Quantitative Assessment of the Accuracy of Real-Time Virtual Sonography for Liver Surgery

被引:12
|
作者
Miyata, Akinori [1 ]
Arita, Junichi [1 ]
Shirata, Chikara [1 ]
Abe, Satoru [1 ]
Akamatsu, Nobuhisa [1 ]
Kaneko, Junichi [1 ]
Kokudo, Norihiro [1 ,2 ]
Hasegawa, Kiyoshi [1 ]
机构
[1] Univ Tokyo, Grad Sch Med, Dept Surg, Hepatobiliary Pancreat Surg Div, Tokyo, Japan
[2] Natl Ctr Global Hlth & Med, Dept Surg, Hepatobiliary Pancreat Surg Div, Tokyo, Japan
关键词
hepato-biliary; image-guided surgery; simulation; HEPATOCELLULAR-CARCINOMA; RADIOFREQUENCY ABLATION; AUGMENTED REALITY; NAVIGATION; RESECTION; FEASIBILITY; ULTRASOUND; BIOPSY; CT;
D O I
10.1177/1553350619875301
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. Real-time virtual sonography (RVS) is a navigation system for liver surgery. In this study, the degree of misalignment of intraoperative RVS images with computed tomographic (CT) images was measured. Methods. Between December 2014 and July 2015, intraoperative RVS was performed in a total of 33 patients undergoing liver surgery. Reconstructed CT images, rendered like intraoperative ultrasonographic (IOUS) images, were adjusted with the IOUS images and visualized side by side. The degree of misalignment between the reconstructed CT images and IOUS images was measured at anterior section, posterior section, and left liver in each patient. Furthermore, the time required for the adjustment was measured as the "adjustment time." Results. The degree of misalignment between the images could potentially be measured for a total of 96 points in the 33 patients. Of these, the actual measurement could not be conducted for 35 points due to poor visualization of the intrahepatic vasculature (n = 20) or to a large misalignment that hampered continuation of further adjustment (n = 15). The median degree of misalignment was 9.8 mm (range = 2.4-37.6 mm) in the right anterior section, 9.8 mm (range = 2.7-71.5 mm) in the right posterior section, and 9.5 mm (range = 0.9-37.6 mm) in the left liver. The median adjustment time was 105 seconds (range = 51-245 seconds). Conclusions. Although some misalignment occurred, it might be acceptable for selected situations. Further investigation is needed to reduce the frequency of adjustment failure.
引用
收藏
页码:60 / 67
页数:8
相关论文
共 50 条
  • [1] Real-time Contrast-enhanced and Real-time Virtual Sonography in the Assessment of Benign Liver Lesions
    Sandulescu, Larisa
    Saftoiu, Adrian
    Dumitrescu, Daniela
    Ciurea, Tudorel
    [J]. JOURNAL OF GASTROINTESTINAL AND LIVER DISEASES, 2008, 17 (04) : 475 - 478
  • [2] The Role of Real-time Contrast-Enhanced and Real-time Virtual Sonography in the Assessment of Malignant Liver Lesions
    Sandulescu, Larisa
    Saftoiu, Adrian
    Dumitrescu, Daniela
    Ciurea, Tudorel
    [J]. JOURNAL OF GASTROINTESTINAL AND LIVER DISEASES, 2009, 18 (01) : 103 - 108
  • [3] REAL-TIME SONOGRAPHY DURING SPINAL SURGERY
    KNAKE, JE
    GABRIELSEN, TO
    CHANDLER, WF
    LATACK, JT
    GEBARSKI, SS
    YANG, PJ
    [J]. RADIOLOGY, 1984, 151 (02) : 461 - 465
  • [4] ACCURACY OF REAL-TIME SONOGRAPHY IN DETERMINATION OF FETAL VIABILITY
    LEVINE, SC
    FILLY, RA
    [J]. OBSTETRICS AND GYNECOLOGY, 1977, 49 (04): : 475 - 477
  • [5] Real-Time Virtual Sonography For The Peripheral Lesion Of The Lung
    Numanami, H.
    Yamaji, M.
    Tanaka, M.
    Koyama, S.
    Yamaguchi, E.
    Haniuda, M.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [6] Incorporating PET/CT into the novel real-time virtual sonography
    Hung, Guang-Uei
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [7] MRI, US or real-time virtual sonography in the evaluation of adenomyosis?
    Vinci, Valeria
    Saldari, Matteo
    Sergi, Maria Eleonora
    Bernardo, Silvia
    Rizzo, Giuseppe
    Porpora, Maria Grazia
    Catalano, Carlo
    Manganaro, Lucia
    [J]. RADIOLOGIA MEDICA, 2017, 122 (05): : 361 - 368
  • [8] MRI, US or real-time virtual sonography in the evaluation of adenomyosis?
    Valeria Vinci
    Matteo Saldari
    Maria Eleonora Sergi
    Silvia Bernardo
    Giuseppe Rizzo
    Maria Grazia Porpora
    Carlo Catalano
    Lucia Manganaro
    [J]. La radiologia medica, 2017, 122 : 361 - 368
  • [9] Accuracy of quantitative real-time PCR analysis
    Sochivko, D. G.
    Fedorov, A. A.
    Varlamov, D. A.
    Kurochkin, V. E.
    Petrov, R. V.
    [J]. DOKLADY BIOCHEMISTRY AND BIOPHYSICS, 2013, 449 (01) : 105 - 108
  • [10] Accuracy of quantitative real-time PCR analysis
    D. G. Sochivko
    A. A. Fedorov
    D. A. Varlamov
    V. E. Kurochkin
    R. V. Petrov
    [J]. Doklady Biochemistry and Biophysics, 2013, 449 : 105 - 108