A Comparison Between the NIST PJVS-Based Power Standard and the NRC Current-Comparator-Based Power Standard

被引:0
|
作者
Waltrip, Bryan C. [1 ]
Nelson, Thomas L. [1 ]
So, Eddy [2 ]
Angelo, David [2 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
[2] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
关键词
Bridge circuits; current-comparator; Josephson junctions; power measurement; wattmeters; METER CALIBRATIONS; 60; HZ;
D O I
10.1109/TIM.2014.2329386
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents the results of a comparison of active/reactive power meter calibrations between the National Institute of Standards and Technology and National Research Council. The comparison was implemented using a transfer standard consisting of a highly stable commercial sampling-type power/energy meter. Active and reactive power measurements were made at 120 V, 5 A, 50 Hz, and 60 Hz. For active power, the measurements were made at applied current phase angles of 0 degrees, +60 degrees, and -60 degrees. For reactive power, the measurements were made at applied current phase angles of +60 degrees, +90 degrees, -60 degrees, and -90 degrees. The results of the comparison indicate agreement to within the stated uncertainties of the participants.(1)
引用
收藏
页码:14 / 18
页数:5
相关论文
共 50 条
  • [1] An Improved Current-Comparator-Based Power Standard With an Uncertainty of 2.5 μW/VA (k=1)
    So, Eddy
    Arseneau, Rejean
    Angelo, Dave
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (06) : 1704 - 1709
  • [2] Development of a PJVS Based Primary AC Power Standard
    Ihlenfeld, W. G. Kurten
    Landim, R. P.
    Franco, A. M. R.
    Debatin, R. M.
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [3] Design of DC Power Standard Source Based on PJVS
    Qian, Lushuai
    Zhai, Changwei
    Zhao, Jianting
    Qu, Jifeng
    Shi, Yang
    Fu, Yaqiong
    2024 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CPEM 2024, 2024,
  • [4] A MODIFIED TECHNIQUE FOR THE CALIBRATION OF A CURRENT-COMPARATOR-BASED HIGH VOLTAGE CAPACITANCE BRIDGE AND ITS COMPARISON BETWEEN KRISS AND NRC
    Faisal, Agah
    Jung, Jae Kap
    So, Eddy
    2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 671 - +
  • [5] A CURRENT-COMPARATOR-BASED SYSTEM FOR CALIBRATING ACTIVE REACTIVE POWER AND ENERGY METERS
    MOORE, WJM
    SO, E
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1983, 32 (01) : 147 - 149
  • [7] A Synchronized Current-Comparator-Based Power Bridge for Calibrating Analog Merging Units
    Djokic, Branislav
    Parks, Harold
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [8] A current-comparator-based bridge for calibrating power and energy standards at 50/60 Hz
    Kyriazis, GA
    Oldham, NM
    Werneck, M
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 256 - 257
  • [10] A two-way Josephson voltage standard comparison between NIST and NRC
    Tang, Y.
    Wood, B.
    Hamilton, C. A.
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 600 - +