Feasibility of late acquisition [68Ga]Ga-PSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer-first clinical experiences

被引:34
|
作者
Alberts, Ian [1 ]
Prenosil, George [1 ]
Mingels, Clemens [1 ]
Bohn, Karl Peter [1 ]
Viscione, Marco [1 ]
Sari, Hasan [1 ,2 ]
Rominger, Axel [1 ]
Afshar-Oromieh, Ali [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Nucl Med, Inselspital, Freiburgstr 18, CH-3010 Bern, Switzerland
[2] Siemens Healthcare AG, Adv Clin Imaging Technol, Lausanne, Switzerland
关键词
Total body; Ultra-long FOV PET; Whole body; PET; CT; Positron-emission tomography; Digital PET; GA-68-PSMA-11; HBED-CC; RADIATION-DOSIMETRY;
D O I
10.1007/s00259-021-05438-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose While acquisition of images in [(68) Ga]Ga-PSMA-11 following longer uptake times can improve lesion uptake and contrast, resultant imaging quality and count statistics are limited by the isotope's half-life (68 min). Here, we present a series of cases demonstrating that when performed using a long axial field-of-view (LAFOV) PET/CT system, late imaging is feasible and can even provide improved image quality compared to regular acquisitions. Methods In this retrospective case series, we report our initial experiences with 10 patients who underwent standard imaging at 1 h p.i. following administration of 192 +/- 36 MBq [(68) Ga]Ga-PSMA-11 with additional late imaging performed at 4 h p.i. Images were acquired in a single bed position for 6 min at 1 h p.i. and 16 min p.i. at 4 h p.i. using a LAFOV scanner (106 cm axial FOV). Two experienced nuclear medicine physicians reviewed all scans in consensus and evaluated overall image quality (5-point Likert scale), lesion uptake in terms of standardised uptake values (SUV), tumour to background ratio (TBR) and target-lesion signal to background noise (SNR). Results Subjective image quality as rated on a 5-point Likert scale was only modestly lower for late acquisitions (4.2/5 at 4 h p.i.; 5/5 1 h p.i.), TBR was significantly improved (4 h: 3.41 vs 1 h: 1.93, p < 0.001) and SNR was improved with borderline significance (4 h: 33.02 vs 1 h: 24.80, p = 0.062) at later imaging. Images were obtained with total acquisition times comparable to routine examinations on standard axial FOV scanners. Conclusion Late acquisition in tandem with a LAFOV PET/CT resulted in improvements in TBR and SNR and was associated with only modest impairment in subjective visual imaging quality. These data show that later acquisition times for [(68) Ga]Ga-PSMA-11 may be preferable when performed on LAFOV systems.
引用
收藏
页码:4456 / 4462
页数:7
相关论文
共 50 条
  • [1] Feasibility of late acquisition [68Ga]Ga-PSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer—first clinical experiences
    Ian Alberts
    George Prenosil
    Clemens Mingels
    Karl Peter Bohn
    Marco Viscione
    Hasan Sari
    Axel Rominger
    Ali Afshar-Oromieh
    European Journal of Nuclear Medicine and Molecular Imaging, 2021, 48 : 4456 - 4462
  • [2] First experiences with late acquisition [68Ga]GaPSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer
    Alberts, I.
    Prenosil, G.
    Mingels, C.
    Bohn, K.
    Viscione, M.
    Sari, H.
    Rominger, A.
    Afhsar-Oromieh, A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S264 - S265
  • [3] The superior detection rate of total-body [68Ga]Ga-PSMA-11 PET/CT compared to short axial field-of-view [68Ga]Ga-PSMA-11 PET/CT for early recurrent prostate cancer patients with PSA &lt;0.2 ng/mL after radical prostatectomy
    Wang, Yining
    Dong, Liang
    Zhao, Haitao
    Li, Lianghua
    Huang, Gang
    Liu, Jianjun
    Chen, Ruohua
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [4] The image quality and feasibility of solitary delayed [68 Ga]Ga-PSMA-11 PET/CT using long field-of-view scanning in patients with prostate cancer
    Yu, Xiaofeng
    Xu, Lian
    Huang, Gang
    Liu, Jianjun
    Chen, Ruohua
    Chen, Yumei
    EJNMMI RESEARCH, 2024, 14 (01)
  • [5] The image quality and feasibility of solitary delayed [68 Ga]Ga-PSMA-11 PET/CT using long field-of-view scanning in patients with prostate cancer
    Xiaofeng Yu
    Lian Xu
    Gang Huang
    Jianjun Liu
    Ruohua Chen
    Yumei Chen
    EJNMMI Research, 14
  • [6] The superior detection rate of total-body [68Ga]Ga-PSMA-11 PET/CT compared to short axial field-of-view [68Ga]Ga-PSMA-11 PET/CT for early recurrent prostate cancer patients with PSA &lt; 0.2 ng/mL after radical prostatectomy
    Wang, Yining
    Dong, Liang
    Zhao, Haitao
    Li, Lianghua
    Huang, Gang
    Xue, Wei
    Liu, Jianjun
    Chen, Ruohua
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 (08) : 2484 - 2494
  • [7] The impact of (68Ga)-Ga-PSMA-11 PET/CT in primary staging of prostate cancer
    Weitzer, F.
    Pernthaler, B.
    Plhak, E.
    Riedl, R.
    Aigner, R. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S458 - S458
  • [8] The influence of digital PET/CT on diagnostic certainty and interrater reliability in [68Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer
    Ian Alberts
    Jan-Niklas Hünermund
    Christos Sachpekidis
    Clemens Mingels
    Viktor Fech
    Karl Peter Bohn
    Axel Rominger
    Ali Afshar-Oromieh
    European Radiology, 2021, 31 : 8030 - 8039
  • [9] The influence of digital PET/CT on diagnostic certainty and interrater reliability in [68Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer
    Alberts, Ian
    Huenermund, Jan-Niklas
    Sachpekidis, Christos
    Mingels, Clemens
    Fech, Viktor
    Bohn, Karl Peter
    Rominger, Axel
    Afshar-Oromieh, Ali
    EUROPEAN RADIOLOGY, 2021, 31 (10) : 8030 - 8039
  • [10] Total-body PET/CT with half-dose [68 Ga]Ga-PSMA-11 for biochemical recurrent prostate cancer: comparable diagnostic value to short axial field-of-view PET/CT with full-dose [68 Ga]Ga-PSMA-11
    Liu, Yuheng
    Li, Lianghua
    Qin, Yanyu
    Chen, Zijun
    Zhao, Haitao
    Wang, Xinlu
    Chen, Ruohua
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 (02) : 581 - 589