Crystallinity determination of polylactide by FT-Raman spectrometry

被引:64
|
作者
Qin, D [1 ]
Kean, RT [1 ]
机构
[1] Cargill Inc, Cent Res, Minneapolis, MN 55440 USA
关键词
FT-Raman spectrometry; chemometrics; polylactide; crystallinity analysis;
D O I
10.1366/0003702981943950
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The complete Raman spectra of isotropically crystalline and amorphous polylactide (PLA) have been successfully obtained with the use of a Fourier transform instrument equipped with a near-IR laser source. The Raman spectra of PLA were recorded at a resolution of 4 cm(-1) from a backscattering sampling geometry. The changes in band intensity and shape of Raman spectra in the crystalline PLA samples facilitated the quantification of the crystallinity, which was primarily determined by the change in enthalpy via differential scanning calorimetry (DSC). A comparison of quantitative analysis of crystallinity by a multivariate technique, partial least-squares (PLS), and a univariate method at the carbonyl stretching band is described. Quantitative analysis of PLA crystallinity was performed after the data preprocessing with the standard normal variate (SNV) method in the combined spectral regions 3100-2800 cm(-1) and 2000-200 cm(-1) that were determined by the analysis of the correlation spectrum. The use of the calibration model containing mostly 1% D-PLA samples to separately analyze the samples with 1% and 5% D-PLA is discussed. Superior results obtained from the multivariate over the univariate method make PLS the preferred method of choice in quantification of crystallinity. The standard error of prediction (SEP) for the analysis of 1% D-PLA samples with the use of the calibration model described is 0.85 (-J/g), and the SEP for 5% D-PLA samples is 1.35 (-J/g) with the outliers excluded. The discrepancy in analyzing these two types of PLA samples with the use of the same calibration model is also discussed.
引用
收藏
页码:488 / 495
页数:8
相关论文
共 50 条
  • [1] FT-Raman spectrometry determination of Malathion in pesticide formulations
    Quintás, G
    Garrigues, S
    de la Guardia, M
    [J]. TALANTA, 2004, 63 (02) : 345 - 350
  • [2] The Quantitative Analysis of Crystallinity Using FT-Raman Spectroscopy
    Lynne S. Taylor
    George Zografi
    [J]. Pharmaceutical Research, 1998, 15 : 755 - 761
  • [3] The quantitative analysis of crystallinity using FT-Raman spectroscopy
    Taylor, LS
    Zografi, G
    [J]. PHARMACEUTICAL RESEARCH, 1998, 15 (05) : 755 - 761
  • [4] Cellulose I crystallinity determination using FT-Raman spectroscopy: univariate and multivariate methods
    Agarwal, Umesh P.
    Reiner, Richard S.
    Ralph, Sally A.
    [J]. CELLULOSE, 2010, 17 (04) : 721 - 733
  • [5] EVOLUTION OF CRYSTALLINITY IN PHOTODEGRADED POLYETHYLENE FILMS STUDIED BY FT-RAMAN SPECTROSCOPY
    RULL, F
    RODRIGUEZ, J
    ALIA, JM
    ARROYO, F
    EDWARDS, H
    [J]. MACROMOLECULAR SYMPOSIA, 1995, 94 : 189 - 200
  • [6] FT-Raman quantitative determination of ambroxol in tablets
    Szostak, R
    Mazurek, S
    [J]. JOURNAL OF MOLECULAR STRUCTURE, 2004, 704 (1-3) : 229 - 233
  • [7] Pharmaceutical powders analysis using FT-Raman spectrometry:: Simultaneous determination of sulfathiazole and sulfanilamide
    Lopez-Sanchez, Macarena
    Jose Ruedas-Rama, Maria
    Ruiz-Medina, Antonio
    Molina-Diaz, Antonio
    Jose Ayora-Canada, Maria
    [J]. TALANTA, 2008, 74 (05) : 1603 - 1607
  • [8] Sweeteners determination in table top formulations using FT-Raman spectrometry and chemometric analysis
    Armenta, S
    Garrigues, S
    de la Guardia, M
    [J]. ANALYTICA CHIMICA ACTA, 2004, 521 (02) : 149 - 155
  • [9] FT-Raman determination of Mepiquat chloride in agrochemical products
    Quintás, G
    Garrigues, S
    Pastor, A
    de la Guardia, M
    [J]. VIBRATIONAL SPECTROSCOPY, 2004, 36 (01) : 41 - 46
  • [10] Isotacticity determination of polypropylene using FT-Raman spectroscopy
    Sundell, T
    Fagerholm, H
    Crozier, H
    [J]. POLYMER, 1996, 37 (15) : 3227 - 3231