The effect of cement on hip stem fixation: a biomechanical study

被引:9
|
作者
Celik, Talip [1 ]
Mutlu, Ibrahim [1 ]
Ozkan, Arif [2 ]
Kisioglu, Yasin [1 ]
机构
[1] Kocaeli Univ, Dept Biomed Engn, Umuttepe Campus, TR-41380 Kocaeli, Turkey
[2] Duzce Univ, Dept Biomed Engn, Konuralp Campus, TR-81620 Duzce, Turkey
关键词
Hip surgery; Cement; Cemented method; Uncemented method; Finite element analysis; Hip stem; 3-DIMENSIONAL SHAPE OPTIMIZATION; FINITE-ELEMENT-ANALYSIS; ARTHROPLASTY REGISTER; FEMORAL COMPONENT; FATIGUE BEHAVIOR; PROXIMAL FEMUR; BONE; REPLACEMENT; PROSTHESIS; SIMULATION;
D O I
10.1007/s13246-017-0539-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study presents the numerical analysis of stem fixation in hip surgery using with/without cement methods since the use of cement is still controversial based on the clinical studies in the literature. Many different factors such as stress shielding, aseptic loosening, material properties of the stem, surgeon experiences etc. play an important role in the failure of the stem fixations. The stem fixation methods, cemented and uncemented, were evaluated in terms of mechanical failure aspects using computerized finite element method. For the modeling processes, three dimensional (3D) femur model was generated from computerized tomography (CT) images taken from a patient using the MIMICS Software. The design of the stem was also generated as 3D CAD model using the design parameters taken from the manufacturer catalogue. These 3D CAD models were generated and combined with/without cement considering the surgical procedure using SolidWorks program and then imported into ANSYS Workbench Software. Two different material properties, CoCrMo and Ti6Al4V, for the stem model and Poly Methyl Methacrylate (PMMA) for the cement were assigned. The material properties of the femur were described according to a density calculated from the CT images. Body weight and muscle forces were applied on the femur and the distal femur was fixed for the boundary conditions. The calculations of the stress distributions of the models including cement and relative movements of the contacts examined to evaluate the effects of the cement and different stem material usage on the failure of stem fixation. According to the results, the use of cement for the stem fixation reduces the stress shielding but increases the aseptic loosening depending on the cement crack formations. Additionally, using the stiffer material for the stem reduces the cement stress but increases the stress shielding. Based on the results obtained in the study, even when taking the disadvantages into account, the cement usage is more suitable for the hip fixations.
引用
收藏
页码:349 / 357
页数:9
相关论文
共 50 条
  • [1] The effect of cement on hip stem fixation: a biomechanical study
    Talip Çelik
    İbrahim Mutlu
    Arif Özkan
    Yasin Kişioğlu
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 349 - 357
  • [2] Effect of hip stem taper in cement stresses
    New, AMR
    Taylor, M
    Wroblewski, BM
    ORTHOPEDICS, 2005, 28 (08) : S857 - S862
  • [3] The Effect of Screw Length and Bone Cement Augmentation on the Fixation Strength of Iliac Screws A Biomechanical Study
    Zheng, Zhao-Min
    Zhang, Kui-Bo
    Zhang, Jia-Fang
    Yu, Bin-Sheng
    Liu, Hui
    Zhuang, Xin-Ming
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (08): : 545 - 550
  • [4] Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study
    Selvan, VT
    Oakley, MJ
    Rangan, A
    Al-Lami, MK
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (02): : 136 - 141
  • [5] Effect of flexibility of the femoral stem on bone-remodeling and fixation of the stem in a canine total hip arthroplasty model without cement
    Harvey, EJ
    Bobyn, JD
    Tanzer, M
    Stackpool, GJ
    Krygier, JJ
    Hacking, SA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (01): : 93 - 107
  • [6] FEMORAL STEM FIXATION IN HIP-REPLACEMENT - A BIOMECHANICAL COMPARISON OF CEMENTLESS AND CEMENTED PROSTHESES
    PHILLIPS, TW
    MESSIEH, SS
    MCDONALD, PD
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (03): : 431 - 434
  • [7] The biomechanical effect of the collar of a femoral stem on total hip arthroplasty
    Jeon, Insu
    Bae, Ji-Yong
    Park, Jin-Hong
    Yoon, Taek-Rim
    Todo, Mitsugu
    Mawatari, Masaaki
    Hotokebuchi, Takao
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 (01) : 103 - 112
  • [8] Proximal cement fixation in total hip arthroplasty—first results with a new stem design
    Thomas Einsiedel
    Florian Gebhard
    Ilaria Bregolato
    Anja Hiemeier
    Lothar Kinzl
    Markus Schultheiss
    International Orthopaedics, 2008, 32 : 295 - 306
  • [9] An experimental study on the impact of prosthesis temperature on the biomechanical properties of bone cement fixation
    Wanzhuo Chen
    Haining Zhang
    BMC Surgery, 23
  • [10] An experimental study on the impact of prosthesis temperature on the biomechanical properties of bone cement fixation
    Chen, Wanzhuo
    Zhang, Haining
    BMC SURGERY, 2023, 23 (01)