Physiological artifacts in scalp EEG and ear-EEG

被引:47
|
作者
Kappel, Simon L. [1 ]
Looney, David [2 ,3 ]
Mandic, Danilo P. [3 ]
Kidmose, Preben [1 ]
机构
[1] Aarhus Univ, Dept Engn, Finlandsgade 22, DK-8200 Aarhus N, Denmark
[2] Pindrop, 817 West Peachtree St NW,Suite 770, Atlanta, GA 24105 USA
[3] Imperial Coll, Dept Elect & Elect Engn, London SW7 2BT, England
来源
基金
英国工程与自然科学研究理事会;
关键词
Ear-EEG; Physiological artifacts; Wearable EEG; Alpha band modulation; STEADY-STATE RESPONSES; EYE-MOVEMENTS; WIRELESS; BLINKS; MOTION; CORTEX;
D O I
10.1186/s12938-017-0391-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A problem inherent to recording EEG is the interference arising from noise and artifacts. While in a laboratory environment, artifacts and interference can, to a large extent, be avoided or controlled, in real-life scenarios this is a challenge. Ear-EEG is a concept where EEG is acquired from electrodes in the ear. Methods: We present a characterization of physiological artifacts generated in a controlled environment for nine subjects. The influence of the artifacts was quantified in terms of the signal-to-noise ratio (SNR) deterioration of the auditory steady-state response. Alpha band modulation was also studied in an open/closed eyes paradigm. Results: Artifacts related to jaw muscle contractions were present all over the scalp and in the ear, with the highest SNR deteriorations in the gamma band. The SNR deterioration for jaw artifacts were in general higher in the ear compared to the scalp. Whereas eye-blinking did not influence the SNR in the ear, it was significant for all groups of scalps electrodes in the delta and theta bands. Eye movements resulted in statistical significant SNR deterioration in both frontal, temporal and ear electrodes. Recordings of alpha band modulation showed increased power and coherence of the EEG for ear and scalp electrodes in the closed-eyes periods. Conclusions: Ear-EEG is a method developed for unobtrusive and discreet recording over long periods of time and in real-life environments. This study investigated the influence of the most important types of physiological artifacts, and demonstrated that spontaneous activity, in terms of alpha band oscillations, could be recorded from the ear-EEG platform. In its present form ear-EEG was more prone to jaw related artifacts and less prone to eye-blinking artifacts compared to state-of-the-art scalp based systems.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Physiological artifacts in scalp EEG and ear-EEG
    Simon L. Kappel
    David Looney
    Danilo P. Mandic
    Preben Kidmose
    [J]. BioMedical Engineering OnLine, 16
  • [2] Case comparison of sleep features from ear-EEG and scalp-EEG
    Zibrandtsen, I.
    Kidmose, P.
    Otto, M.
    Ibsen, J.
    Kjaer, T. W.
    [J]. SLEEP SCIENCE, 2016, 9 (02) : 69 - 72
  • [3] Alpha Oscillations During Effortful Continuous Speech: From Scalp EEG to Ear-EEG
    Ala, Tirdad Seifi
    Alickovic, Emina
    Cabrera, Alvaro Fuentes
    Whitmer, William M. M.
    Hadley, Lauren V. V.
    Rank, Mike L. L.
    Lunner, Thomas
    Graversen, Carina
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2023, 70 (04) : 1264 - 1273
  • [4] Ear-EEG sensitivity modeling for neural sources and ocular artifacts
    Yarici, Metin C.
    Thornton, Mike
    Mandic, Danilo P.
    [J]. FRONTIERS IN NEUROSCIENCE, 2023, 16
  • [5] Ear-EEG Forward Models: Improved Head-Models for Ear-EEG
    Kappel, Simon L.
    Makeig, Scott
    Kidmose, Preben
    [J]. FRONTIERS IN NEUROSCIENCE, 2019, 13
  • [6] EEG Recorded from the Ear: Characterizing the Ear-EEG Method
    Mikkelsen, Kaare B.
    Kappel, Simon L.
    Mandic, Danilo P.
    Kidmose, Preben
    [J]. FRONTIERS IN NEUROSCIENCE, 2015, 9
  • [7] Auditory Steady-State Responses across Chirp Repetition Rates for Ear-EEG and Scalp EEG
    Christensen, C. B.
    Kappel, S. L.
    Kidmose, P.
    [J]. 2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 1376 - 1379
  • [8] Ear-EEG detects ictal and interictal abnormalities in focal and generalized epilepsy - A comparison with scalp EEG monitoring
    Zibrandtsen, I. C.
    Kidmose, P.
    Christensen, C. B.
    Kjaer, T. W.
    [J]. CLINICAL NEUROPHYSIOLOGY, 2017, 128 (12) : 2454 - 2461
  • [9] High-density ear-EEG for understanding ear-centered EEG
    Meiser, Arnd
    Knoll, Anna Lena
    Bleichner, Martin G.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2024, 21 (01)
  • [10] Portable Ear-EEG Acquisition System
    Ying, Shaofei
    [J]. INTERNATIONAL JOURNAL OF PSYCHOPHYSIOLOGY, 2021, 168 : S225 - S226