A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training

被引:4
|
作者
Vieira Leocadio, Rodolfo Rocha [1 ,2 ]
Rego Segundo, Alan Kardek [1 ]
Louzada, Cibelle Ferreira [2 ]
机构
[1] Univ Fed Ouro Preto, Escola Minas, Dept Control & Automat Engn DECAT, BR-35400000 Ouro Preto, MG, Brazil
[2] Univ Fed Ouro Preto, Escola Med, Dept Pediat & Adult Clin DECPA, BR-35400000 Ouro Preto, MG, Brazil
关键词
cardiopulmonary resuscitation; sensor; medical mannequins; spirometric techniques; LUNG-VOLUME; FLOW-RATE; AIR VOLUME; IMPEDANCE; RESONANCE; TOMOGRAPHY; ULTRASOUND; CHILDREN; DESIGN; ADULTS;
D O I
10.3390/s19235095
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF20-75%). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 +/- 22, 450 +/- 23, 603 +/- 24, 751 +/- 26, 922 +/- 27, 1021 +/- 30, 1182 +/- 33, 1326 +/- 36, 1476 +/- 37, 1618 +/- 45 and 1786 +/- 56) x 10(-6) m(3) for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) x 10(-6) m(3), respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF20-75%), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] Systematic Review of Capnography with Mask Ventilation during Cardiopulmonary Resuscitation Maneuvers
    Jose Cereceda-Sanchez, Francisco
    Molina-Mula, Jesus
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (03)
  • [2] Ventilation during cardiopulmonary resuscitation
    Segond, N.
    Demoulin, L.
    Viglino, D.
    Savary, D.
    Debaty, G.
    ANNALES FRANCAISES DE MEDECINE D URGENCE, 2024, 14 (06): : 376 - 384
  • [3] Feedback during cardiopulmonary resuscitation
    Bohn, Andreas
    Gude, Philipp
    CURRENT OPINION IN ANESTHESIOLOGY, 2008, 21 (02) : 200 - 203
  • [4] Controlling ventilation during cardiopulmonary resuscitation
    Bowden, Kevin
    RESUSCITATION, 2014, 85 (06) : E81 - E81
  • [5] MEASUREMENT OF VENTILATION DURING CARDIOPULMONARY RESUSCITATION
    ORNATO, JP
    BRYSON, BL
    DONOVAN, PJ
    FARQUHARSON, RR
    JAEGER, C
    CRITICAL CARE MEDICINE, 1983, 11 (02) : 79 - 82
  • [6] Presence of a close relative during cardiopulmonary resuscitation maneuvers
    Carpentier, F.
    ANNALES FRANCAISES DE MEDECINE D URGENCE, 2013, 3 (03): : 200 - 201
  • [7] Human or monitor feedback to improve mask ventilation during simulated neonatal cardiopulmonary resuscitation
    Binder, Corinna
    Schmoelzer, Georg M.
    O'Reilly, Megan
    Schwaberger, Bernhard
    Urlesberger, Berndt
    Pichler, Gerhard
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2014, 99 (02): : F120 - F123
  • [8] Cardiopulmonary resuscitation algorithms, defibrillation and optimized ventilation during resuscitation
    Samson, Ricardo A.
    Berg, Marc D.
    Berg, Robert A.
    CURRENT OPINION IN ANESTHESIOLOGY, 2006, 19 (02) : 146 - 156
  • [9] Hand injuries caused by feedback device usage during cardiopulmonary resuscitation training
    Hong, Jun Young
    Oh, Je Hyeok
    Kim, Chan Woong
    Lee, Dong Hoon
    RESUSCITATION, 2016, 107 : E3 - E4
  • [10] A survey of ventilation strategies during cardiopulmonary resuscitation
    Ye-cheng Liu
    Yan-meng Qi
    Hui Zhang
    Joseph Walline
    Hua-dong Zhu
    World Journal of Emergency Medicine, 2019, (04) : 222 - 227