Evaluation of z-axis resolution and image noise for nonconstant velocity spiral CT data reconstructed using a weighted 3D filtered backprojection (WFBP) reconstruction algorithm

被引:10
|
作者
Christner, Jodie A. [1 ]
Stierstorfer, Karl [2 ]
Primak, Andrew N. [1 ]
Eusemann, Christian D. [3 ]
Flohr, Thomas G. [2 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
[2] Siemens Healthcare Forchheim, D-91301 Forchheim, Germany
[3] Siemens Med Solut USA, Malvern, PA 19355 USA
关键词
computerised tomography; image reconstruction; medical image processing; phantoms; COMPUTED-TOMOGRAPHY; QUALITY EVALUATION; VARIABLE PITCH; PERFUSION; SCANNER;
D O I
10.1118/1.3271110
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A WFBP reconstruction algorithm was used to reconstruct stationary (axial, pitch=0), constant velocity spiral (pitch=0.35-1.5) and nonconstant velocity spiral CT data acquired using a 128x0.6 mm acquisition mode (38.4 mm total detector length, z-flying focal spot technique), and a gantry rotation time of 0.30 s. Nonconstant velocity scans used the system's periodic spiral mode, where the table moved in and out of the gantry in a cyclical manner. For all scan types, the volume CTDI was 10 mGy. Measurements of CT number, image noise, and the slice sensitivity profile were made for all scan types as a function of the nominal slice width, table velocity, and position within the scan field of view. A thorax phantom was scanned using all modes and reconstructed transverse and coronal plane images were compared. Results: Negligible differences in slice thickness, CT number, noise, or artifacts were found between scan modes for data taken at two positions within the scan field of view. For nominal slices of 1.0-3.0 mm, FWHM values of the slice sensitivity profiles were essentially independent of the scan type. For periodic spiral scans, FWHM values measured at the center of the scan range were indistinguishable from those taken 5 mm from one end of the scan range. All CT numbers were within +/- 5 HU, and CT number and noise values were similar for all scan modes assessed. A slight increase in noise and artifact level was observed 5 mm from the start of the scan on the first pass of the periodic spiral. On subsequent passes, noise and artifact level in the transverse and coronal plane images were the same for all scan modes. Conclusions: Nonconstant velocity periodic spiral scans can achieve z-axis spatial resolution, CT number accuracy, image noise and artifact level equivalent to those for stationary (axial), and constant velocity spiral scans. Thus, periodic spiral scans are expected to allow assessment of four-dimensional CT data for scan lengths greater than the detector width without sacrificing image quality.
引用
收藏
页码:897 / 906
页数:10
相关论文
共 4 条
  • [1] Image reconstruction using Z-axis spatio-temporal sampling in 3D electrical capacitance tomography
    Wanta, Damian
    Smolik, Waldemar T.
    Kryszyn, Jacek
    Midura, Mateusz
    Wroblewski, Przemyslaw
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2022, 33 (11)
  • [2] 3D image reconstruction of CT-data using the ITEM algorithm
    Durst, J
    Pauli, J
    Anton, G
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1008 - 1012
  • [3] 3D-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction at low-helical pitches to improve noise characteristics and dose efficiency
    Tang, Xiangyang
    Hsieh, Jiang
    Nilsen, Roy A.
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [4] Computation-efficient cone beam image reconstruction for image-guided radiation therapy applications using 3D weighted filtered backfirojection (CB-FBP) algorithm
    Tang, X.
    Hsieh, J.
    Dong, F.
    MEDICAL PHYSICS, 2006, 33 (06) : 1997 - 1997