Estimation of organ absorbed dose in pediatric chest X-ray examination: A phantom study

被引:7
|
作者
Jamal, Nurul H. M. [1 ]
Sayed, Inayatullah S. [1 ]
Syed, Waliullah S. [2 ]
机构
[1] Int Islamic Univ Malaysia, Dept Diagnost Imaging & Radiotherapy, Kulliyyah Allied Hlth Sci, Jalan Sultan Ahmad Shah, Kuantan 25200, Pahang, Malaysia
[2] Univ Sains Malaysia, Sch Hlth Sci, Hlth Campus, Kota Baharu 16150, Kelantan, Malaysia
关键词
Absorbed dose; Chest X-ray; Pediatrics; Phantom; OSLD; TISSUE-EQUIVALENT MATERIALS; BEESWAX; CONSTRUCTION;
D O I
10.1016/j.radphyschem.2019.108472
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Children have a greater risk of developing lifetime cancer and other biological effects from ionizing radiation exposure than adults. The aim of this study was to measure the absorbed dose received by lungs and heart in pediatric chest X-ray examination using nanoDot optically stimulated luminescent dosimeter (OSLD). The X-ray system, Siemens Multix Top was used. A pediatric phantom developed by using beeswax and polyurethane foam was exposed at 50 kVp, 52 kVp, 55 kVp, 57 kVp and 60 kVp, with fixed tube current-exposure time (3 mAs), which is normally used in pediatric clinical chest X-ray examinations. The nanoDot OSLDs were placed in different parts in the thorax of the phantom according to the position of organs in the chest area, which are lungs and heart. For lungs, absorbed dose measurement nanoDot OSLDs were placed in the apex and base at three different depths. The phantom was exposed three times for each kVp value, and the absorbed doses were measured in mGy. The findings show that the measured absorbed dose to the heart increased with the increase in kVp. Overall, a 22% increase in absorbed dose to heart and a 29% increase in lungs with the increase in kVp was recorded. In addition, absorbed dose to the base of left and right lungs was recorded higher up to 9% as compared to the apex of lungs. In conclusion, the absorbed dosage increases with exposure, while the absorbed dose decreases with depth. It is necessary for the radiographer to select an appropriate exposure setting based on the physical characteristics of the pediatric patient.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study
    Pengpan, Thanyawee
    Rattanarungruangchai, Natch
    Dechjaithat, Juthathip
    Panthim, Phawinee
    Siricharuwong, Puntarika
    Prapan, Ausanai
    [J]. RADIOLOGY RESEARCH AND PRACTICE, 2022, 2022
  • [2] The influence of calculated CT X-ray spectra on organ absorbed dose to a human phantom
    Ferreira, C. C.
    Vieira, J. W.
    Maia, A. F.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2009, 267 (20): : 3447 - 3455
  • [3] An absorbed dose due to examination on the VolumeScope X-ray device
    Vasiliev, VN
    Zaytsev, MY
    [J]. X-Ray and Neutron Capillary Optics II, 2005, 5943 : 242 - 247
  • [4] Fast method for the estimation of the absorbed dose in X-ray microtomography
    De Muynck, Amelie
    Bonte, Stijn
    Dhaene, Jelle
    Dierick, Manuel
    Bacher, Klaus
    Van Hoorebeke, Luc
    Boone, Matthieu N.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2019, 452 : 40 - 47
  • [5] Measurement of radiotherapy x-ray skin dose on a chest wall phantom
    Quach, KY
    Morales, J
    Butson, MJ
    Rosenfeld, AB
    Metcalfe, PE
    [J]. MEDICAL PHYSICS, 2000, 27 (07) : 1676 - 1680
  • [6] Assessment of organ absorbed doses in patients undergoing chest X-ray examinations by Monte Carlo based softwares and phantom dosimetry
    Campos de Oliveira, Paulo Marcio
    Squair, Peterson L.
    Lacerda, Marco A.
    da Silva, Teogenes A.
    [J]. RADIATION MEASUREMENTS, 2011, 46 (12) : 2073 - 2076
  • [7] Chest X-ray examination
    Hasan, S. Shakir
    [J]. ANAESTHESIA PAIN & INTENSIVE CARE, 2021, 25 (04)
  • [8] MONTE CARLO STUDY OF THE CARDIAC ABSORBED DOSE DURING X-RAY EXAMINATION OF AN ADULT PATIENT
    Kadri, O.
    Manai, K.
    Alfuraih, A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 171 (04) : 431 - 437
  • [9] Effects of tube potential and scatter rejection on image quality and effective dose in digital chest X-ray examination: An anthropomorphic phantom study
    Shaw, D. J.
    Crawshaw, I.
    Rimmer, S. D.
    [J]. RADIOGRAPHY, 2013, 19 (04) : 321 - 325
  • [10] X-ray technique in examination of the chest
    Richards, GE
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 1924, 14 : 1077 - 1080