Application of naphthalene-captopril column for pre-concentration and determination of trace amount of copper(II) in waste water samples by flame atomic absorption spectrometry

被引:0
|
作者
Mousavi, H. Zavvar [1 ]
机构
[1] Semnan Univ, Coll Sci, Dept Chem, Semnan, Iran
关键词
pre-concentration; copper; captopril; flame atomic absorption spectrometry;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A sensitive and selective pre-concentration method for the determination of copper(II) Py flame atomic absorption spectrometry has been developed. Copper forms a complex with captopril. This complex is retained by naphthalene- captopril adsorbent in a column with a height of ca. 2.5 cm. The adsorbed metal complex is then eluted from the column with nitric acid and its copper content is determined by flame atomic absorption spectrometry. The effect of different variables such as pH, reagent concentration, flow rate and interfering ions on the recovery of the analyte was investigated. The calibration graph was linear in the range 75-350 mu g L-1 of copper in the initial solution. Detection limits (3 sigma, n 10) of 23 mu g L-1 and relative standard deviation 2.4 and 1.6 % at 200 and 300 mu g L-1 level were obtained. This procedure was successfully applied to the determination of copper in waste water samples. Analytical recoveries of copper added to samples were 98-103%.
引用
收藏
页码:3519 / 3524
页数:6
相关论文
共 50 条