Sol-gel-derived thick-film amperometric immunosensors

被引:207
|
作者
Wang, J [1 ]
Pamidi, PVA
Rogers, KR
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
[2] US EPA, Natl Exposure Res Lab, Las Vegas, NV 89139 USA
关键词
D O I
10.1021/ac971093e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sol-gel processing is used for the first time for the preparation of electrochemical immunosensors. One-step sensor fabrication, based on the coupling of sol-gel and screen-printing technologies, is employed. A low-temperature cured ink is prepared by dispersion of rabbit immunoglobulin G (RIgG), graphite powder, and a binder in the sol-gel solution. The enzyme-labeled antibody can readily diffuse toward the encapsulated antigen, which retains its binding properties, and the association reaction is easily detected at the dispersed graphite surface. Use of anti-RIgG labeled with alkaline phosphatase, naphthyl phosphate as the substrate, and amperometric detection at +400 mV (vs Ag/AgCl) results in a low detection limit of 5 ng/mL (32 pM) for the solution antigen. Tailoring the porosity of the ceramic-carbon matrix can be used for tuning the assay performance. The high sensitivity, low cost, durability, and simplicity of the new single-use immunosensors make them well suited for various onsite applications.
引用
收藏
页码:1171 / 1175
页数:5
相关论文
共 50 条
  • [1] Sol-gel-derived thick-film conductometric biosensor for urea determination in serum
    Lee, WY
    Kim, SR
    Kim, TH
    Lee, KS
    Shin, MC
    Park, JK
    ANALYTICA CHIMICA ACTA, 2000, 404 (02) : 195 - 203
  • [2] Sol-gel-derived percolative copper film
    Szu, Sungping
    Cheng, Ching-Lung
    MATERIALS RESEARCH BULLETIN, 2008, 43 (10) : 2687 - 2696
  • [3] Dewetting of a sol-gel-derived thin film
    Kwon, SJ
    Park, JG
    LANGMUIR, 2006, 22 (08) : 3895 - 3898
  • [4] Wrinkling of a sol-gel-derived thin film
    Kwon, SJ
    Park, JH
    Park, JG
    PHYSICAL REVIEW E, 2005, 71 (01)
  • [5] Sol-gel-derived amperometric biosensor for hydrogen peroxide based on methylene green incorporated in Nafion film
    Wang, BQ
    Dong, SJ
    TALANTA, 2000, 51 (03) : 565 - 572
  • [6] Dynamic instability of a sol-gel-derived thin film
    Kwon, S. Joon
    Park, Jae-Gwan
    JOURNAL OF PHYSICAL CHEMISTRY B, 2008, 112 (07): : 2016 - 2023
  • [7] Amperometric glucose biosensor based on sol-gel-derived zirconia/Nafion composite film as encapsulation matrix
    Kim, HJ
    Yoon, SH
    Choi, HN
    Lyu, YK
    Lee, WY
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2006, 27 (01): : 65 - 70
  • [8] Sol-gel-derived metal-dispersed carbon composite amperometric biosensors
    Wang, J
    Pamidi, PVA
    Park, DS
    ELECTROANALYSIS, 1997, 9 (01) : 52 - 55
  • [9] Horseradish peroxidase sol-gel-derived bioelectrode and amperometric determination of hydrogen peroxide
    Wang, ZH
    Li, QW
    Liu, J
    Luo, G
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 29 (12) : 1482 - 1482
  • [10] Sol-gel-derived prussian blue-silicate amperometric glucose biosensor
    Bharathi, S
    Lev, O
    APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 2000, 89 (2-3) : 209 - 216