Failures of stainless steel orthopaedic devices - Causes and remedies

被引:0
|
作者
Mudali, UK [1 ]
Sridhar, TM
Eliaz, N
Raj, B
机构
[1] Indira Gandhi Ctr Atom Res, Met & Mat Grp, Kalpakkam 603102, Tamil Nadu, India
[2] Tel Aviv Univ, Dept Solid Mech Mat & Syst, Biomat Med Devices & Corros Lab, IL-69978 Tel Aviv, Israel
关键词
D O I
暂无
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Orthopaedic implants are exposed to the biochemical and dynamic environments of the human body; their design is dictated by anatomy and restricted by physiological conditions. In every failure of An orthopaedic implant, the concerned patient is made to experience the trauma of repeated surgeries, besides the severe pain experienced during the process of rejection of the device. The removal of the failed implant will cause great expense and hardship to the patient. Therefore, it is highly desirable to keep the number of failures to a minimum. The determination of the mechanism by which failure of an implant occurred is important, but it is also necessary to explore the event, or sequence of events, that had caused a particular mechanism or mechanisms to be operative. Furthermore, failure analyses can help improving the overall performance of implant devices through revision engineering. In the framework of this Study, a survey of failure investigations of stainless steel implants was undertaken. The use of advanced stainless steels and the application of nitrogen ion implantation and bioceramic hydroxyapatite coatings as potential remedies are evaluated. Failure analyses reveal the occurrence of significant localized corrosion attack viz., pitting and crevice corrosion. These findings support the need for the development of some innovative implant materials exhibiting resistance to localized corrosion attack. Cooperative efforts by the materials and medical/surgical disciplines may result in even greater improvements in the future with respect to the durability and safety of these implant devices.
引用
收藏
页码:231 / 267
页数:37
相关论文
共 50 条
  • [1] DENTURES - CAUSES OF FAILURES AND REMEDIES
    MEYER, FS
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1951, 1 (06): : 672 - 691
  • [2] BANK FAILURES - CAUSES AND REMEDIES
    Thomas, R. G.
    [J]. JOURNAL OF BUSINESS OF THE UNIVERSITY OF CHICAGO, 1935, 8 (03): : 297 - 318
  • [3] FAILURES IN STAINLESS-STEEL ORTHOPEDIC IMPLANT DEVICES - A SURVEY
    SIVAKUMAR, M
    DHANADURAI, KSK
    RAJESWARI, S
    THULASIRAMAN, V
    [J]. JOURNAL OF MATERIALS SCIENCE LETTERS, 1995, 14 (05) : 351 - 354
  • [4] Fatigue failures of austenitic stainless steel orthopedic fixation devices
    M. E. Stevenson
    M. E. Barkey
    R. C. Bradt
    [J]. Practical Failure Analysis, 2002, 2 (3) : 57 - 64
  • [5] FAILURE OF STEEL STRUCTURES - CAUSES AND REMEDIES
    WECK, R
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1965, 285 (1400): : 3 - &
  • [6] MIC CAUSES STAINLESS-STEEL TUBE FAILURES - DESPITE HIGH WATER VELOCITY
    KOBRIN, G
    [J]. MATERIALS PERFORMANCE, 1994, 33 (04) : 62 - 63
  • [7] The stainless steel screw as an orthopaedic implant
    Rozman, J
    Kmetic, M
    Bunc, M
    Pihlar, B
    [J]. MATERIALS AND CORROSION-WERKSTOFFE UND KORROSION, 2002, 53 (12): : 908 - 912
  • [8] Unsuccessful hydraulic fracturing cases in Australia: Investigation into causes of failures and their remedies
    Rahman, M. K.
    Suarez, Y. A.
    Chen, Z.
    Rahman, S. S.
    [J]. JOURNAL OF PETROLEUM SCIENCE AND ENGINEERING, 2007, 57 (1-2) : 70 - 81
  • [9] INVESTIGATION OF FAILURES IN STAINLESS-STEEL ORTHOPEDIC IMPLANT DEVICES - PIT-INDUCED FATIGUE CRACKS
    SIVAKUMAR, M
    MUDALI, UK
    RAJESWARI, S
    [J]. JOURNAL OF MATERIALS SCIENCE LETTERS, 1995, 14 (02) : 148 - 151
  • [10] AUSTENITIC STAINLESS-STEEL FERRITIC STEEL WELD JOINT FAILURES
    KLUEH, RL
    KING, JF
    [J]. WELDING JOURNAL, 1982, 61 (09) : S302 - S311