Performance characteristics of a new prototype for a portable GC using ambient air as carrier gas for on-site analysis

被引:13
|
作者
Sanchez, Juan M. [1 ]
Sacks, Richard D.
机构
[1] Univ Girona, Dept Chem, Girona 17071, Spain
[2] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
关键词
at-column heating; atmospheric air carrier gas; portable GC; semivolatile organic compounds; vacuum-outlet GC;
D O I
10.1002/jssc.200600367
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance characteristics of a portable GC instrument requiring no compressed gas supplies and using relatively lightweight transportable components for the analysis of volatile organic components in large-volume air samples are described. To avoid the need for compressed gas tanks, ambient air is used as the carrier gas, and a vacuum pump is used to pull the carrier gas and injected samples through the wall-coated capillary column and a photoionization detector (PID). At-column heating is used eliminating the need for a conventional oven. The fused silica column is wrapped with heater wire and sensor wire so that heating is provided directly at the column. A PID is used since it requires no external gas supplies and has high sensitivity for many compounds of interest in environmental air monitoring. In order to achieve detection limits in the ppb range, an online multibed preconcentrator containing beds of graphitized carbons and carbon molecular sieves is used. After sample collection, the flow direction through the preconcentrator is reversed, and the sample is thermally desorbed directly into the column. Decomposition of sensitive compounds during desorption is greater with air as the carrier gas than with hydrogen.
引用
收藏
页码:1052 / 1060
页数:9
相关论文
共 50 条
  • [2] Performance evaluation of a new portable GC with photoionization detector for on-site real-time monitoring of environmental VOC
    Nam, Paul
    Bose, Mousumi
    He, Xiaolong
    Shi, Honglan
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 256
  • [3] A Portable and Autonomous Mass Spectrometric System for On-Site Environmental Gas Analysis
    Brennwald, Matthias S.
    Schmidt, Mark
    Oser, Julian
    Kipfer, Rolf
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2016, 50 (24) : 13455 - 13463
  • [4] Towards on-site drug evidence confirmation using ambient sampling, portable mass spectrometry
    Mulligan, Christopher
    Fatigante, William
    Stelmack, Ashley
    Burr, Daniel
    Harms, John
    Kim, Jun-Hyun
    Driskell, Jeremy
    Wieland, Jamie
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2019, 257
  • [5] A new integrated approach to field GC/MS for on-site analysis
    Eckenrode, B
    FIELD ANALYTICAL METHODS FOR HAZARDOUS WASTES AND TOXIC CHEMICALS, 1997, : 480 - 489
  • [6] A Gas Chromatographic System for the Detection of Ethylene Gas Using Ambient Air as a Carrier Gas
    Zaidi, Nayyer Abbas
    Tahir, Muhammad Waseem
    Vellekoop, Michael J.
    Lang, Walter
    SENSORS, 2017, 17 (10)
  • [7] AMBIENT AIR ANALYSIS USING PORTABLE INFRARED ANALYSERS.
    Syrjala, Richard
    Process Engineering (Sydney), 1980, 8 (02):
  • [8] A method for on-site analysis of urinary benzene by means of a portable gas-chromatograph
    Suna, S
    Jitsunari, F
    Asakawa, F
    Hirao, T
    Mannami, T
    Suzue, T
    JOURNAL OF OCCUPATIONAL HEALTH, 2005, 47 (01) : 74 - 77
  • [9] Preliminary evaluation of a next-generation portable gas chromatograph mass spectrometer (GC-MS) for the on-site analysis of ignitable liquid residues
    Visotin, Alexander
    Lennard, Chris
    AUSTRALIAN JOURNAL OF FORENSIC SCIENCES, 2016, 48 (02) : 203 - 221
  • [10] A simple and rapid chemiluminescence assay for on-site analysis of paraquat using a portable luminometer
    Naoya Kishikawa
    Sho Higuchi
    Kaname Ohyama
    Kenichiro Nakashima
    Naotaka Kuroda
    Forensic Toxicology, 2013, 31 : 301 - 306