Screen-printed amperometric microcell for proline iminopeptidase enzyme activity assay

被引:10
|
作者
Nagy, G
Gyurcsányi, RE
Cristalli, A
Neuman, MR
Lindner, E [1 ]
机构
[1] Univ Memphis, Joint Grad Program Biomed Engn, Memphis, TN 38152 USA
[2] Univ Tennessee, Ctr Hlth Sci, Herff Coll Engn, Memphis, TN 38152 USA
[3] Univ Pecs, Dept Gen & Phys Chem, H-7624 Pecs, Hungary
[4] Tech Univ Budapest, Hungarian Acad Sci, Dept Analyt Chem, Res Grp Tech Analyt Chem, H-1111 Budapest, Hungary
来源
BIOSENSORS & BIOELECTRONICS | 2000年 / 15卷 / 5-6期
基金
匈牙利科学研究基金会;
关键词
proline iminopeptidase activity; screen-printed graphite electrode; microfabricated amperometric cell; enzyme assay;
D O I
10.1016/S0956-5663(00)00081-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A microfabricated amperometric microcell was designed and used for the determination of proline iminopeptidase (PIP) enzyme activity in 2-10-mu l samples. The measurements were made in the range of 10.3-841.5 mU/ml enzyme activities. The sensitivity of the determinations was between -0.0195 and -0.0203 mu A ml/mU per min. The coefficient of variation of the determined values ranged between 2.8 (at 561.2 mU/ml) and 24.1% (at 10.3 mU/ml). The microcell was manufactured on an alumina substrate using screen-printed graphite working and Ag/AgCl reference electrodes. Elevated PIP activity in the vaginal fluid is a biochemical indicator of bacterial vaginosis. The method is appropriate to differentiate between normal (66 +/- 145 mU/ml) and elevated, diseased (704 +/- 145 mU/ml), values. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:265 / 272
页数:8
相关论文
共 50 条
  • [1] Amperometric determination of phosphoglucomutase activity with a bienzyme screen-printed biosensor
    Cui, Y
    Barford, JP
    Renneberg, R
    ANALYTICAL BIOCHEMISTRY, 2006, 354 (01) : 162 - 164
  • [2] Amperometric microcell for enzyme activity measurements
    Nagy, G
    Xu, CX
    Buck, RP
    Lindner, E
    Neuman, MR
    ANALYTICAL CHEMISTRY, 1998, 70 (10) : 2156 - 2162
  • [3] Amperometric determination of tyrosinase catecholase activity using a screen-printed electrode
    Cui, Yue
    Barford, John P.
    Renneberg, Reinhard
    CHEMIA ANALITYCZNA, 2007, 52 (01): : 5 - 14
  • [4] Screen-Printed Electrodes for Amperometric Determination of Iodide
    Khodari, M.
    Bilitewski, U.
    Basry, A. A. H.
    ELECTROANALYSIS, 2015, 27 (02) : 281 - 284
  • [5] Amperometric bienzyme screen-printed biosensor for the determination of leucine
    Labroo, Pratima
    Cui, Yue
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (01) : 367 - 372
  • [6] Amperometric bienzyme screen-printed biosensor for the determination of leucine
    Pratima Labroo
    Yue Cui
    Analytical and Bioanalytical Chemistry, 2014, 406 : 367 - 372
  • [7] AMPEROMETRIC DETERMINATION OF COPPER USING SCREEN-PRINTED ELECTRODES
    SOMASUNDRUM, M
    BANNISTER, JV
    SENSORS AND ACTUATORS B-CHEMICAL, 1993, 15 (1-3) : 203 - 208
  • [8] Amperometric determination of copper using screen-printed electrodes
    Somasundrum, Mithran
    Bannister, Joe V.
    Sensors and Actuators, B: Chemical, 1993, B15 (1 -3 pt 1) : 203 - 208
  • [9] Application of an ELISA-type amperometric assay to the detection of Vibrio species with screen-printed electrodes
    Laczka, Olivier
    Labbate, Maurizio
    Doblin, Martina
    ANALYTICAL METHODS, 2014, 6 (07) : 2020 - 2023
  • [10] Immobilization of enzyme on screen-printed electrode by exposure to glutaraldehyde vapour for the construction of amperometric acetylcholinesterase electrodes
    Li, YG
    Zhou, YX
    Feng, JL
    Jiang, ZH
    Ma, LR
    ANALYTICA CHIMICA ACTA, 1999, 382 (03) : 277 - 282