Assessing Body Composition in Healthy Newborn Infants: Reliability of Dual-Energy X-Ray Absorptiometry

被引:49
|
作者
Godang, Kristin [1 ]
Qvigstad, Elisabeth
Voldner, Nanna [2 ]
Isaksen, Gunhild A.
Froslie, Kathrine F. [2 ,3 ,4 ]
Notthellen, Jacob [5 ]
Henriksen, Tore [2 ]
Bollerslev, Jens
机构
[1] Natl Hosp Norway, Endocrinol Sect, Oslo Univ Hosp, Dept Endocrinol, N-0027 Oslo, Norway
[2] Natl Hosp Norway, Dept Obstet & Gynecol, Oslo Univ Hosp, N-0027 Oslo, Norway
[3] Natl Hosp Norway, Natl Resource Ctr Womens Hlth, Oslo Univ Hosp, N-0027 Oslo, Norway
[4] Univ Oslo, Inst Basic Med Sci, Dept Biostat, Oslo, Norway
[5] Natl Hosp Norway, Dept Med Phys, Oslo Univ Hosp, N-0027 Oslo, Norway
关键词
Body composition; caliper; DXA; infants; reliability; BONE-MINERAL MEASUREMENTS; SOFT-TISSUE COMPOSITION; SKINFOLD THICKNESS MEASUREMENTS; PHOTON-ABSORPTIOMETRY; TECHNICAL CONSIDERATIONS; PREMATURE-INFANTS; FAT MEASUREMENTS; LUMBAR SPINE; VALIDATION; DENSITOMETRY;
D O I
10.1016/j.jocd.2010.01.121
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Dual-energy X-ray absorptiometry (DXA) is used to measure body composition in newborns; however, data on DXA accuracy are limited. We investigated the reliability of body composition measurements by DXA. The present study included 207 normal-term newborn babies, recruited from a larger study on the determinants of birth weight in healthy pregnancies (STORK) between 2005 and 2008. Reliability analysis of total fat mass (FM(DxA)), fat-free mass, lean mass (LM(DxA)), bone mineral content (BMC), and bone mineral density (BMD) were based on 2 DXA scans of 50 neonates. We also performed a comparison analysis for DXA (FM(DxA)) measurements and caliper (CLP) or circumference (CF) measurements of trunk and extremities (performed on all neonates, n = 207). Reliability: All intraclass correlation coefficients (ICC) were satisfactory to excellent for total body and the extremity-compartment FM(DxA), LM(DxA), BMD, and BMC; ICC ranged from 0.86 to 0.96 but with a lower ICC for trunk FM(DxA). For comparison analysis, the Pearson correlation coefficients for CLP vs DXA and CF vs DXA ranged from 0.48 to 0.79 and 0.41 to 0.77, respectively. Quadriceps CLP and CF measurements correlated best with the most reliable DXA results, whereas more modest correlations were found for the trunk region. DXA measurements of body composition demonstrated good reliability and can be used as a reference method in neonates. CLP and CF measurements are appropriate for larger cohorts or when DXA is unavailable, and they provide fair rough estimations of fat mass.
引用
收藏
页码:151 / 160
页数:10
相关论文
共 50 条
  • [1] Dual-energy X-ray absorptiometry and body composition
    Laskey, MA
    NUTRITION, 1996, 12 (01) : 45 - 51
  • [2] Dual-energy X-ray absorptiometry and body composition
    Plank, LD
    CURRENT OPINION IN CLINICAL NUTRITION AND METABOLIC CARE, 2005, 8 (03): : 305 - 309
  • [3] DUAL-ENERGY X-RAY ABSORPTIOMETRY IN THE NEWBORN
    SALLE, BL
    LAPILLONNE, A
    BRAILLON, P
    ARCHIVES FRANCAISES DE PEDIATRIE, 1993, 50 (07): : 545 - 546
  • [4] Assessing body composition and changes in body composition - Another look at dual-energy X-ray absorptiometry
    Lohman, TG
    Harris, M
    Teixeira, PJ
    Weiss, L
    IN VIVO BODY COMPOSITION STUDIES, 2000, 904 : 45 - 54
  • [5] Dual-Energy X-Ray Absorptiometry in Small Subjects: Influence of Dual-Energy X-Ray Equipment on Assessment of Mineralization and Body Composition in Newborn Piglets
    Jean-Charles Picaud
    Kindja Nyamugabo
    Pierre Braillon
    Alexandre Lapillonne
    Olivier Claris
    Pierre Delmas
    Pierre Meunier
    Bernard Salle
    Jacques Rigo
    Pediatric Research, 1999, 46 : 772 - 772
  • [6] Dual-energy x-ray absorptiometry in small subjects: Influence of dual-energy x-ray equipment on assessment of mineralization and body composition in newborn piglets
    Picaud, JC
    Nyamugabo, K
    Braillon, P
    Lapillonne, A
    Claris, O
    Delmas, P
    Meunier, P
    Salle, B
    Rigo, J
    PEDIATRIC RESEARCH, 1999, 46 (06) : 772 - 777
  • [7] Body composition by dual-energy x-ray absorptiometry in women with fibromyalgia
    Marques Teles Lobo, Marcia Maria
    Paiva, Eduardo dos Santos
    Andretta, Aline
    Madalozzo Schieferdecker, Maria Eliana
    REVISTA BRASILEIRA DE REUMATOLOGIA, 2014, 54 (04) : 273 - 278
  • [8] Pediatric body composition analysis with dual-energy X-ray absorptiometry
    Maura Helba
    Larry A. Binkovitz
    Pediatric Radiology, 2009, 39 : 647 - 656
  • [9] Reliability of 2 Different Positioning Protocols for Dual-Energy X-ray Absorptiometry Measurement of Body Composition in Healthy Adults
    Kerr, Ava
    Slater, Gary J.
    Byrne, Nuala
    Nana, Alisa
    JOURNAL OF CLINICAL DENSITOMETRY, 2016, 19 (03) : 282 - 289
  • [10] Body Composition in Karate: A Dual-Energy X-ray Absorptiometry Study
    Cavedon, Valentina
    Milanese, Chiara
    Sacristani, Franco
    Zancanaro, Carlo
    APPLIED SCIENCES-BASEL, 2023, 13 (01):