A simple method for determining the coagulation threshold temperature of transparent tissue-mimicking thermal therapy gel phantoms: Validated by magnetic resonance imaging thermometry

被引:7
|
作者
Brodin, N. Patrik [1 ,2 ]
Partanen, Ari [3 ]
Asp, Patrik [4 ]
Branch, Craig A. [5 ]
Guha, Chandan [1 ,2 ]
Tome, Wolfgang A. [1 ,2 ]
机构
[1] Albert Einstein Coll Med, Inst Oncophys, Bronx, NY 10461 USA
[2] Montefiore Med Ctr, Dept Radiat Oncol, Bronx, NY 10461 USA
[3] Clin Sci MR Therapy Philips, Andover, MA 01810 USA
[4] Albert Einstein Coll Med, Liver Res Ctr & Radiat Oncol, Bronx, NY 10461 USA
[5] Albert Einstein Coll Med, Dept Radiol, Bronx, NY 10461 USA
基金
美国国家卫生研究院;
关键词
thermal therapy; focused ultrasound; MRI; tissue-mimicking phantoms; CANCER-THERAPY; HYPERTHERMIA; PRECISE; SHIFT;
D O I
10.1118/1.4941361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Tissue-mimicking thermal therapy phantoms that coagulate at specific temperatures are valuable tools for developing and evaluating treatment strategies related to thermal therapy. Here, the authors propose a simple and efficient method for determining the coagulation threshold temperature of transparent thermal therapy gel phantoms. Methods: The authors used a previously published gel phantom recipe with 2% (w/v) of bovine serum albumin as the temperature-sensitive protein. Using the programmable heating settings of a polymerase chain reaction (PCR) machine, the authors heated 50 mu l gel samples to various temperatures for 3 min and then imaged them using the BioRad Gel Doc system to determine the coagulation temperature using an opacity quantification method. The estimated coagulation temperatures were then validated for gel phantoms prepared with different pH levels using high-intensity focused ultrasound (HIFU) heating and magnetic resonance imaging (MRI) thermometry methods on a clinical MR-HIFU system. Results: The PCR heating method produced consistent and reproducible coagulation of gel samples in precise correlation with the set incubation temperatures. The resulting coagulation threshold temperatures for gel phantoms of varying pH levels were found to be 44.1 +/- 0.1, 53.4 +/- 0.9, and 60.3 +/- 0.9 degrees C for pH levels of 4.25, 4.50, and 4.75, respectively. This corresponded well with the coagulation threshold temperatures determined by MR-thermometry, with coagulation defined as a 95% decrease in T2 relaxation time, which were estimated at 53.6 +/- 1.9 and 62.9 +/- 2.4 degrees C for a pH of 4.50 and 4.75, respectively. Conclusions: The opacity quantification method provides a fast and reproducible estimate of the coagulation threshold temperature of transparent temperature-sensitive gel phantoms. The temperatures determined using this method were well within the range of temperatures estimated using MR-thermometry. Due to the specific heating capabilities of the PCR machine, and the robust determination of coagulation threshold temperatures based on the statistically significant increase in the opacity of gel samples, coagulation temperatures can be determined more precisely and with less variability compared to MRI-based methods. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:1167 / 1174
页数:8
相关论文
共 3 条
  • [1] A Fast and Efficient Method for Determining Coagulation Temperatures of Tissue-Mimicking Thermal Therapy Gel Phantoms: Validated by Magnetic Resonance Thermometry
    Brodin, P.
    Partanen, A.
    Asp, P.
    Branch, C.
    Guha, C.
    Tome, W.
    MEDICAL PHYSICS, 2015, 42 (06) : 3625 - 3625
  • [2] Tissue-Mimicking Gel Phantoms for Thermal Therapy Studies
    Dabbagh, Ali
    Abdullah, Basri Johan Jeet
    Ramasindarum, Chanthiriga
    Abu Kasim, Noor Hayaty
    ULTRASONIC IMAGING, 2014, 36 (04) : 291 - 316
  • [3] TISSUE-MIMICKING GELATIN AGAR GELS FOR USE IN MAGNETIC-RESONANCE IMAGING PHANTOMS
    BLECHINGER, JC