRT Delivery QA and Decreased Gamma Criteria
    Price, R.
    Veltchev, I.
    Lin, T.
    Eldib, A.
    Chen, L.
    Chen, X.
    Jin, L.
    Wang, L.
    Liu, J.
    Ma, C.
    MEDICAL PHYSICS, 2020, 47 (06) : E810 - E810
  • [8] A Radiobiological Comparative Study of Robotic and LINAC-Based Lung SBRT
    Roth, T.
    Rankine, L.
    Schreiber, E.
    Das, S.
    Mavroidis, P.
    MEDICAL PHYSICS, 2016, 43 (06) : 3588 - 3589
  • [9] Real-time intrafraction prostate motion during dose-escalated linac-based SBRT
    Panizza, D.
    Trivellato, S.
    Montanari, G.
    Pisoni, V.
    Faccenda, V.
    Caricato, P.
    Daniotti, M. C.
    Lucchini, R.
    De Ponti, E.
    Arcangeli, S.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S690 - S691
  • [10] Linac-based SBRT as a feasible salvage option for local recurrences in previously irradiated prostate cancer
    Cuccia, Francesco
    Nicosia, Luca
    Mazzola, Rosario
    Figlia, Vanessa
    Giaj-Levra, Niccolo
    Ricchetti, Francesco
    Rigo, Michele
    Vitale, Claudio
    Corradini, Stefanie
    Ruggieri, Ruggero
    Alongi, Filippo
    STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (07) : 628 - 636