Ion-optical studies for a range adaptation method in ion beam therapy using a static wedge degrader combined with magnetic beam deflection

被引:8
|
作者
Chaudhri, Naved [1 ]
Saito, Nami [1 ]
Bert, Christoph [1 ]
Franczak, Bernhard [1 ]
Steidl, Peter [1 ]
Durante, Marco [1 ,2 ]
Rietzel, Eike [3 ]
Schardt, Dieter [1 ]
机构
[1] GSI Helmholtz Ctr Heavy Ion Res GmbH, D-64291 Darmstadt, Germany
[2] Tech Univ Darmstadt, Inst Festkorperphys, D-64289 Darmstadt, Germany
[3] Siemens Healthcare Workflow & Solut, Particle Therapy, D-91052 Erlangen, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 12期
关键词
TRACKING SYSTEM; MOTION; ACCURACY;
D O I
10.1088/0031-9155/55/12/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fast radiological range adaptation of the ion beam is essential when target motion is mitigated by beam tracking using scanned ion beams for dose delivery. Electromagnetically controlled deflection of a well-focused ion beam on a small static wedge degrader positioned between two dipole magnets, inside the beam delivery system, has been considered as a fast range adaptation method. The principle of the range adaptation method was tested in experiments and Monte Carlo simulations for the therapy beam line at the GSI Helmholtz Centre for Heavy Ions Research. Based on the simulations, ion optical settings of beam deflection and realignment of the adapted beam were experimentally applied to the beam line, and additional tuning was manually performed. Different degrader shapes were employed for the energy adaptation. Measured and simulated beam profiles, i.e. lateral distribution and range in water at isocentre, were analysed and compared with the therapy beam values for beam scanning. Deflected beam positions of up to +/- 28 mm on degrader were performed which resulted in a range adaptation of up to +/- 15 mm water equivalence (WE). The maximum deviation between the measured adapted range from the nominal range adaptation was below 0.4 mm WE. In experiments, the width of the adapted beam at the isocentre was adjustable between 5 and 11 mm full width at half maximum. The results demonstrate the feasibility/proof of the proposed range adaptation method for beam tracking from the beam quality point of view.
引用
下载
收藏
页码:3499 / 3513
页数:15
相关论文
共 26 条
  • [1] METHOD OF MATCHING EXTERNAL BEAM OF A CYCLOTRON TO AN ION-OPTICAL AXIS
    SCHUTTE, F
    EHRNREICH, KR
    VANHEUSD.GC
    NUCLEAR INSTRUMENTS & METHODS, 1971, 97 (02): : 347 - +
  • [2] Low energy nitrogen ion doping into GaAs using combined ion-beam and molecular-beam epitaxy method
    Shima, T
    Makita, Y
    Kimura, S
    Iida, T
    Fang, XH
    Jiang, DS
    Kudo, K
    Tanaka, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1996, 120 (1-4): : 293 - 297
  • [3] On the Ion Beam Range and Dose Verification in Hadron Therapy Using Sound Waves
    Fourkal, E.
    Veltchev, I.
    Gayou, O.
    Nahirnyak, V.
    MEDICAL PHYSICS, 2015, 42 (06) : 3296 - 3296
  • [4] CARBON DOPING INTO GAAS USING COMBINED ION-BEAM AND MOLECULAR-BEAM EPITAXY METHOD
    IIDA, T
    MAKITA, Y
    KIMURA, S
    KAWASUMI, Y
    YAMADA, A
    UEKUSA, S
    TSUKAMOTO, T
    JOURNAL OF CRYSTAL GROWTH, 1995, 150 (1-4) : 236 - 240
  • [5] Compensation of beam deflection due to the magnetic field using beam steering by aperture displacement technique in the multibeamlet negative ion source
    Hamabe, M
    Takeiri, Y
    Ikeda, K
    Oka, Y
    Osakabe, M
    Tsumori, K
    Asano, E
    Kawamoto, T
    Kaneko, O
    Tanaka, M
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (08): : 3237 - 3244
  • [6] Combined figuring technology for high-precision optical surfaces using a deterministic ion beam material adding and removal method
    Liao, Wenlin
    Dai, Yifan
    Xie, Xuhui
    Zhou, Lin
    OPTICAL ENGINEERING, 2013, 52 (01)
  • [7] Studies of magnetic dipole transitions in highly charged argon and barium using an electron beam ion trap
    Bieber, D.J.
    Margolis, H.S.
    Oxley, P.K.
    Silver, J.D.
    Physica Scripta T, 1997, T73 : 64 - 66
  • [8] Studies of magnetic dipole transitions in highly charged argon and barium using an electron beam ion trap
    Bieber, DJ
    Margolis, HS
    Oxley, PK
    Silver, JD
    PHYSICA SCRIPTA, 1997, T73 : 64 - 66
  • [9] Influence of ion beam on optical properties of GaAs: Calculation using the fractional-derivative-spectrum method
    Kulik, Miroslaw
    Rzodkiewicz, Witold
    Drozdziel, Andrzej
    Pyszniak, Krzysztof
    PHYSICA STATUS SOLIDI C: CURRENT TOPICS IN SOLID STATE PHYSICS, VOL 8, NO 4, 2011, 8 (04): : 1315 - 1318
  • [10] Scanning near-field optical microscopy in the near-infrared spectral range using ion beam treated silicon probes
    Danzebrink, HU
    Dziomba, T
    Lehrer, C
    Frey, L
    Sulzbach, T
    Ohlsson, O
    PRECISION ENGINEERING, NANOTECHNOLOGY, VOL. 2, 1999, : 168 - 171