Verification of absorbed dose using diodes in cobalt-60 radiation therapy

被引:2
|
作者
Gadhi, Muhammad Asghar [1 ,4 ]
Fatmi, Shahab [1 ]
Chughtai, Gul M. [2 ]
Arshad, Muhammad [1 ]
Shakil, Muhammad [3 ]
Rahmani, Uzma Mahmood [1 ]
Imran, Malik Younas [1 ]
Buzdar, Saeed Ahmad [4 ]
机构
[1] Bahawalpur Inst Nucl Med & Oncol BINO, Dept Med Phys, Bahawalpur, Pakistan
[2] MPS MEDIC LLC, Philadelphia, PA USA
[3] Univ Gujrat, Dept Phys, Gujrat, Pakistan
[4] Islamia Univ Bahawalpur, Dept Phys, Med Phys Res Grp, Bahawalpur, Pakistan
关键词
Radiation oncology; Dosimetry; Absorbed dose; Diodes; Entrance/exit dose; Ionizing radiation; RADIOTHERAPY; INCIDENT; ERRORS;
D O I
10.1007/s13246-016-0422-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this work was to enhance the quality and safety of dose delivery in the practice of radiation oncology. To achieve this goal, the absorbed dose verification program was initiated by using the diode in vivo dosimetry (IVD) system (for entrance and exit). This practice was implemented at BINO, Bahawalpur, Pakistan. Diodes were calibrated for making absorbed dose measurements. Various correction factors (SSD, dose non-linearity, field size, angle of incidence, and wedge) were determined for diode IVD system. The measurements were performed in phantom in order to validate the IVD procedure. One hundred and nineteen patients were monitored and 995 measurements were performed. For phantom, the percentage difference between measured and calculated dose for entrance setting remained within +/- 2 % and for exit setting +/- 3 %. For patient measurements, the percentage difference between measured and calculated dose remained within +/- 5 % for entrance/open fields and +/- 7 % for exit/wedge/oblique fields. One hundred and nineteen patients and 995 fields have been monitored during the period of 6 months. The analysis of all available measurements gave a mean percent deviation of +/- 1.19 % and standard deviation of +/- 2.87 %. Larger variations have been noticed in oblique, wedge and exit measurements. This investigation revealed that clinical dosimetry using diodes is simple, provides immediate results and is a useful quality assurance tool for dose delivery. It has enhanced the quality of radiation dose delivery and increased/improved the reliability of the radiation therapy practice in BINO.
引用
收藏
页码:211 / 219
页数:9
相关论文
共 50 条
  • [1] Verification of absorbed dose using diodes in cobalt-60 radiation therapy
    Muhammad Asghar Gadhi
    Shahab Fatmi
    Gul M. Chughtai
    Muhammad Arshad
    Muhammad Shakil
    Uzma Mahmood Rahmani
    Malik Younas Imran
    Saeed Ahmad Buzdar
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 211 - 219
  • [3] Radiation absorbed dose for cobalt-60 gamma source in phantoms for different materials
    Akram, Muhammad
    Zulkafal, Hafiz Muhibb Ullah
    Altaf, Saima
    Iqbal, Khalid
    Khan, Muhammad Afzal
    Buzdar, Saeed Ahmed
    JOURNAL OF THE PAKISTAN MEDICAL ASSOCIATION, 2018, 68 (02) : 264 - 267
  • [4] ABSOLUTE DOSE MEASUREMENT OF COBALT-60 RADIATION
    JOHNS, HE
    CORMACK, DV
    WHITMORE, GF
    GARRETT, C
    RADIOLOGY, 1955, 64 (01) : 113 - 114
  • [5] Ionometric determination of absorbed dose to water for cobalt-60 gamma rays
    Boutillon, M.
    Perroche, A.-M.
    Physics in Medicine and Biology, 1993, 38 (03):
  • [6] MULTICURIE COBALT-60 UNITS FOR RADIATION THERAPY
    MORRISON, A
    DIXON, WR
    GARRETT, C
    JOHNS, HE
    BATES, LM
    EPP, ER
    CORMACK, DV
    FEDORUK, SO
    SCIENCE, 1952, 115 (2986) : 310 - 312
  • [7] TRANSMISSION DOSE MEASUREMENT FOR COBALT-60 RADIATION WITH SPECIAL REFERENCE TO ROTATION THERAPY
    FEDORUK, SO
    JOHNS, HE
    BRITISH JOURNAL OF RADIOLOGY, 1957, 30 (352): : 190 - 195
  • [8] The role of Cobalt-60 in modern radiation therapy: Dose delivery and image guidance
    Schreiner, L. John
    Joshi, Chandra P.
    Darko, Johnson
    Kerr, Andrew
    Salomons, Greg
    Dhanesar, Sandeep
    JOURNAL OF MEDICAL PHYSICS, 2009, 34 (03) : 133 - 136
  • [9] Megavoltage Digital Tomosynthesis Using a Radioactive Cobalt-60 Gamma Ray Source for Radiation Therapy Treatment Verification
    MacDonald, A.
    Darko, J.
    Schreiner, L. J.
    MEDICAL PHYSICS, 2009, 36 (09) : 4314 - 4314
  • [10] THE EVALUATION OF THE DOSE TO THE THORAX IN ROTATIONAL COBALT-60 THERAPY
    BURLIN, TE
    BRITISH JOURNAL OF RADIOLOGY, 1957, 30 (358): : 543 - 549