Optimally Enhanced Heating for Focused Ultrasound Surgery with Split Foci, Dual-Frequency, or Multi Foci

被引:7
|
作者
Lu, Mingzhu [1 ]
Guan, Yubo [1 ]
Dong, Tengju [1 ]
Liu, Fenfen [1 ]
Wan, Mingxi [1 ]
机构
[1] Xi An Jiao Tong Univ, Sch Life Sci & Technol, Dept Biomed Engn, Key Lab Biomed Informat Engn,Minist Educ, Xian 710049, Peoples R China
基金
中国国家自然科学基金;
关键词
ACOUSTIC CAVITATION; PHASED-ARRAY;
D O I
10.1063/1.4976602
中图分类号
O59 [应用物理学];
学科分类号
摘要
' To substantially enhance heating in HIFU treatment, several methods such as split foci, multi foci, and dual-frequency modes are used. The enhanced-cavitation heating protocols are implemented experimentally in BSA gel-phantom using four-element split-focus array. Using dual frequency of 1.2 and 2.4 MHz, the superimposing of two frequency pressures at confocal region can enhance nucleation cavitation and inertial cavitation activity. When using 135 degrees phase shift combined with dual frequency of 1.2 and 2.4 MHz, the peak negative pressure reach maximum due to peak-negative pressures of two frequencies occur at same time, resulting strong cavitation activities. When using dual frequency of 1.2 and 2.4 MHz, 25-Hz pulse-repetition frequency (PRF), both 135 degrees and 180 degrees phase shift protocols, the experiment results show the largest lesion size of 10.5 x 10.5 x 11 mm(3), quickest lesion inception time of less than 0.2 s, therefore, both 135 degrees and 180 degrees phase shift protocols are most efficient in enhanced-cavitation heating. The filtered-PCD mean square waveforms reveal that the strong inertial-cavitation activities involve in those two treatments. The lesion size of four foci of 180 degrees phase shift, single frequency, 25 Hz PRF, is 2 times that of 0 degrees phase shift, single frequency even if the peak intensity of 180 degrees case is half less than that of 0 degrees phase shift case. When arrange multi foci using phased array in a style of a wavelength distance between neighbor foci in focal plane, the result is the same as that using split foci of 180 degrees phase shift and single frequency.
引用
收藏
页数:5
相关论文
共 40 条
  • [1] Enhanced Cavitation Activities from Axial Split Foci Using Second/Third-Harmonic Superimposition for Focused Ultrasound Surgery
    Lu, Mingzhu
    Guan, Yubo
    Li, Yujiao
    Huang, Weijun
    Ma, Fengcao
    Wan, Mingxi
    [J]. 2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [2] Thrombolysis enhanced by dual-frequency high-intensity focused ultrasound
    Guo, Sijia
    Suo, Dingjie
    Jing, Yun
    Jiang, Xiaoning
    Frank, Jonathan
    Lin, Weili
    [J]. 2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 2173 - 2176
  • [3] Highly efficient cavitation-enhanced heating with dual-frequency ultrasound exposure in high-intensity focused ultrasound treatment
    Sasaki, Hiroshi
    Yasuda, Jun
    Takagi, Ryo
    Miyashita, Takuya
    Goto, Kota
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2014, 53 (07)
  • [4] Reducing lesion aberration by dual-frequency focused ultrasound ablations
    Chen, Wen-Shiang
    Liu, Hao-Li
    Tung, Yao-Sheng
    Wang, Jen-Chieh
    Ding, Yu-Hui
    Jan, Chen-Kai
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2011, 27 (07) : 637 - 647
  • [5] Dual-frequency high intensity focused ultrasound (HIFU) accelerating therapy
    He, P. Z.
    Shou, W. D.
    Duan, S. M.
    Xia, R. M.
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 213 - 216
  • [6] Enhanced-Cavitation Heating Protocols in Focused Ultrasound Surgery With Broadband Split-Focus Approach
    Lu, Mingzhu
    Shi, Yu
    Fang, Li
    Guan, Yubo
    Jiang, Hujie
    Wan, Mingxi
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2014, 61 (04) : 631 - 646
  • [7] High-Intensity Focused Ultrasound Ablation by the Dual-Frequency Excitation
    Law, Siu Kwun Berlinda
    Zhou, Yufeng
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2019, 66 (01) : 18 - 25
  • [8] Collisionless heating in capacitive discharges enhanced by dual-frequency excitation
    Turner, MM
    Chabert, P
    [J]. PHYSICAL REVIEW LETTERS, 2006, 96 (20)
  • [9] Dual-Frequency Piezoelectric Transducers for Contrast Enhanced Ultrasound Imaging
    Martin, K. Heath
    Lindsey, Brooks D.
    Ma, Jianguo
    Lee, Mike
    Li, Sibo
    Foster, F. Stuart
    Jiang, Xiaoning
    Dayton, Paul A.
    [J]. SENSORS, 2014, 14 (11) : 20825 - 20842
  • [10] Split-focused ultrasound for breast tumor thermal surgery with multidirectional heating
    Cheng, Tze-Yuan
    Ju, Kuen-Cheng
    Ho, Cheng-Shiao
    Chen, Yung-Yaw
    Lin, Win-Li
    [J]. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1776 - +