Applying index of circulating anticoagulant to mixing tests with lupus anticoagulant screen and confirm reagents can distinguish with high specificity between lupus anticoagulants and direct factor Xa inhibitors
被引:1
|作者:
Kumano, Osamu
[1
]
Amiral, Jean
[2
]
Dunois, Claire
[2
]
Peyrafitte, Marie
[2
]
Moore, Gary W.
[3
,4
,5
]
机构:
[1] Sysmex Corp, Prot Technol, Engn 1, Kobe, Hyogo, Japan
[2] Hyphen BioMed, Res Dept, Neuville Sur Oise, France
[3] Addenbrookes Hosp, Dept Haematol, Specialist Haemostasis Unit, Cambridge, England
[4] Middlesex Univ, Fac Sci & Technol, London, England
[5] St Thomas Hosp, Dept Haemostasis & Thrombosis, Viapath Analyt, London, England
关键词:
activated partial thromboplastin time;
antiphospholipid antibodies;
antiphospholipid syndrome;
dilute russell;
lupus anticoagulant;
viper venom time;
TEST-SPECIFIC CUTOFF;
GUIDELINES;
DIAGNOSIS;
UPDATE;
ASSAYS;
D O I:
10.1111/ijlh.13571
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Background Lupus anticoagulants (LA) are detected by prolongation of clotting times for dilute Russell's viper venom time (dRVVT) and activated partial thromboplastin time (APTT) screening tests. Direct oral anticoagulants (DOACs) can interfere with both screening and confirmatory tests. The present study aimed to investigate the influence of direct factor Xa inhibitors (DiXaIs) on screen, confirm and mixing tests and establish a method for differentiation from other sample types. Materials and methods A total of 257 samples including nonanticoagulated LA positive, LA positive with DiXaIs, factor deficiency, FVIII inhibitors, warfarin and non-APS DiXaIs were tested. APTT reagents Cephen LS/Cephen and dRVVT reagents LA1/LA2 were used, respectively, to screen/confirm the study group. Index of circulating anticoagulant (ICA) was calculated from clotting times based on the following formula as ICA screening and ICA confirmation. ICA= (1:1 Mix sample - Normal pooled plasma) / Screen patient x 100. An ICA matrix was established which suggested the presence of a DiXaI when both ICA screening and confirmation were above the cut-off. When only ICA screening is elevated, LA is suspected. Results Sensitivity and specificity of the ICA matrix were 52.2% and 92.8% for DiXaIs and 38.1% and 96.7% for LA in APTT, and 61.2% and 92.9% for DiXaIs and 22.2% and 88.4% for LA in dRVVT, respectively. Conclusion The ICA matrix achieved high specificity with a lower apparent sensitivity for DiXaI samples comparatively to other devices but due only to less interferences: the matrix could contribute to differentiating DiXaIs from LA in samples where anticoagulation status is unknown.
引用
收藏
页码:771 / 778
页数:8