Determination of the detective quantum efficiency of gamma camera systems: a Monte Carlo study

被引:2
|
作者
Eriksson, Ida [1 ]
Starck, Sven-Ake [2 ]
Bath, Magnus [3 ,4 ]
机构
[1] Karlstad Hosp, Dept Med Phys, SE-65185 Karlstad, Sweden
[2] Cty Hosp Ryhov, Dept Oncol, Hosp Phys Unit, SE-55185 Jonkoping, Sweden
[3] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, SE-41345 Gothenburg, Sweden
[4] Univ Gothenburg, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
关键词
DIGITAL CHEST RADIOGRAPHY; NODULE DETECTION; SCINTILLATION CAMERA; NOISE; PERFORMANCE; GENERATIONS; RESOLUTION; SCANNER; SPECT; TRIAL;
D O I
10.1093/rpd/ncq055
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of the present work was to investigate the validity of using the Monte Carlo technique for determining the detective quantum efficiency (DQE) of a gamma camera system and to use this technique in investigating the DQE behaviour of a gamma camera system and its dependency on a number of relevant parameters. The Monte Carlo-based software SIMIND, simulating a complete gamma camera system, was used in the present study. The modulation transfer function (MTF) of the system was determined from simulated images of a point source of (99m)Tc, positioned at different depths in a water phantom. Simulations were performed using different collimators and energy windows. The MTF of the system was combined with the photon yield and the sensitivity, obtained from the simulations, to form the frequency-dependent DQE of the system. As figure-of-merit (FOM), the integral of the 2D DQE was used. The simulated DQE curves agreed well with published data. As expected, there was a strong dependency of the shape and magnitude of the DQE curve on the collimator, energy window and imaging position. The highest FOM was obtained for a lower energy threshold of 127 keV for objects close to the detector and 131 keV for objects deeper in the phantom, supporting an asymmetric window setting to reduce scatter. The Monte Carlo software SIMIND can be used to determine the DQE of a gamma camera system from a simulated point source alone. The optimal DQE results in the present study were obtained for parameter settings close to the clinically used settings.
引用
收藏
页码:219 / 227
页数:9
相关论文
共 50 条
  • [1] Comparison of gamma ( Anger) camera systems in terms of detective quantum efficiency using Monte Carlo simulation
    Eriksson, Ida
    Starck, Sven-Ake
    Bath, Magnus
    NUCLEAR MEDICINE COMMUNICATIONS, 2014, 35 (04) : 405 - 415
  • [2] The use of detective quantum efficiency (DQE) in evaluating the performance of gamma camera systems
    Starck, SÅ
    Båth, M
    Carlsson, S
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07): : 1601 - 1609
  • [3] Determination of the DQE of gamma camera systems using the Monte Carlo technique
    Eriksson, I.
    Starck, S.
    Bath, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S393 - S393
  • [4] Detective quantum efficiency of a storage phosphor imaging plate: Monte Carlo study and experimental results
    Barnea, G
    PPC-2003: 14TH IEEE INTERNATIONAL PULSED POWER CONFERENCE, VOLS 1 AND 2, DIGEST OF TECHNICAL PAPERS, 2003, : 82 - 85
  • [5] Monte Carlo study on image shrinkage of a small gamma camera
    Zeng, HN
    Xu, ZZ
    Wang, ZM
    Li, C
    Wang, CR
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2000, 24 (02): : 166 - 171
  • [6] Monte Carlo study on image shrinkage of a small gamma camera
    Zeng, Haining
    Xu, Zizong
    Wang, Zhaomin
    Li, Cheng
    Wang, Congrong
    Kao Neng Wu Li Yu Ho Wu Li/High Energy Physics and Nuclear Physics, 2000, 24 (02): : 166 - 171
  • [7] ON THE DETECTIVE QUANTUM EFFICIENCY OF TELEVISION CAMERA TUBES
    JONES, RC
    JOURNAL OF THE SMPTE-SOCIETY OF MOTION PICTURE AND TELEVISION ENGINEERS, 1959, 68 (07): : 462 - 466
  • [8] Monte Carlo detective quantum efficiency and scatter studies of a metal/a-Se portal detector
    Lachaine, M
    Fallone, BG
    MEDICAL PHYSICS, 1998, 25 (07) : 1186 - 1194
  • [9] Monte Carlo detective quantum efficiency and scatter studies of a metal/a-Se portal detector
    McGill University, Medical Physics Unit, Montreal General Hospital, 1650 Cedar Ave., Montreal, Que. H3G 1A4, Canada
    Med. Phys., 7 I (1186-1194):
  • [10] The Influence of Inter-Crystal Scattering on Detection Efficiency of Dedicated Breast Gamma Camera: A Monte Carlo Study
    Rasouli, M.
    Ay, M. R.
    Takavar, A.
    Lashkari, S.
    Loudos, G.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 2451 - 2454