Evaluation of methodologies for computing the deep brain stimulation volume of tissue activated

被引:53
|
作者
Duffley, Gordon [1 ,2 ]
Anderson, Daria Nesterovich [1 ,2 ,3 ]
Vorwerk, Johannes [2 ,6 ]
Dorval, Alan D. [1 ]
Butson, Christopher R. [1 ,2 ,3 ,4 ,5 ]
机构
[1] Univ Utah, Dept Biomed Engn, Salt Lake City, UT 84112 USA
[2] Univ Utah, SCI Inst, Salt Lake City, UT 84112 USA
[3] Univ Utah, Dept Neurosurg, Salt Lake City, UT 84112 USA
[4] Univ Utah, Dept Psychiat, Salt Lake City, UT 84112 USA
[5] Univ Utah, Dept Neurol, Salt Lake City, UT 84112 USA
[6] UMIT Private Univ Hlth Sci Med Informat & Technol, Inst Elect & Biomed Engn, Hall In Tirol, Austria
基金
美国国家科学基金会;
关键词
Hessian; fiber orientation; activating function; deep brain stimuation; volume of tissue activated; axon orientation; PATIENT-SPECIFIC MODELS; GLOBUS-PALLIDUS; CONNECTIVITY; METAANALYSIS; EFFICACY; NUCLEUS; SAFETY; TRIAL;
D O I
10.1088/1741-2552/ab3c95
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Computational models are a popular tool for predicting the effects of deep brain stimulation (DBS) on neural tissue. One commonly used model, the volume of tissue activated (VTA), is computed using multiple methodologies. We quantified differences in the VTAs generated by five methodologies: the traditional axon model method, the electric field norm, and three activating function based approaches?the activating function at each grid point in the tangential direction (AF-Tan) or in the maximally activating direction (AF-3D), and the maximum activating function along the entire length of a tangential fiber (AF-Max). Approach. We computed the VTA using each method across multiple stimulation settings. The resulting volumes were compared for similarity, and the methodologies were analyzed for their differences in behavior. Main results. Activation threshold values for both the electric field norm and the activating function varied with regards to electrode configuration, pulse width, and frequency. All methods produced highly similar volumes for monopolar stimulation. For bipolar electrode configurations, only the maximum activating function along the tangential axon method, AF-Max, produced similar volumes to those produced by the axon model method. Further analysis revealed that both of these methods are biased by their exclusive use of tangential fiber orientations. In contrast, the activating function in the maximally activating direction method, AF-3D, produces a VTA that is free of axon orientation and projection bias. Significance. Simulating tangentially oriented axons, the standard approach of computing the VTA, is too computationally expensive for widespread implementation and yields results biased by the assumption of tangential fiber orientation. In this work, we show that a computationally efficient method based on the activating function, AF-Max, reliably reproduces the VTAs generated by direct axon modeling. Further, we propose another method, AF-3D as a potentially superior model for representing generic neural tissue activation.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Linking volume of tissue activated to neural oscillations in Deep Brain Stimulation
    Kutuzova, A.
    Graef, C.
    Lonergan, B.
    Tai, Y.
    Haar, S.
    MOVEMENT DISORDERS, 2023, 38 : S541 - S542
  • [2] Current steering to control the volume of tissue activated during deep brain stimulation
    Butson, Christopher R.
    McIntyre, Cameron C.
    BRAIN STIMULATION, 2008, 1 (01) : 7 - 15
  • [3] EFFECT OF REALISTIC MODELING OF DEEP BRAIN STIMULATION ON THE PREDICTION OF VOLUME OF ACTIVATED TISSUE
    Golestanirad, L.
    Izquierdo, A. P.
    Graham, S. J.
    Mosig, J. R.
    Pollo, C.
    PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2012, 126 : 1 - 16
  • [4] Role of electrode design on the volume of tissue activated during deep brain stimulation
    Butson, Christopher R.
    McIntyre, Cameron C.
    JOURNAL OF NEURAL ENGINEERING, 2006, 3 (01) : 1 - 8
  • [5] Visualization of volume of tissue activated modeling in a clinical planning system for deep brain stimulation
    Carl, Barbara
    Bopp, Miriam
    Sass, Benjamin
    Waldthaler, Josefine
    Timmermann, Lars
    Nimsky, Christopher
    JOURNAL OF NEUROSURGICAL SCIENCES, 2024, 68 (01) : 59 - 69
  • [6] Patient-speciftic analysis of the volume of tissue activated during deep brain stimulation
    Butson, Christopher R.
    Cooper, Scott E.
    Henderson, Jaimie M.
    McIntyre, Cameron C.
    NEUROIMAGE, 2007, 34 (02) : 661 - 670
  • [7] Modeling the volume of tissue activated in deep brain stimulation and its clinical influence: a review
    Patrick, Erin E.
    Fleeting, Chance R.
    Patel, Drashti R.
    Casauay, Jed T.
    Patel, Aashay
    Shepherd, Hunter
    Wong, Joshua K.
    FRONTIERS IN HUMAN NEUROSCIENCE, 2024, 18
  • [8] A Gaussian Process Emulator for Estimating the Volume of Tissue Activated During Deep Brain Stimulation
    De La Pava, Ivan
    Gomez, Viviana
    Alvarez, Mauricio A.
    Henao, Oscar A.
    Daza-Santacoloma, Genaro
    Orozco, Alvaro A.
    PATTERN RECOGNITION AND IMAGE ANALYSIS (IBPRIA 2015), 2015, 9117 : 691 - 699
  • [9] Estimation of the neuromodulation parameters from the planned volume of tissue activated in deep brain stimulation
    Gomez-Orozco, Viviana
    Alexander Alvarez-Lopez, Mauricio
    Alberto Henao-Gallo, Oscar
    Daza-Santacoloma, Genaro
    Angel Orozco-Gutierrez, Alvaro
    REVISTA FACULTAD DE INGENIERIA-UNIVERSIDAD DE ANTIOQUIA, 2016, (79): : 9 - 18
  • [10] Visualization of volume of tissue activated modeling in a clinical planning system for deep brain stimulation
    Carl, Barbara
    Bopp, Miriam
    Sass, Benjamin
    Waldthaler, Josefine
    Timmermann, Lars
    Nimsky, Christopher
    JOURNAL OF NEUROSURGICAL SCIENCES, 2020,