FDA Regulation of Clinical High Intensity Focused Ultrasound (HIFU) Devices

被引:13
|
作者
Harris, Gerald R. [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Silver Spring, MD 20992 USA
关键词
CALIBRATION; MODEL;
D O I
10.1109/IEMBS.2009.5332444
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the U. S., medical devices are regulated under the authority of the 1976 Medical Device Amendments to the Food, Drug, and Cosmetic Act, with the Food and Drug Administration's Center for Devices and Radiological Health having primary responsibility. The Act defines several regulatory paths to market depending on the complexity of the device and indications for use. For most high intensity focused ultrasound (HIFU) devices the premarket submissions include both pre-clinical and clinical data. Pre-clinical testing generally comprises ultrasound power measurements and field characterization, in vitro and in vivo temperature measurements, thermal computational modeling, and demonstrating the accuracy for targeting the region of interest and monitoring treatment progress. Protocols for clinical trials are developed by the device sponsor in conjunction with FDA medical and scientific staff. Currently there are no recognized guidance or standards documents for HIFU testing that could be used in the regulatory review process, but such work is underway within the International Electrotechnical Commission.
引用
收藏
页码:145 / 148
页数:4
相关论文
共 50 条
  • [1] High intensity focused ultrasound (HIFU)
    Barkin, Jack
    CANADIAN JOURNAL OF UROLOGY, 2011, 18 (02) : 5634 - 5643
  • [2] High intensity focused ultrasound (HIFU)
    Ganzer, R.
    RADIOLOGE, 2017, 57 (08): : 659 - 664
  • [3] Issues in Clinical Applications of High Intensity Focused Ultrasound (HIFU)
    Qian, Zu Wen
    CURRENT MEDICAL IMAGING REVIEWS, 2010, 6 (03) : 136 - 144
  • [4] The FDA perspective on pre-clinical testing for high intensity focused ultrasound devices
    Harris, Gerald R.
    Therapeutic Ultrasound, 2006, 829 : 595 - 598
  • [5] Anaesthesia for high intensity focused ultrasound (HIFU) therapy
    Yao, C. -L.
    Trinh, T.
    Wong, G. T. C.
    Irwin, M. G.
    ANAESTHESIA, 2008, 63 (08) : 865 - 872
  • [6] High Intensity Focused Ultrasound (HIFU) in urology.
    Chartier-Kastler, E
    Yonneau, L
    Conort, P
    Haertig, A
    Bitker, MO
    Richard, F
    PROGRES EN UROLOGIE, 2000, 10 (06): : 1108 - 1117
  • [7] Coagulation Using High Intensity Focused Ultrasound (HIFU)
    Nguyen, Phuc V.
    Oh, Junghwan
    Kang, Hyun Wook
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS X, 2014, 8926
  • [8] High intensity focused ultrasound at 3 mhz (hifu)
    Thueoff, S.
    Chaussy, C.
    EUROPEAN UROLOGY SUPPLEMENTS, 2007, 6 (02) : 293 - 293
  • [9] CT and ultrasound guided stereotactic high intensity focused ultrasound (HIFU)
    Wood, Bradford J.
    Yanof, J.
    Frenkel, V.
    Viswanathan, A.
    Dromi, S.
    Oh, K.
    Kruecker, J.
    Bauer, C.
    Seip, R.
    Kam, A.
    Li, K. C. P.
    THERAPEUTIC ULTRASOUND, 2006, 829 : 122 - +
  • [10] Advances of high intensity focused ultrasound (HIFU) for pancreatic cancer
    Li Xiaoping
    Zheng Leizhen
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2013, 29 (07) : 678 - 682