Detection of heavy metals in aqueous solution using PGNAA technique

被引:9
|
作者
Cheng, Can [1 ,2 ]
Hei, Da-Qian [1 ]
Jia, Wen-Bao [1 ,2 ]
Jiang, Zhou [1 ]
Wang, Hong-Tao [1 ]
机构
[1] Nanjing Univ Aeronaut & Astronaut, Coll Mat Sci & Engn, Dept Nucl Sci & Engn, Nanjing 211106, Jiangsu, Peoples R China
[2] Jiangsu Higher Educ Inst, Collaborat Innovat Ctr Radiat Med, Suzhou 215000, Peoples R China
基金
中国国家自然科学基金;
关键词
Prompt gamma neutron activation analysis; Heavy metals; Linearity response; Minimum detectable concentrations; SCINTILLATION DETECTOR; ELEMENTS DETECTION; SET-UP; OPTIMIZATION; WATER; CONCRETE; CADMIUM; SAMPLES; BORON;
D O I
10.1007/s41365-016-0010-0
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A prompt gamma neutron activation analysis setup was developed for heavy metal detection in aqueous solutions with a 300 mCi Am-241-Be neutron source and a 4 x 4 inch (diameter x height) BGO detector. In the present work, heavy metals, including Mn, Cu, Ni, Cr and Zn, were measured by the setup. The minimum detectable concentrations of Mn, Cu, Ni, Cr and Zn were 246.6, 391.2, 218.1, 301.5 and 2804.1 ppm, respectively. The minimum detectable concentration of each element and the linearity response between the characteristic peak counts and elements concentrations have been studied. And the results showed that all heavy metals had a good linear relationship between characteristic peak counts and concentrations.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Detection of heavy metals in aqueous solution using PGNAA technique
    Can Cheng
    Da-Qian Hei
    Wen-Bao Jia
    Zhou Jiang
    Hong-Tao Wang
    Nuclear Science and Techniques, 2016, 27
  • [2] Detection of heavy metals in aqueous solution using PGNAA technique
    Can Cheng
    Da-Qian Hei
    Wen-Bao Jia
    Zhou Jiang
    Hong-Tao Wang
    NuclearScienceandTechniques, 2016, 27 (01) : 72 - 76
  • [3] Heavy metals detection in sediments using PGNAA method
    Hei Da-Qian
    Jia Wen-Bao
    Jiang Zhou
    Cheng Can
    Li Jia-Tong
    Wang Hong-Tao
    APPLIED RADIATION AND ISOTOPES, 2016, 112 : 50 - 54
  • [4] Improving a PGNAA Technique to Detect Heavy Metals in Solid Samples
    AlMisned, Ghada
    Al-Abdullah, Tariq
    Liadi, Fatai
    Hawsawi, Adil
    APPLIED SCIENCES-BASEL, 2022, 12 (08):
  • [5] Optimization of PGNAA set-up for the elements detection in aqueous solution
    WenBao Jia
    DaQian Hei
    Can Cheng
    HaoJia Zhang
    Qing Shan
    Science China Technological Sciences, 2014, 57 : 625 - 629
  • [6] Optimization of PGNAA set-up for the elements detection in aqueous solution
    Jia WenBao
    Hei DaQian
    Cheng Can
    Zhang HaoJia
    Shan Qing
    SCIENCE CHINA-TECHNOLOGICAL SCIENCES, 2014, 57 (03) : 625 - 629
  • [7] Optimization of PGNAA set-up for the elements detection in aqueous solution
    JIA WenBao
    HEI DaQian
    CHENG Can
    ZHANG HaoJia
    SHAN Qing
    Science China(Technological Sciences), 2014, (03) : 625 - 629
  • [8] Optimization of PGNAA set-up for the elements detection in aqueous solution
    JIA WenBao
    HEI DaQian
    CHENG Can
    ZHANG HaoJia
    SHAN Qing
    Science China(Technological Sciences), 2014, 57 (03) : 625 - 629
  • [9] Study on the PGNAA. measurement of heavy metals in aqueous solution by the Monte Carlo Library Least-Squares (MCLLS) approach
    Zhang, Yan
    Jia, WenBao
    Gardner, Robin
    Shan, Qing
    Hei, Daqian
    APPLIED RADIATION AND ISOTOPES, 2018, 132 : 13 - 17
  • [10] Removal of heavy metals from aqueous solution by using mango biomass
    Ashraf, Muhammad Aqeel
    Wajid, Abdul
    Mahmood, Karamat
    Maah, Mohd Jamil
    Yusoff, Ismail