Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion

被引:99
|
作者
Segal, HC [1 ]
Briggs, C
Kunka, S
Casbard, A
Harrison, P
Machin, SJ
Murphy, MF
机构
[1] Churchill Hosp, Oxford Haemophilia Ctr, Oxford OX3 7LJ, England
[2] Churchill Hosp, Thrombosis Unit, Oxford OX3 7LJ, England
[3] UCL Hosp, Dept Haematol, London, England
[4] Natl Blood Serv, Oxford, England
[5] MRC, Clin Trials Unit, London, England
[6] Univ Oxford, John Radcliffe Hosp, Dept Haematol, Oxford OX3 9DU, England
[7] Univ Oxford, Oxford OX1 2JD, England
关键词
thrombocytopenia; platelet counting; haematology analysers; prophylactic platelet transfusion thresholds;
D O I
10.1111/j.1365-2141.2004.05352.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Although haematology analysers provide reliable full blood counts, they are known to be inaccurate at enumerating platelets in severe thrombocytopenia. If the thresholds for platelet transfusion, currently set at 10 x 10(9)/l, are to be further reduced, it is vital that the limitations of current analysers are fully understood. The aim of this large multicentre study was to determine the accuracy of haematology analysers in current routine practice for platelet counts below 20 x 10(9)/l. Platelet counts estimated by analysers using optical, impedance and immunological methods were compared with the International Reference Method for platelet counting. The results demonstrated variation in platelet counting between different analysers and even the same type of analyser at different sites. Optical methods for platelet counting on the XE 2100, Advia 120, Cell-Dyn 4000 and H3* were not superior to impedance methods on the XE 2100, LH750 and Pentra analysers. All analysers except one overestimated the platelet count, which would result in under transfusion of platelets. This study highlights the inaccuracies of haematology analysers in platelet counting in severe thrombocytopenia. It re-emphasizes the need for external quality control to improve analyser calibration for samples with low platelet counts, and suggests that the optimal thresholds for prophylactic platelet transfusions should be re-evaluated.
引用
收藏
页码:520 / 525
页数:6
相关论文
共 50 条
  • [1] ACCURACY OF PLATELET COUNTING BY HAEMATOLOGY ANALYSERS AND ITS POTENTIAL IMPACT ON THE TRANSFUSION DECISION-MAKING PROCESS
    Lopez Rubio, M.
    Magro, E.
    Pascual, T.
    Garcia Suarez, J.
    Gil Fernandez, J. J.
    Calero, M. A.
    Martin Guerrero, Y.
    Callejas, M.
    Perez Calvo, C.
    Burgaleta, C.
    HAEMATOLOGICA-THE HEMATOLOGY JOURNAL, 2009, 94 : 316 - 316
  • [2] PLATELET COUNTING IN THROMBOCYTOPENIA
    KOEPKE, JA
    HAEMOSTASIS, 1982, 12 (1-2) : 13 - 13
  • [3] A large multi-centre evaluation of the accuracy of platelet counting in severe thrombocytopenia.
    Machin, SJ
    Segal, HC
    Briggs, C
    Harrison, P
    Kunka, S
    Murphy, MF
    BLOOD, 2003, 102 (11) : 82B - 82B
  • [4] Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusion
    Hong, K. H.
    Kim, M. J.
    Lee, K. W.
    Park, K. U.
    Kim, H. S.
    Song, J.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (03) : 298 - 306
  • [5] Platelet Transfusion in Patients with Thrombocytopenia
    Gilstad, Colleen W.
    NEW ENGLAND JOURNAL OF MEDICINE, 2023, 389 (05): : 479 - 480
  • [6] Thrombocytopenia and platelet transfusion in the neonate
    Cremer, Malte
    Sallmon, Hannes
    Kling, Pamela J.
    Buehrer, Christoph
    Dame, Christof
    SEMINARS IN FETAL & NEONATAL MEDICINE, 2016, 21 (01): : 10 - 18
  • [7] DONOR PLATELET KINETICS IN THROMBOCYTOPENIA REFRACTORY TO PLATELET TRANSFUSION
    PETERS, AM
    SAVERYMUTTU, SH
    MALIK, F
    ZUIABLE, A
    SCHWALZ, M
    LEWIS, SM
    GORDONSMITH, EC
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1985, 11 (2-3): : A29 - A29
  • [8] Potential Clinical Impact of Inaccurate Automated Platelet Counts in the Setting of Severe Thrombocytopenia
    Marionneaux, Steven M.
    Maslak, Peter G.
    Francisco, Nenita
    Chan, Virgil
    Hanenberg, Justin
    Lynch, Jean
    Rafael, Josephine
    Yao, Jingqiu
    Chua, Conchita
    BLOOD, 2012, 120 (21)
  • [9] The Performance of Haematology Analyzers and the Platelet Transfusion Threshold
    Papa, F.
    Cedrone, M.
    Rongioletti, M.
    Fenu, S.
    Chierichini, A.
    Panetta, V.
    Pasqualetti, P.
    Lenti, L.
    Annino, L.
    Liumbruno, G. M.
    TRANSFUSION, 2009, 49 : 163A - 163A
  • [10] The effects of various platelet transfusion thresholds in preterm infants with severe thrombocytopenia
    Shah, Rita
    Suresh, Gautham
    ACTA PAEDIATRICA, 2020, 109 (01) : 206 - 206