Finite element analysis of a transmandibular implant

被引:6
|
作者
Nomura, Tsutomu [1 ]
Powers, Michael P.
Katz, J. Lawrence
Saito, Chikara
机构
[1] Niigata Univ, Grad Sch Med & Dent Sci, Div Reconstruct Surg Oral & Maxillofacial Reg, Dept Tissue Regenerat & Reconstruct Course Oral L, Niigata 9518514, Japan
[2] Case Western Reserve Univ, Sch Dent, Dept Oral & Maxillofacial Surg, Cleveland, OH 44109 USA
[3] Case Western Reserve Univ, Case Sch Engn, Dept Biomed Engn, Cleveland, OH 44109 USA
[4] Univ Missouri, Sch Dent, Dept Oral Biol, Kansas City, MO 64108 USA
关键词
transmandibular implant; von Mises stress; finite element analysis; bone height; baseplate;
D O I
10.1002/jbm.b.30606
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Finite element analysis (FEA) was used to study the influence of bony height and baseplate effect on the transmandibular implant (TMI). FEA was used to assess the stress distribution in the mandible, with and without a baseplate. Forces were applied to the center of a bar space. The von Mises stress (SEQV) was displayed in bone, and shear stress also was measured. Tensile stress was always seen at the so-called saddle area; compressive stress was seen at the inferior border. Tensile strain at the saddle area of a 6-mm height mandible is always 9.1 times greater (corresponding to more SEQV) than that of an 18-mm mandible. At same bone heights, saddle area stresses are similar with or without a baseplate. However, stress at the lateral transossseous post interface with bone was significantly different between different bone height groups. Shear stress around the lateral transossseous post was smaller than does the SEQV of the same site. Decreased bone height dramatically increases stress at the saddle area independent of the baseplate presence. These data suggest that bone apposition is found in the severely atrophic mandibles. Thus, there is no need for a baseplate in a slight to mild atrophic mandible. (c) 2006 Wiley Periodicals, Inc.
引用
收藏
页码:370 / 376
页数:7
相关论文
共 50 条
  • [1] A finite element stress analysis of the transmandibular implant system
    Williams, KR
    Murphy, WM
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2001, 14 (02) : 115 - 119
  • [2] A finite element stress analysis of the transmandibular implant system (TMI)
    Williams, KR
    Murphy, WM
    JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1175 - 1175
  • [3] Designing a transmandibular implant with the aid of biostereometric analysis
    Potgieter, HT
    Jooste, CH
    Adams, LP
    JOURNAL OF PROSTHETIC DENTISTRY, 1996, 75 (05): : 525 - 528
  • [4] Finite Element Analysis of a TMJ Implant
    Kashi, A.
    Chowdhury, A. Roy
    Saha, S.
    JOURNAL OF DENTAL RESEARCH, 2010, 89 (03) : 241 - 245
  • [5] FINITE ELEMENT ANALYSIS OF CRANIAL IMPLANT
    Chamrad, J.
    Marcian, P.
    Borak, L.
    Wolff, J.
    ENGINEERING MECHANICS 2016, 2016, : 234 - 237
  • [6] FINITE ELEMENT ANALYSIS OF TMJ IMPLANT
    Kashi, Ajay
    Chowdhury, Amit Roy
    Saha, Subrata
    PROCEEDINGS OF THE 4TH FRONTIERS IN BIOMEDICAL DEVICES CONFERENCE AND EXPOSITION - 2009, 2009, : 93 - 94
  • [7] A technique to improve positioning of the transmandibular posts of the transmandibular implant
    Franco, PF
    Steinberg, MJ
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 56 (06) : 797 - 797
  • [8] Histomorphometric comparison of transmandibular implant and titanium implant
    Nomura T.
    Powers M.P.
    Katz J.L.
    Ejiri S.
    Kobayashi T.
    Journal of Long-Term Effects of Medical Implants, 2016, 26 (03) : 209 - 215
  • [9] Finite Element Analysis of a Temporomandibular Joint Implant
    Mall, Vivek Kumar
    Wahi, Pankaj
    Sinha, Niraj
    Journal of Engineering and Science in Medical Diagnostics and Therapy, 2024, 71 (01)
  • [10] Finite element analysis of dental implant prosthetics
    Aumnakmanee, Suchat
    Yodpiji, Nantakrit
    Jantong, Nattawut
    Jongprasithporn, Manutchanok
    MATERIALS TODAY-PROCEEDINGS, 2018, 5 (03) : 9525 - 9534