Comparison of the anterior capsulotomy edge created by manual capsulorhexis and 2 femtosecond laser platforms: Scanning electron microscopy study

被引:22
|
作者
Al Harthi, Khaled [1 ]
Al Shahwan, Sami [1 ]
Al Towerki, Abdulelah [1 ]
Banerjee, P. Pat [2 ]
Behrens, Ashley [1 ,3 ]
Edward, Deepak P. [1 ,3 ]
机构
[1] King Khalid Eye Specialist Hosp, Riyadh 11462, Saudi Arabia
[2] Univ Illinois, Dept Mech & Ind Engn, Chicago, IL USA
[3] Johns Hopkins Univ, Wilmer Eye Inst, Baltimore, MD 21218 USA
来源
关键词
INTRAOCULAR-LENS DISLOCATION; ASSISTED CATARACT-SURGERY; IN-THE-BAG; INTERFACE;
D O I
10.1016/j.jcrs.2014.10.001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the scanning electron microscopy (SEM) features of the anterior capsule edge created by continuous curvilinear capsulorhexis (CCC) and femtosecond laser-assisted capsulotomy using angular second moment and contrast. SETTING: King Khaled Eye Specialist Hospital, Riyadh, Kingdom of Saudi Arabia. DESIGN: Prospective comparative series. METHODS: Anterior capsule specimens following CCC (n = 10) or femtosecond laser-assisted capsulotomy using 2 platforms (Lensx, n = 9, and Victus, n = 10) were studied by SEM. Irregularity of the capsule edge was quantified using 2 parameters (angular second moment and contrast) using ImageJ software. The clinical features and laser parameters were correlated with angular second moment and contrast. RESULTS: By SEM, the femtosecond laser-assisted capsulotomy edge surfaces created by both lasers showed marked irregularity compared with the smooth edge of the CCC. The angular second moment and contrast measures for both lasers differed significantly from those obtained for CCC (P<.001). There was no between-laser difference in angular second moment and contrast measures. The angular second moment showed only a weak negative correlation with increasing laser power, whereas contrast showed a weak positive correlation with increasing power. CONCLUSIONS: Both laser platforms created an irregular capsulotonny edge. The angular second moment and contrast quantified capsule edge irregularities for further comparisons. These measures could be used to quantify efforts to reduce capsule-edge irregularity from femtosecond laser-lens capsule interaction.
引用
下载
收藏
页码:2106 / 2112
页数:7
相关论文
共 30 条
  • [1] Scanning Electron Microscopy Analysis of the Anterior Capsulotomy Edge: A Comparative Study between Femtosecond Laser-Assisted Capsulotomy and Manual Capsulorhexis
    Tognetto, Daniele
    De Giacinto, Chiara
    Perrotta, Alberto Armando
    Candian, Tommaso
    Bova, Alessandro
    Rinaldi, Silvia
    Turco, Gianluca
    JOURNAL OF OPHTHALMOLOGY, 2018, 2018
  • [2] Analysis of Femtosecond Laser Assisted Capsulotomy Cutting Edges and Manual Capsulorhexis Using Environmental Scanning Electron Microscopy
    Serrao, Sebastiano
    Lombardo, Giuseppe
    Desiderio, Giovanni
    Buratto, Lucio
    Schiano-Lomoriello, Domenico
    Pileri, Marco
    Lombardo, Marco
    JOURNAL OF OPHTHALMOLOGY, 2014, 2014
  • [3] Comparison of the Mechanical Properties of the Anterior Lens Capsule Following Manual Capsulorhexis and Femtosecond Laser Capsulotomy
    Sandor, Gabor L.
    Kiss, Zoltan
    Bocskai, Zoltan I.
    Kolev, Krasimir
    Takacs, Agnes I.
    Juhasz, Eva
    Kranitz, Kinga
    Toth, Gabor
    Gyenes, Andrea
    Bojtar, Imre
    Juhasz, Tibor
    Nagy, Zoltan Z.
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (10) : 660 - 664
  • [4] Intereye comparison of femtosecond laser-assisted cataract surgery capsulotomy and manual capsulorhexis edge strength
    Chan, Thomas
    Pattamatta, Ushasree
    Butlin, Mark
    Meades, Kerrie
    Bala, Chandra
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (04): : 480 - 485
  • [5] INTEREYE COMPARISON OF FEMTOSECOND LASER-ASSISTED CATARACT SURGERY CAPSULOTOMY AND MANUAL CAPSULORHEXIS EDGE STRENGTH
    Meades, Kerrie
    Pattamatta, Usharee
    Chan, Thomas
    Butlin, Mark
    Bala, Chandra
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2017, 45 : 44 - 44
  • [6] Light microscopy and scanning electron microscopy analysis of rigid curved interface femtosecond laser-assisted and manual anterior capsulotomy
    Ostovic, Marko
    Klaproth, Oliver K.
    Hengerer, Fritz H.
    Mayer, Wolfgang J.
    Kohnen, Thomas
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10): : 1587 - 1592
  • [7] Comparison of anterior capsulotomy techniques: continuous curvilinear capsulorhexis, femtosecond laser-assisted capsulotomy and selective laser capsulotomy
    Daya, Sheraz
    Chee, Soon-Phaik
    Ti, Seng-Ei
    Packard, Richard
    Mordaunt, David H.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2020, 104 (03) : 437 - 442
  • [8] Electron microscopy of laser capsulotomy edge: Interplatform comparison
    Bala, Chandra
    Xia, Yong
    Meades, Kerrie
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (08): : 1382 - 1389
  • [9] Anterior Capsulotomy Outcomes: A Comparison Between Two Femtosecond Laser Cataract Surgery Platforms
    Pantanelli, Seth M.
    Diakonis, Vasilios F.
    Al-Mohtaseb, Zaina
    Cabot, Florence
    Yesilirmak, Nilufer
    Kounis, George A.
    Sayed-Ahmed, Ibrahim O.
    Waren, Daniel
    Yoo, Sonia H.
    Donaldson, Kendall E.
    JOURNAL OF REFRACTIVE SURGERY, 2015, 31 (12) : 821 - 825
  • [10] Scanning electron microscopy evaluation of capsulorhexis in femtosecond laser-assisted cataract surgery
    Mastropasqua, Leonardo
    Toto, Lisa
    Calienno, Roberta
    Mattei, Peter A.
    Mastropasqua, Alessandra
    Vecchiarino, Luca
    Di Iorio, Donato
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10): : 1581 - 1586