Novel beta-gamma coincidence measurements using phoswich detectors

被引:32
|
作者
Ely, JH [1 ]
Aalseth, CE [1 ]
McIntyre, JI [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
Radionuclide; Monitoring Network; Typical Application; Analysis Pipeline; Usage Mode;
D O I
10.1007/s10967-005-0044-y
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel method to measure beta-gamma coincidences using a phoswich detector with state-of-the-art pulse shape discrimination techniques has been investigated. A thin CaF2(Eu) and thick NaI(Tl) crystal phoswich detector has been built and tested using gamma and beta test sources as well as xenon gas. The two types of radiation are distinctly seen in the signal processing electronics based on pulse shape analysis. Preliminary results indicate that it will be difficult to achieve sufficient discrimination of the pulse types in application to beta-gamma coincident counting.
引用
收藏
页码:245 / 250
页数:6
相关论文
共 50 条
  • [1] Novel beta-gamma coincidence measurements using phoswich detectors
    J. H. Ely
    C. E. Aalseth
    J. I. McIntyre
    Journal of Radioanalytical and Nuclear Chemistry, 2005, 263 : 245 - 250
  • [2] Beta-gamma coincidence counting using an yttrium aluminum perovskit and bismuth germanate phoswich scintillator
    McIntyre, Justin I.
    Schrom, Brian
    Aalseth, Craig E.
    Cooper, Matthew
    Hayes, James C.
    Heimbigner, Tom R.
    Hossbach, Todd
    Hubbard, Charlie
    Litke, Kevin E.
    Ripplinger, Michael
    Seifert, Carolyn E.
    Suarez, Reynold
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1301 - 1304
  • [3] Modelling of a Single-Channel Beta-Gamma Coincidence Phoswich Detector Using Geant4 for the Conversion Electron Energy Peak Resolution and Beta-Gamma Coincidence Efficiency Improvement
    Zhang, Weihua
    Mekarski, Pawel
    Bean, Marc
    Yi, Jing
    Ungar, Kurt
    PURE AND APPLIED GEOPHYSICS, 2014, 171 (3-5) : 621 - 627
  • [4] SYSTEMATIC ERRORS IN 4 PI BETA-GAMMA COINCIDENCE MEASUREMENTS
    WILLIAMS, A
    HUGHES, FH
    CAMPION, PJ
    METROLOGIA, 1968, 4 (04) : 178 - &
  • [5] ABSOLUTE MEASUREMENTS OF RADIOACTIVITY USING 4PI BETA-GAMMA COINCIDENCE METHOD
    WOOD, M
    PHYSICS IN MEDICINE AND BIOLOGY, 1970, 15 (01): : 133 - &
  • [6] Geant4 Monte Carlo radioxenon beta-gamma coincidence efficiency simulation for a phoswich detector
    Zhang, Weihua
    Mekarski, Pawel
    Lam, Jonathan
    Ungar, Kurt
    Pellerin, Eric
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2010, 285 (03) : 475 - 482
  • [7] Geant4 analysis and optimization of a double crystal phoswich detector for beta-gamma coincidence detection
    Fan, Xing
    Zhang, Xian-Peng
    Tian, Geng
    Yang, Chao-Wen
    NUCLEAR SCIENCE AND TECHNIQUES, 2018, 29 (04)
  • [8] Geant4 Monte Carlo radioxenon beta-gamma coincidence efficiency simulation for a phoswich detector
    Weihua Zhang
    Pawel Mekarski
    Jonathan Lam
    Kurt Ungar
    Eric Pellerin
    Journal of Radioanalytical and Nuclear Chemistry, 2010, 285 : 475 - 482
  • [9] Digital Coincidence System for beta-gamma Coincidence Counting
    Yuan Daqing
    Liu QiZhi
    Yang Yuandi
    Wang Jianqing
    Chen YunDong
    Liang JunChen
    Chen Xilin
    Yao Shunhe
    Guo Xiaoqing
    Wu Yongle
    NUCLEAR PHYSICS A, 2010, 834 (1-4) : 773C - 776C
  • [10] Single channel beta-gamma coincidence detection of radioactive xenon using digital pulse shape analysis of phoswich detector signals
    Hennig, Wolfgang
    Tan, Hui
    Warburton, William K.
    McIntyre, Justin I.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 510 - 514