A Comparative Study of Single-pulse and Double-pulse Laser-Induced Breakdown Spectroscopy with Uranium-containing Samples

被引:29
|
作者
Skrodzki, Patrick J. [1 ]
Becker, Jason R. [1 ]
Diwakar, Prasoon K. [1 ]
Harilal, Sivanandan S. [2 ]
Hassanein, Ahmed [1 ]
机构
[1] Purdue Univ, Ctr Mat Extreme Environm CMUXE, Sch Nucl Engn, 400 Cent Dr, W Lafayette, IN 47907 USA
[2] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
Double-pulse laser-induced breakdown spectroscopy; DPLIBS; Nuclear forensics; Uranium; Signal enhancement; Plasma diagnostics; WAVELENGTH; PLASMA; ENHANCEMENT; ABLATION; THORIUM;
D O I
10.1177/0003702815626670
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) holds potential advantages in special nuclear material (SNM) sensing and nuclear forensics, which require rapid analysis, minimal sample preparation, and stand-off distance capability. SNM, such as U, however, result in crowded emission spectra with LIBS, and characteristic emission lines are challenging to discern. It is well-known that double-pulse LIBS (DPLIBS) improves the signal intensity for analytes over conventional single-pulse LIBS (SPLIBS). This study investigates the U signal in a glass matrix using DPLIBS and compares it to signal obtained using SPLIBS. Double-pulse LIBS involves sequential firing of a 1.06 mu m Nd:YAG pre-pulse and 10.6 mu m TEA CO2 heating pulse in a near collinear geometry. Optimization of experimental parameters including inter-pulse delay and energy follows identification of characteristic lines for the bulk analyte Ca and the minor constituent analyte U for both DPLIBS and SPLIBS. Spatial and temporal coupling of the two pulses in the proposed DPLIBS technique yields improvements in analytical merits with a negligible increase in damage to the sample compared to SPLIBS. Subsequently, the study discusses optimum plasma emission conditions of U lines and relative figures of merit in both SPLIBS and DPLIBS. Investigation into plasma characteristics also addresses plausible mechanisms related to the observed U analyte signal variation between SPLIBS and DPLIBS.
引用
下载
收藏
页码:467 / 473
页数:7
相关论文
共 50 条
  • [1] Experimental and theoretical comparison of single-pulse and double-pulse laser induced breakdown spectroscopy on metallic samples
    De Giacomo, A.
    Dell'Aglio, M.
    Bruno, D.
    Gaudiuso, R.
    De Pascale, O.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2008, 63 (07) : 805 - 816
  • [2] Double-pulse and single-pulse laser-induced breakdown spectroscopy for distinguishing between gaseous and particulate phase analytes
    Asgill, Michael E.
    Brown, Michael S.
    Frische, Kyle
    Roquemore, William M.
    Hahn, David W.
    APPLIED OPTICS, 2010, 49 (13) : C110 - C119
  • [3] Double-pulse laser ablation coupled to laser-induced breakdown spectroscopy
    Glaus, Reto
    Hahn, David W.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2014, 98 : 48 - 53
  • [4] Comparative Study of Two Methods of Orthogonal Double-Pulse Laser-Induced Breakdown Spectroscopy of Aluminum
    Safi, A.
    Bahreini, M.
    Tavassoli, S. H.
    OPTICS AND SPECTROSCOPY, 2016, 120 (03) : 367 - 378
  • [5] Comparative Study of Two Methods of Orthogonal Double-Pulse Laser-Induced Breakdown Spectroscopy of Aluminum
    A. Safi
    M. Bahreini
    S. H. Tavassoli
    Optics and Spectroscopy, 2016, 120 : 367 - 378
  • [6] Comparison of double-pulse and single-pulse laser-induced breakdown spectroscopy techniques in the analysis of powdered samples of silicate raw materials for the brick-and-tile industry
    Ctvrtnickova, Tereza
    Cabalin, Luisa M.
    Laserna, Javier
    Kanicky, Viktor
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2008, 63 (01) : 42 - 50
  • [7] A Comparative Study of the Laser Induced Breakdown Spectroscopy in Single- and Collinear Double-Pulse Laser Geometry
    孙对兄
    苏茂根
    董晨钟
    温冠宏
    Plasma Science and Technology, 2014, (04) : 374 - 379
  • [8] A Comparative Study of the Laser Induced Breakdown Spectroscopy in Single- and Collinear Double-Pulse Laser Geometry
    Sun Duixiong
    Su Maogen
    Dong Chenzhong
    Wen Guanhong
    PLASMA SCIENCE & TECHNOLOGY, 2014, 16 (04) : 374 - 379
  • [9] A Comparative Study of the Laser Induced Breakdown Spectroscopy in Single- and Collinear Double-Pulse Laser Geometry
    孙对兄
    苏茂根
    董晨钟
    温冠宏
    Plasma Science and Technology, 2014, 16 (04) : 374 - 379
  • [10] Effect of atmosphere on collinear double-pulse laser-induced breakdown spectroscopy
    Andrew J. Effenberger
    Jill R. Scott