Gas chromatography/ion trap tandem mass spectrometry for the analysis of halobenzenes in soils by solid-phase microextraction

被引:0
|
作者
Sarrión, MN [1 ]
Santos, FJ [1 ]
Galceran, MT [1 ]
机构
[1] Univ Barcelona, Fac Quim, Dept Quim Analit, E-08028 Barcelona, Spain
关键词
D O I
10.1002/1097-0231(20001215)14:23<2271::AID-RCM163>3.0.CO;2-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Headspace solid-phase microextraction combined with gas chromatography/ion trap tandem mass spectrometry (HS-SPME/GC/ITMS/MS) was used for the analysis of 12 halobenzenes from soil samples. For MS/MS optimisation, the experiments were performed by precursor ion selection and software controlled operations. Collision-induced dissociation (CID) can be achieved by two different approaches, resonant and non-resonant excitation modes. Different results were obtained using the two approaches, and the resonant excitation mode was chosen as the best for all halobenzenes. Parameters such as the CID excitation amplitude, excitation RF storage level and CID bandwidth frequency were optimised to maximise the formation of halobenzene product ions. A 100-mum polydimethylsiloxane fibre was used for the isolation and preconcentration of the analytes, The HS-SPME/GC/ITMS/MS method was applied to the analysis of halobenzenes in an agricultural soil sample. The halobenzenes were quantified by standard addition, which led to good reproducibility (RSD between 4.7 and 9.2%) and detection limits in the low pg/g range. The method was validated by comparing the results with those obtained in a European inter-laboratory exercise. Copyright (C) 2000 John Wiley & Sons, Ltd.
引用
收藏
页码:2271 / 2281
页数:11
相关论文
共 50 条