A First Study on the Timing Performance of PMT-Quadrant-Sharing LYSO Detector Array for Time-of-Flight PET

被引:3
|
作者
An, Shaohui [1 ]
Li, Hongdi [1 ]
Liu, Shitao [1 ]
Ramirez, Rocio [1 ]
Zhang, Yuxuan [1 ]
Wang, Chao [1 ]
Baghaei, Hossain [1 ]
Wong, Wai-Hoi [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Expt Diagnost Imaging, Houston, TX 77030 USA
关键词
time-of-flight PET; LYSO arrays; digital signal processing; PMT-Quadrant-Sharing; time resolution; DESIGN;
D O I
10.1109/NSSMIC.2009.5401827
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Time-of-flight (TOF) information can improve the imaging quality of positron emission tomography (PET) camera by calculating the original positron-electron annihilation position. In this work, we measured the preliminary TOF resolution of non-TOF-optimized LYSO arrays for TOF PET using our PMT-Quadrant-Sharing (PQS) technology which has higher light collection efficiency (no external light guide). We investigated digital signal processing methods that may achieve better time resolution. Digital oscilloscope Tektronix DPO7054 was used to capture the waveforms. Three timing methods, Linear Fitting (LF) to the pulse rising time, Constant Fraction Discriminator (CFI)) and Leading edge Discriminator (LD), were studied. A new Time-Energy (T-E) correction using Multivariate Least Mean Square Fit was applied to minimize time walk caused by the variation of signal shape. The time resolution as a function of the depth of interaction was also studied. The average block-to-block coincident timing resolution of 13x13 (4x4x20 mm(3) detector pixel) array with fast PMT R9779 (51-mm diameter) was measured to be 432 Ps (FWHM). Another 13x13 crystal array with ultrahigh spatial resolution 1.4x1.4x10 mm(3) detector pixels coupling to small 19-mm diameter poor-timing XP1912 PMTs was also tested and achieved 551 ps. This study shows that our regular PQS detector blocks (without optimization for TOF) have good TOF time resolution suitable for TOF PET systems, in addition to higher spatial resolution and reducing PMT cost.
引用
收藏
页码:3593 / 3596
页数:4
相关论文
共 50 条
  • [1] Timing Performance Evaluation of PMT-Quadrant-Sharing LYSO Detectors for Time-of-Flight PET
    An, Shaohui
    Li, Hongdi
    Liu, Shitao
    Ramirez, Rocio
    Zhang, Yuxuan
    Wang, Chao
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) : 2155 - 2160
  • [2] Time alignment study of the PMT-Quadrant-Sharing (PQS) LYSO array for time-of-flight PET
    An, Shaohui
    Li, Hongdi
    Liu, Shitao
    Ramirez, Rocio
    Zhang, Yuxuan
    Wang, Chao
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [3] Timing alignment study of PMT-Quadrant-Sharing (PQS) detectors for Time-of-Flight PET
    An, Shaohui
    Li, Hongdi
    Liu, Shitao
    Ramirez, Rocio
    Zhang, Yuxuan
    Wang, Chao
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2758 - 2760
  • [4] Monte Carlo simulation study on the time resolution of a PMT-quadrant-sharing LSO detector block for time-of-flight PET
    Liu, Jiguo
    Liu, Shitao
    Li, Hongdi
    Kim, Soonseok
    Zhang, Yuxuan
    Baghaei, Hossain
    Ramirez, Rocio
    Wong, Wai-Hoi
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3294 - 3304
  • [5] Monte Carlo Simulation Study on the Time Resolution of a PMT-Quadrant-Sharing LSO Detector Block for Time-of-Flight PET
    Liu, Shitao
    Liu, Jiguo
    Zhang, Yuxuan
    Li, Hongdi
    Ramirez, Rocio
    Baghaei, Hossain
    An, Shaohui
    Wang, Chao
    Wong, Wai-Hoi
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3895 - 3901
  • [6] Monte Carlo Simulation Study on the Time Resolution of a PMT-Quadrant-Sharing LSO Detector Block for Time-of-Flight PET
    Liu, Shitao
    Li, Hongdi
    Zhang, Yuxuan
    Ramirez, Rocio A.
    Baghaei, Hossain
    An, Shaohui
    Wang, Chao
    Liu, Jiguo
    Wong, Wai-Hoi
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2614 - 2620
  • [7] Preliminary study on the TOF-timing performance of non-optimized PMT-quadrant-sharing (PQS) LYSO array
    An, Shaohui
    Liu, Shitao
    Li, Hongdi
    Zhang, Yuxuan
    Jaramillo, Rocio A. Ramirez
    Wang, Chao
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [8] Design and Development of a Gap less Ring with Modular PMT-Quadrant-Sharing Detector (PQS) for a Time-of-Flight PET Camera
    Li, Hongdi
    Zhang, Yuxuan
    Ramirez, Rocio
    An, Shaohui
    Wang, Chao
    Liu, Shitao
    Dong, Yun
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [9] Comparison of Two Light Reflector Patterns Designed for PMT-Quadrant-Sharing (PQS) Time-of-Flight PET Detectors
    An, Shaohui
    Li, Hongdi
    Ramirez, Rocio
    Liu, Shitao
    Zhang, Yuxuan
    Wang, Chao
    Baghaei, Hossain
    Wong, Wai-Hoi
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3130 - 3132
  • [10] A high-resolution time-of-flight clinical PET detection system using the PMT-quadrant-sharing technology
    Wong, Wai-Hoi
    Li, Hongdi
    Zhang, Yuxuan
    Ramirez, Rocio
    Baghaei, Hossain
    An, Shaohui
    Wang, Chao
    Liu, Shitao
    Dong, Yun
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55