Adaptive planning based on single beam optimization in passive scattering carbon ion radiotherapy for patients with pancreatic cancer

被引:4
|
作者
Li, Yang [1 ,2 ]
Kubota, Yoshiki [3 ]
Okamoto, Masahiko [3 ]
Shiba, Shintaro [3 ]
Okazaki, Shohei [3 ]
Matsui, Toshiaki [1 ]
Tashiro, Mutsumi [3 ]
Nakano, Takashi [3 ]
Ohno, Tatsuya [3 ]
机构
[1] Gunma Univ, Grad Sch Med, Maebashi, Gumma, Japan
[2] Harbin Med Univ, Dept Radiat Oncol, Canc Hosp, Harbin, Peoples R China
[3] Gunma Univ, Heavy Ion Med Ctr, Maebashi, Gumma, Japan
关键词
Carbon-ion radiotherapy; Adaptive planning; Pancreatic cancer; Accumulated dose assessment; Robustness of treatment; ANATOMICAL CHANGES; RADIATION-THERAPY; IMPACT; IRRADIATION; ROBUSTNESS; STRATEGIES; GUIDANCE; RISK;
D O I
10.1186/s13014-021-01841-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Daily anatomical deviations may distort the dose distribution in carbon ion radiotherapy (CIRT), which may cause treatment failure. Therefore, this study aimed to perform re-planning to maintain the dose coverage in patients with pancreatic cancer with passive scattering CIRT. Methods Eight patients with pancreatic cancer and 95 daily computed tomography (CT) sets were examined. Two types of adaptive plans based on new range compensators (RCs) (AP-1) and initial RCs (AP-2) were generated. In AP-2, each beam was optimized by manually adjusting the range shifter thickness and spread-out Bragg peak size to make dose reduction by < 3% of the original plan. Doses of the original plan with bone matching (BM) and tumor matching (TM) were examined for comparison. We calculated the accumulated dose using the contour and intensity-based deformable image registration algorithm. The dosimetric differences in respect to the original plan were compared between methods. Results Using TM and BM, mean +/- standard deviations of daily CTV V95 (%) difference from the original plan was - 5.1 +/- 6.2 and - 8.8 +/- 8.8, respectively, but 1.2 +/- 3.4 in AP-1 and - 0.5 +/- 2.1 in AP-2 (P < 0.001). AP-1 and AP-2 enabled to maintain a satisfactory accumulated dose in all patients. The dose difference was 1.2 +/- 2.8, - 2,1 +/- 1.7, - 7.1 +/- 5.2, and - 16.5 +/- 15.0 for AP-1, AP-2, TM, and BM, respectively. However, AP-2 caused a dose increase in the duodenum, especially in the left-right beam. Conclusions The possible dose deterioration should be considered when performing the BM, even TM. Re-planning based on single beam optimization in passive scattering CIRT seems an effective and safe method of ensuring the treatment robustness in pancreatic cancer. Further study is necessary to spare healthy tissues, especially the duodenum.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Adaptive planning based on single beam optimization in passive scattering carbon ion radiotherapy for patients with pancreatic cancer
    Yang Li
    Yoshiki Kubota
    Masahiko Okamoto
    Shintaro Shiba
    Shohei Okazaki
    Toshiaki Matsui
    Mutsumi Tashiro
    Takashi Nakano
    Tatsuya Ohno
    [J]. Radiation Oncology, 16
  • [2] Robust Beam Selection Based on Water Equivalent Thickness Analysis in Passive Scattering Carbon-Ion Radiotherapy for Pancreatic Cancer
    Zhou, Yuan
    Sakai, Makoto
    Li, Yang
    Kubota, Yoshiki
    Okamoto, Masahiko
    Shiba, Shintaro
    Okazaki, Shohei
    Matsui, Toshiaki
    Ohno, Tatsuya
    [J]. CANCERS, 2023, 15 (09)
  • [3] An adaptive planning strategy in carbon ion therapy of pancreatic cancer involving beam angle selection
    Kawashima, Motohiro
    Tashiro, Mutsumi
    Varnava, Maria
    Shiba, Shintaro
    Matsui, Toshiaki
    Okazaki, Shohei
    Li, Yang
    Komatsu, Shuichiro
    Kawamura, Hidemasa
    Okamoto, Masahiko
    Ohno, Tatsuya
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 21 : 35 - 41
  • [4] The role of multiple anatomical scenarios in plan optimization for carbon ion radiotherapy of pancreatic cancer
    Molinelli, Silvia
    Vai, Alessandro
    Russo, Stefania
    Loap, Pierre
    Meschini, Giorgia
    Paganelli, Chiara
    Barcellini, Amelia
    Vitolo, Viviana
    Orlandi, Ester
    Ciocca, Mario
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 176 : 1 - 8
  • [5] Adaptive radiotherapy based on the daily regression of a tumor in carbon-ion beam irradiation
    Nagano, Ai
    Minohara, Shinichi
    Kato, Shingo
    Kiyohara, Hiroki
    Ando, Ken
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (24):
  • [6] Carbon Ion Radiotherapy in the Treatment of Pancreatic Cancer A Review
    Malouff, Timothy D.
    Krishnan, Sunil
    Hallemeier, Christopher L.
    Haddock, Michael G.
    Hoppe, Bradford S.
    Beltran, Chris
    Mahajan, Anita
    Trifiletti, Daniel M.
    [J]. PANCREAS, 2020, 49 (06) : 737 - 743
  • [7] Deterioration of pancreatic exocrine function in carbon ion radiotherapy for pancreatic cancer
    Shiba, Shintaro
    Miyasaka, Yuhei
    Okamoto, Masahiko
    Komatsu, Shuichiro
    Okazaki, Shohei
    Shibuya, Kei
    Ohno, Tatsuya
    [J]. CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2021, 31 : 80 - 85
  • [8] A single institutional experience of carbon-ion radiotherapy for locally advanced pancreatic cancer.
    Shinoto, Makoto
    Shioyama, Yoshiyuki
    Suefuji, Hiroaki
    Toyama, Shinto
    Matsumoto, Keiji
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2017, 35 (04)
  • [9] Carbon ion radiotherapy in pancreatic cancer: A review of clinical data
    Liermann, Jakob
    Shinoto, Makoto
    Syed, Mustafa
    Debus, Juergen
    Herfarth, Klaus
    Naumann, Patrick
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 147 : 145 - 150
  • [10] The need for treatment adaptation in carbon ion radiotherapy of pancreatic cancer
    Molinelli, S.
    Vai, A.
    Russo, S.
    Loap, P.
    Meschini, G.
    Magro, G.
    Paganelli, C.
    Barcellini, A.
    Vitolo, V.
    Ciocca, M.
    Orlandi, E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S566 - S567