A new sequential injection analysis-capillary electrophoresis system with amperometric detection

被引:3
|
作者
Feito, Rodrigo Fernandez [1 ]
Dinsdale, Richard M. [1 ]
Guwy, Alan J. [1 ]
Premier, Giuliano C. [1 ]
机构
[1] Univ South Wales, Fac Comp Engn & Sci, Sustainable Environm Res Ctr SERC, Pontypridd CF37 1DL, M Glam, Wales
基金
英国自然环境研究理事会; 英国工程与自然科学研究理事会; 英国生物技术与生命科学研究理事会;
关键词
Amperometric detection; Capillary electrophoresis; Electrode alignment; Fully automated; Hydrodynamic injection; CONTACTLESS CONDUCTIVITY DETECTION; ELECTROCHEMICAL DETECTION; ANALYSIS MANIFOLD; ORGANIC-ACIDS; FLOW; PRESSURE; ANIONS;
D O I
10.1002/elps.201700464
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel and fully automated sequential injection analysis manifold coupled to a capillary electrophoresis apparatus with amperometric detection, is described. The sequential injection manifold was isolated from the high voltage by inserting an air plug into the circuit. Small buffer reservoirs were used to avoid the need to pump fresh buffer to the interface during the electrophoretic separation. No decoupling device was used to mitigate the interference from the high voltage electric field, instead the potential shift induced by the separation voltage, was accounted for. The new hydrodynamic injection method presented is based on the overpressure created in the circuit when a pinch valve is closed for a predetermined time. The injection method yields RSD values of peak height and area below 2.55 and 1.82%, respectively, at different durations of valve closure (n=5). The capillary and working electrode alignment was achieved by adapting a commercial available capillary union. When the electrode was replaced, the alignment method proved to be very reliable, yielding RSD values of peak height and area lower than 2.64 and 2.08%, respectively (n=8). Using this system with a gold microelectrode, dopamine, and epinephrine could be quantified within the concentration range of 1-500 M and detected at a concentration of 0.3 M. The methods here presented could be applied for the development of new capillary electrophoresis systems with amperometric detection and/or to the design of fully automated systems for online process monitoring purposes.
引用
收藏
页码:1754 / 1762
页数:9
相关论文
共 50 条