Quality control of mechanical ventilation at the patient's home

被引:13
|
作者
Farré, R
Giró, E
Casolivé, V
Navajas, D
Escarrabill, J
机构
[1] Univ Barcelona, Inst Invest Biomed August Pi Sunyer, Fac Med, Unitat Biofis & Bioengn, E-08036 Barcelona, Spain
[2] Univ Barcelona, Bellvitge Hosp, UFISS Resp, Serv Pneumol, Barcelona, Spain
关键词
home mechanical ventilation; chronic respiratory failure; ventilator performance;
D O I
10.1007/s00134-002-1593-y
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: During home mechanical ventilation the prescribed settings are applied without permanent supervision of health professionals. After a long-time period of unattended operation at home the ventilator may not apply the ventilation parameters prescribed. This quality control study of home mechanical ventilation assessed whether tidal volume (V-T), frequency (f), and minute ventilation (V'(E)) actually applied by the ventilator coincide with the values set on the ventilator control panel and with those prescribed. Measurements: Actual V-T, f, and V'(E) applied by the ventilator in 30 patients on nocturnal HMV were measured at the patients' homes. The patients were subjected to volume targeted assist ventilation through nasal mask (n = 28) or tracheostomy ( n = 2). The values of V-T, f, and V'(E) set at the ventilator were recorded. The actual and set V-T, f, and V'(E) values were compared with those prescribed. Results: Considerable differences were found between actual, set and prescribed V-T, f, and V'(E). Actual VE was significantly lower than VE set: mean difference was 0.82 l/min, with considerable individual differences. Differences between actual and prescribed VE were caused both by a poor performance of the ventilator and by a discrepancy between the values prescribed and those set at the ventilator control panel. Conclusions: Regularly assessing the actual performance of ventilators at the patient's home is a quality control procedure useful for detecting malfunctions which could improve compliance and outcome of home mechanical ventilation.
引用
收藏
页码:484 / 486
页数:3
相关论文
共 50 条
  • [1] Quality control of mechanical ventilation at the patient's home
    Ramon Farré
    Esther Giró
    Vinyet Casolivé
    Daniel Navajas
    Joan Escarrabill
    [J]. Intensive Care Medicine, 2003, 29 : 484 - 486
  • [2] Mechanical home ventilation in Duchenne's disease - Control of quality of life.
    Kirisits, J
    Reisecker, F
    [J]. NEUROPSYCHIATRIE, 1995, 9 (04): : 192 - 195
  • [3] Quality control of equipment in home mechanical ventilation: a European survey
    Farre, R
    Lloyd-Owen, SJ
    Ambrosino, N
    Donaldson, G
    Escarrabill, J
    Fauroux, B
    Robert, D
    Schoenhofer, B
    Simonds, A
    Wedzicha, JA
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2005, 26 (01) : 86 - 94
  • [4] Home Mechanical Ventilation: A Patient's Perspective Survey Study
    Kwiatosz-Muc, Magdalena
    Kopacz, Bozena
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (08)
  • [5] Performance of mechanical ventilators at the patient's home: a multicentre quality control study
    Farre, R
    Navajas, D
    Prats, E
    Marti, S
    Guell, R
    Montserrat, JM
    Tebe, C
    Escarrabill, J
    [J]. THORAX, 2006, 61 (05) : 400 - 404
  • [6] Quality of Care in Home Mechanical Ventilation
    Klingshirn, Hanna
    Schwarz, Sarah B.
    [J]. PNEUMOLOGIE, 2022, 76 (06): : 397 - 403
  • [7] PATIENT SELECTION AND TECHNIQUES FOR HOME MECHANICAL VENTILATION
    DONNER, CF
    HOWARD, P
    ROBERT, D
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1993, 6 (01) : 3 - 4
  • [8] Quality of life with home mechanical ventilation.
    Windisch, W
    Petermann, F
    LaierGroeneveld, G
    Fischer, S
    Criee, CP
    [J]. MEDIZINISCHE KLINIK, 1997, 92 (01) : 95 - 100
  • [9] Quality of life under home mechanical ventilation
    Hein, H
    Schucher, B
    Magnussen, H
    [J]. MEDIZINISCHE KLINIK, 1999, 94 : 99 - 101
  • [10] Impact of home mechanical ventilation on sleep quality
    Schwarz, Sarah B.
    Magnet, Friederike S.
    Windisch, Wolfram
    [J]. CURRENT OPINION IN PULMONARY MEDICINE, 2017, 23 (06) : 500 - 505